两种三焦点人工晶状体偏心与倾斜对客观视觉质量影响的比较
前瞻性队列研究纳入84例白内障术后患者,比较839MP与Panoptix三焦点人工晶状体的偏心、倾斜及客观视觉质量。839MP组总偏心距离和垂直偏心距离更小,全眼及眼内总高阶像差更低;Panoptix组60 cm裸眼中距离视力更优。两组总倾斜角度和不良视觉现象发生率无显著差异。
要点
- 839MP组总偏心距离(0.27±0.16)mm低于Panoptix组(0.44±0.26)mm,垂直偏心距离也更小。
- 839MP组全眼总高阶像差(0.222±0.094)μm低于Panoptix组(0.324±0.243)μm。
- Panoptix组60 cm裸眼中距离视力优于839MP组(P<0.001),其余视力无显著差异。
核心结论
- 三焦点人工晶状体植入术后早期,相比Panoptix人工晶状体,839MP人工晶状体偏心距离更小、客观视觉质量更好。
- 相同倾斜程度下839MP人工晶状体客观视觉质量更好,相同偏心程度下Panoptix人工晶状体客观视觉质量更好。
研究设计与样本
- 前瞻性队列研究,连续纳入2023年1月至2024年6月在北京大学第三医院行双眼超声乳化白内障吸除联合人工晶状体植入术的白内障患者单眼。
- 按植入人工晶状体类型分为839MP组(AT LISA tri 839MP)和Panoptix组(AcrySof IQ PanOptix),每组招募45例(45只眼),完成随访各42例(42只眼),共84例(84只眼)。
- 术后3个月比较视力、等效球镜度数、人工晶状体偏心距离和倾斜角度、全眼和眼内高阶像差及调制传递函数,并评估不良视觉现象。
主要发现
- 839MP组总偏心距离(0.27±0.16)mm和垂直偏心距离(0.17±0.11)mm低于Panoptix组的(0.44±0.26)mm和(0.35±0.28)mm(均P<0.05)。
- 839MP组总倾斜角度(3.83°±2.15°)与Panoptix组(3.08°±1.70°)差异无统计学意义(P>0.05)。
- 839MP组全眼总高阶像差(0.222±0.094)μm、全眼像差(0.099±0.071)μm和眼内总高阶像差(0.195±0.094)μm均低于Panoptix组的(0.324±0.243)μm、(0.168±0.166)μm和(0.304±0.240)μm(均P<0.05)。
- Panoptix组60 cm裸眼中距离视力优于839MP组(P<0.001),其余裸眼远、中(80 cm)、近视力及最佳矫正远视力组间差异无统计学意义(均P>0.05)。
- 两组人工晶状体倾斜角度和偏心距离与眼内高阶像差无相关性(P>0.05),不良视觉现象发生情况差异无统计学意义(均P>0.05)。
关键数值与推荐
- 全眼高阶像差>0.3 μm时,839MP组偏心距离和倾斜角度截断值分别为0.29 mm(AUC=0.69)和5.1°(AUC=0.79),Panoptix组分别为0.33 mm(AUC=0.76)和3.2°(AUC=0.81)。
- 全眼高阶像差>0.5 μm时,839MP组偏心距离和倾斜角度截断值分别为0.54 mm(AUC=0.64)和6.9°(AUC=0.85),Panoptix组分别为0.61 mm(AUC=0.72)和4.7°(AUC=0.88)。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Objective: To compare the differences in decentration and tilt between two types of trifocal intraocular lenses (IOL) and their impacts on postoperative objective visual quality. Methods: This was a prospective cohort study. One eye was consecutively enrolled from each cataract patient who underwent bilateral phacoemulsification cataract extraction combined with IOL implantation at Peking University Third Hospital between January 2023 and June 2024. Patients were divided into two groups based on the type of implanted IOL: the 839MP group (AT LISA tri 839MP IOL) and the Panoptix group (AcrySof IQ PanOptix IOL). At 3 months postoperatively, the following parameters were compared between the two groups: uncorrected distance, intermediate (80 and 60 cm), and near visual acuities; best-corrected distance visual acuity; spherical equivalent refraction; IOL decentration distance and tilt angle. Additionally, total higher-order aberrations (HOAs), coma, trefoil, spherical aberration of the whole eye and intraocular, and modulation transfer function (MTF) were measured, and the severity of adverse visual phenomena was graded. Pearson correlation analysis was used to explore the correlation between IOL decentration/tilt and total HOAs of the whole eye. Receiver operating characteristic (ROC) curve analysis was performed to determine the cut-off values of IOL decentration and tilt for predicting total HOAs of the whole eye >0.3 μm and >0.5 μm, with the area under the curve (AUC) used to evaluate predictive efficacy. Results: Forty-five patients (45 eyes) were recruited in each group, and 42 patients (42 eyes) in each group completed the follow-up, with a total of 84 patients (84 eyes). Among them, there were 28 males and 56 females, with a mean age of 65 years. No statistically significant differences were observed between the two groups in terms of gender, age, eye laterality, axial length category, IOL power, preoperative lens decentration and tilt, Kappa angle, or Alpha angle (all P>0.05). At 3 months postoperatively, there were no significant differences between the 839MP and Panoptix groups in uncorrected distance, intermediate (80 cm), and near visual acuities or best-corrected distance visual acuity (all P>0.05). However, for uncorrected intermediate visual acuity at 60 cm, the