PAUL青光眼引流器术后滤过泡的超声显像
本研究对65例患者共70只眼在PAUL青光眼引流器植入术后行15 MHz超声检查,描述滤过泡形态并分析其与眼压的相关性。结果显示55只眼(78.6%)为双层滤过泡,双层滤过泡眼压显著低于单层(12.09 mm Hg vs 16.47 mm Hg,p<0.01);滤过泡总高度>2 mm者眼压显著低于<2 mm者(p=0.041)。
要点
- PAUL青光眼引流器术后超声可评估滤过泡形态,双层滤过泡与较低眼压显著相关。
- 双层滤过泡眼压12.09 mm Hg,单层滤过泡眼压16.47 mm Hg,差异有统计学意义(p<0.01)。
- 滤过泡总高度>2 mm者平均眼压显著低于<2 mm者(p=0.041)。
核心结论
- 超声是评估PAUL青光眼引流器植入术后滤过泡特征的合适方法。
- 双层液体滤过泡与较低眼压值显著相关。
- 滤过泡高度与眼压呈反向关联,较高滤过泡(>2 mm)显示显著较低眼压。
研究设计与样本
- 研究纳入65例患者共70只眼,均为接受PAUL青光眼引流器植入术后行15 MHz超声检查(Compact Touch,Quantel Medical)者。
- 测量滤过泡尺寸并评估滤过泡形态,同时记录所有临床数据。
主要发现
- 55只眼(78.6%)为双层滤过泡,15只眼(21.4%)为单层滤过泡,其中6只眼(8.6%)位于底板以上、9只眼(12.9%)位于底板以下。
- 双层滤过泡总高度平均3.52 mm,平均眼压13.03 mm Hg。
- 单层与双层滤过泡间眼压差异显著(p<0.01):单层16.47 mm Hg vs 双层12.09 mm Hg。
- 滤过泡总高度(p=0.143)及滤过泡延伸(T:p=0.629;L:p=0.742)与眼压无显著相关性。
- 滤过泡高度>2 mm者平均眼压显著低于<2 mm者(p=0.041)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND/AIMS: Bleb characteristics in the area around the plate following glaucoma drainage device (GDD) surgery are difficult to evaluate on slit lamp examination. Ultrasound sonography could be used for more detailed visualisation. The aim of this study was to describe bleb configuration after PAUL glaucoma implant (PGI) surgery and evaluate the possible correlation with intraocular pressure (IOP) in order to derive new markers for evaluating GDD function. METHODS: Patients underwent a 15 MHz ultrasound examination (Compact Touch, Quantel Medical) after undergoing PGI implantation. The filtering bleb dimensions were measured and bleb configuration was assessed along with documentation of all clinical data. RESULTS: A total of 70 eyes from 65 patients were included. 55 eyes (78.6%) had a double-layered bleb whereas 15 eyes (21.4%) had a bleb with only one compartment (six eyes, 8.6% above; nine eyes, 12.9% below the plate). The total height of both compartments averaged 3.52 mm. The average IOP was 13.03 mm Hg. There was a significant correlation between single and double-layered blebs (p<0.01) with regard to IOP (one compartment: 16.47 mm Hg vs two compartments: 12.09 mm Hg). There was no significant correlation between total bleb height (p=0.143) or bleb extension (T: p=0.629; L: p=0.742). However, total bleb height showed an inverse association with IOP. Blebs with a height >2 mm had significantly lower mean IOP than those having <2 mm bleb heights (p=0.041). CONCLUSIONS: Sonography is a suitable modality to evaluate bleb characteristics after PGI implantation. Blebs with a double fluid layer are significantly correlated with lower IOP values. Bleb height has an inverse association with IOP, with higher blebs (>2 mm) showing a significantly lower IOP.
引用
- Weber C, Weinhold L, Walz W, et al. Sonographic bleb visualisation after PAUL glaucoma implant surgery. Br J Ophthalmol. 2025. PMID: 39689979. doi:10.1136/bjo-2024-326168