IMO视觉功能分析仪与Humphrey视野分析仪的结构-功能相关性比较
前瞻性横断面研究纳入57眼(57例),比较IMOvifa与HFA 24-2视野及OCT结构参数的相关性。IMOvifa检查时间更短(3.7 vs 5.3分钟),平均缺损更低(差值-0.6 dB),视野指数更高(差值+1.0%),与OCT参数的结构-功能相关强度与HFA无显著差异。
要点
- IMOvifa检查时间3.7分钟,短于HFA的5.3分钟。
- IMOvifa平均缺损较HFA低0.6 dB,视野指数较HFA高1.0%。
- 两种视野计与OCT参数的结构-功能相关强度无显著差异。
核心结论
- IMOvifa检查耗时更短,所致视野缺损略轻于HFA,各分区平均敏感度呈中度至强相关,两种视野计的结构-功能相关性相似。
研究设计与样本
- 前瞻性横断面研究,纳入青光眼疑似、青光眼及高眼压患者57眼(57例),平均年龄59.2岁,女性占61.4%。
- 受试者在12周内完成IMOvifa 24-2、HFA 24-2 Swedish Interactive Threshold Algorithm-Standard及OCT检查。
主要发现
- IMOvifa检查时间3.7分钟 vs HFA 5.3分钟,平均缺损较HFA低0.6 dB,视野指数较HFA高1.0%(p<0.001)。
- IMOvifa与HFA各分区平均敏感度呈中度至强相关(p<0.001)。
- IMOvifa与下方(r=0.46)、上方(r=0.45)、颞侧(r=0.28)视网膜神经纤维层厚度分区及下方(r=0.63)、上方(r=0.53)黄斑区神经节细胞-内丛状层半球厚度显著相关(p≤0.03)。
- IMOvifa与HFA结果和OCT参数的相关强度无显著差异(p≥0.06)。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To compare the structure-function correlations between optical coherence tomography (OCT) thickness parameters and the corresponding sectorial mean sensitivity (MS) values obtained with IMO visual function analyser (IMOvifa) and Humphrey field analyser (HFA) perimeters. DESIGN: Prospective cross-sectional study. METHODS: Glaucoma suspect, glaucoma and ocular hypertension patients underwent IMOvifa 24-2 Ambient Interactive Zippy Estimated Sequential Testing, HFA 24-2 Swedish Interactive Threshold Algorithm-Standard and OCT tests within 12 weeks. Based on the Garway-Heath map, the sensitivity points were divided in corresponding OCT peripapillary retinal nerve fibre layer (pRNFL) sectors, and also the central 10° points in corresponding ganglion cell-inner plexiform layer thickness over the macula. The structure-function correlations were analysed by Spearman's rank coefficient and compared by Steiger's test. RESULTS: 57 eyes of 57 patients (mean age 59.2 years, 61.4% female) with reliable field and adequate OCT strength were included in this analysis. IMOvifa had significantly shorter testing time (3.7 vs 5.3 mins), lower mean deviation (difference -0.6 dB) and greater visual field index (difference +1.0%) compared with HFA (p<0.001 for all). There was moderate to strong correlation between IMOvifa and HFA MS sectors (p<0.001 for all). IMOvifa had significant correlations in the inferior (I, r=0.46), superior (S, r=0.45), temporal (T, r=0.28) pRNFL sectors and in the I (r=0.63) and S (r=0.53) mGCPIPL hemispheres (p≤0.03 for all). There was no significant difference in the strength of the correlation of IMOvifa versus HFA results with OCT parameters (p≥0.06). CONCLUSIONS: The IMOvifa test took less time and resulted in slightly less severe defects than HFA. MS in all sectors was moderately to strongly correlated. Function-structure correlations were similar when using either perimeter.
引用
- Kang J, Nascimento E Silva R, Kim JA, et al. Comparison of structure-function correlation among IMO visual function analyser and Humphrey field analyser. Br J Ophthalmol. 2025. PMID: 40086867. doi:10.1136/bjo-2023-324846
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。