Panoptix group performed better than the 839MP group (P<0.001). Regarding IOL decentration, the total decentration distance and vertical decentration distance in the 839MP group were (0.27±0.16) mm and (0.17±0.11) mm, respectively, which were significantly lower than those in the Panoptix group [(0.44±0.26) mm and (0.35±0.28) mm, respectively; all P<0.05]. For IOL tilt, no significant difference was found in the total tilt angle between the 839MP group (3.83°±2.15°) and the Panoptix group (3.08°±1.70°) (P>0.05). In terms of objective visual quality, the total HOAs of the whole eye, ocular HOAs, and total intraocular HOAs in the 839MP group were (0.222±0.094) μm, (0.099±0.071) μm, and (0.195±0.094) μm, respectively, which were significantly lower than those in the Panoptix group [(0.324±0.243) μm, (0.168±0.166) μm, and (0.304±0.240) μm, respectively; all P<0.05]. No correlation was observed between IOL tilt/decentration and intraocular HOAs in either group (P>0.05). ROC curve analysis showed that for evaluating IOL tolerance to decentration and tilt: when total HOAs of the whole eye>0.3 μm, the cut-off values of IOL decentration and tilt were 0.29 mm (AUC=0.69) and 5.1° (AUC=0.79) in the 839MP group, and 0.33 mm (AUC=0.76) and 3.2° (AUC=0.81) in the Panoptix group; when total HOAs of the whole eye>0.5 μm, the cut-off values of IOL decentration and tilt were 0.54 mm (AUC=0.64) and 6.9° (AUC=0.85) in the 839MP group, and 0.61 mm (AUC=0.72) and 4.7° (AUC=0.88) in the Panoptix group. No significant difference was noted in the incidence of adverse visual phenomena between the two groups (all P>0.05). Conclusions: In the early postoperative period after trifocal IOL implantation, compared with the Panoptix IOL, the 839MP IOL exhibits smaller decentration and better objective visual quality. Furthermore, at the same tilt level, the 839MP IOL provides better objective visual quality; whereas at the same decentration level, the Panoptix IOL shows superior objective visual quality. 目的: 比较两种三焦点人工晶状体(IOL)偏心和倾斜情况的差异,以及对术后客观视觉质量的影响。 方法: 前瞻性队列研究。连续纳入2023年1月至2024年6月在北京大学第三医院行双眼超声乳化白内障吸除联合IOL植入术的白内障患者的单眼进行研究。根据植入IOL类型不同,分为839MP组(AT LISA tri 839MP IOL)和Panoptix组(AcrySof IQ PanOptix IOL)。比较两组术后3个月远、中(80和60 cm)、近距离下的裸眼视力、最佳矫正远视力、等效球镜度数及IOL的偏心距离和倾斜角度,测量全眼和眼内总高阶像差、彗差、三叶草像差、球差以及传递调制函数(MTF),并对不良视觉现象程度进行评分。采用Pearson相关性分析探究IOL偏心距离、倾斜角度与全眼高阶像差的相关性;采用受试者操作特征(ROC)曲线分析,以全眼高阶像差>0.3 μm及>0.5 μm为预测目标,确定IOL偏心距离、倾斜角度的截断值,并通过曲线下面积(AUC)评估其预测效能。 结果: 每组均各招募45例(45只眼)患者,完成随访42例(42只眼)患者,共84例(84只眼),其中男性28例,女性56例,平均年龄65岁。两组间性别、年龄、眼别、眼轴程度、IOL度数、术前晶状体偏心距离和倾斜角度、Kappa角和Alpha角差异均无统计学意义(均P>0.05)。术后3个月,839MP组和Panoptix组间裸眼远、中(80 cm)、近距离视力和最佳矫远正视力差异无统计学意义(均P>0.05),而裸眼中(60 cm)距离视力,Panoptix组优于839MP组(P<0.001)。IOL 偏心方面,839MP 组的总偏心距离和垂直偏心距离分别为(0.27±0.16)和(0.17±0.11)mm明显低于Panoptix组的(0.44±0.26)和(0.35±0.28)mm(均P<0.05);IOL倾斜方面,839MP组总倾斜角度(3.83°±2.15°)与Panoptix组(3.08°±1.70°)比较,两组间差异无统计学意义(P>0.05)。客观视觉质量方面,839MP组的全眼总高阶像差、全眼像差和眼内总高阶像差分别为(0.222±0.094)、(0.099±0.071)和(0.195±0.094)μm,均显著低于Panoptix组的(0.324±0.243)、(0.168±0.166)和(0.304±0.240)μm(均P<0.05)。两组IOL倾斜角度和偏心距离和眼内高阶像差无相关性(P>0.05)。ROC曲线分析结果显示,在IOL偏心与倾斜耐受性评估中,当全眼高阶像差>0.3 μm 时:839MP组IOL偏心距离、倾斜角度的截断值分别为0.29 mm(AUC=0.69)和5.1°(AUC=0.79);Panoptix组分别为0.33 mm(AUC=0.76)和3.2°(AUC=0.81)。当全眼高阶像差>0.5 μm时,839MP组IOL偏心距离、倾斜角度的截断值分别为0.54 mm(AUC=0.64)和6.9°(AUC=0.85);Panoptix组分别为0.61 mm(AUC=0.72)和4.7°(AUC=0.88)。两组患者不良视觉现象发生情况的差异无统计学意义(均P>0.05)。 结论: 三焦点IOL植入术后早期,相比于Panoptix IOL,839MP IOL偏心距离更小,客观视觉质量更好。而在相同倾斜程度下,839MP IOL客观视觉质量更好;在相同偏心程度下,Panoptix IOL客观视觉质量更好。.
引用
- Qin R, Liu YY, Li ZK, et al. [Comparison of the influence of decentration and tilt of two types of trifocal intraocular lenses on objective visual quality]. Zhonghua Yan Ke Za Zhi. 2025. PMID: 41073325. doi:10.3760/cma.j.cn112142-20250205-00053
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。