近视小切口角膜基质透镜取出术后真实角膜屈光力评估方法的比较
前瞻性横断面试验纳入48例89眼SMILE术后眼,比较TCRP、TK及Haigis、Shammas、Maloney、Actual Ka+p校正参数与临床病史法(CHM)的一致性。TK与CHM差异最小(0.41±0.48 D),95%一致性界限宽度最窄(1.88 D),ΔTK与ΔSEco差异0.42 D。结论认为IOLMaster 700的TK参数可提供SMILE术后准确客观的角膜屈光力评估。
要点
- TK与CHM差异最小(0.41±0.48 D),95%一致性界限宽度最窄(1.88 D)。
- ΔTK与ΔSEco差异0.42 D,小于ΔTCRP与ΔSEco的0.62 D。
- ΔTK的95%一致性界限宽度(1.88 D)和相关系数(0.964)均大于ΔTCRP。
核心结论
- IOLMaster 700的TK参数可提供SMILE术后准确客观的角膜屈光力评估。
研究设计与样本
- 前瞻性横断面试验,纳入上海市眼病防治中心接受SMILE的48例患者共89眼。
主要发现
- SMILE术后TK与CHM差异最小(0.41±0.48 D),TK的95%一致性界限宽度最窄(1.88 D),其次为Ka+p(1.94 D)。
- Pearson分析显示各参数与CHM数据均具有良好的相关性。
- ΔTK与ΔSEco差异为0.42 D,ΔTCRP与ΔSEco差异为0.62 D。
- ΔTK的95%一致性界限宽度(1.88 D)和相关系数(0.964)均大于ΔTCRP。
- 一例SMILE术后白内障手术病例证实了上述结论(绝对误差0.005 D)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: We sought to compare the accuracy of different corneal curvature parameters in estimating corneal power change in eyes that underwent myopic small-incision lenticule extraction (SMILE). DESIGN: Prospective cross-sectional trial. METHODS: 89 eyes of 48 patients who had undergone SMILE at Shanghai Eye Disease and Prevention & Treatment Center were enrolled. Total corneal refractive power (TCRP) and total keratometry (TK) values, along with corrected parameters calculated using the Haigis, Shammas, Maloney, and Actual Ka+p methods, were compared with data obtained using the clinical history method (CHM). We also compared the surgically induced changes in TCRP (ΔTCRP) and TK (ΔTK) to changes in the spherical equivalent on the corneal plane (ΔSEco). RESULTS: After SMILE, differences were smallest between values of TK (0.41 ± 0.48 D) and CHM. The width of the 95% limits of agreement of TK (1.88 D) was narrowest, followed by that of Ka+p (1.94 D). Pearson analysis showed that each parameter had a good correlation with CHM data. The differences between ΔTK and ΔSEco was 0.42 D, when between ΔTCRP and ΔSEco was 0.62 D. The width of the 95% limits of agreement of ΔTK (1.88 D) and the correlation coefficient of ΔTK (0.964) were greater than those of ΔTCRP. A separate case of cataract surgery following SMILE confirmed the validity of the above points in our study (the absolute error was 0.005 D). CONCLUSIONS: The TK parameter of the IOLMaster 700 can provide accurate and objective corneal power evaluation after SMILE.
引用
- Fang X, Yu C, Xue W, et al. Comparison of Corneal Power Assessment Methods for True Corneal Power After Myopic Small-Incision Lenticule Extraction. Am J Ophthalmol. 2025. PMID: 40158647. doi:10.1016/j.ajo.2025.03.026
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。