光学相干断层扫描血管成像测量青光眼黄斑局灶性灌注丢失

Glaucomatous Focal Perfusion Loss in the Macula Measured by Optical Coherence Tomographic Angiography.
PubMed 文献 · Am J Ophthalmol · 2024-07-14

本前瞻性病例对照研究纳入60例原发性开角型青光眼(POAG)患者和37名健康对照,采用OCTA测量黄斑低灌注面积(LPA)和局灶性灌注丢失(FPL)。结果显示,POAG组LPA和FPL显著高于正常组(P < .001),对视野缺损期青光眼的诊断准确性高(AROC分别为0.993和0.990),且与中心视野平均偏差相关性好。

要点

  • POAG组LPA为5.78±6.30 mm²,显著高于正常组的0.16±0.38 mm²(P < .001)。
  • POAG组FPL为7.52%±8.84%,显著高于正常组的0.20%±0.47%(P < .001)。
  • LPA和FPL对视野缺损期青光眼的诊断AROC分别为0.993和0.990,优于GCLP VD和GCC厚度。

核心结论

  • 使用OCTA评估黄斑FPL有助于评估青光眼性损害。

研究设计与样本

  • 前瞻性、横断面病例对照比较研究,共分析60例POAG患者和37名健康参与者,每名参与者一只眼接受AngioVue 6×6毫米高清黄斑OCTA扫描。

主要发现

  • 正常组LPA(GCLP-VD)为0.16±0.38 mm²,青光眼组为5.78±6.30 mm²(P < .001)。
  • 正常组FPL(GCLP-VD)为0.20%±0.47%,青光眼组为7.52%±8.84%(P < .001)。
  • 视野缺损期青光眼诊断AROC:LPA为0.993,FPL为0.990;在95%特异性下敏感性分别为96.7%和93.3%。
  • LPA和FPL的组内相关系数分别为0.957和0.952,重复性良好。
  • LPA和FPL与中心视野平均偏差的Pearson r分别为-0.716和-0.705(均P < .001)。

关键数值与推荐

  • LPA定义为局部血管密度低于健康参考人群0.5百分位以下、连续面积超过健康参考人群98.5百分位的区域。
  • FPL为LPA范围内相对于正常均值整合的血管密度丢失百分比。
  • 诊断准确性比较:LPA和FPL的AROC分别为0.993和0.990,优于GCLP VD(AROC 0.950,敏感性83.3%)、OCT GCC厚度(AROC 0.927,敏感性80.0%)和GCC局灶性丢失体积(AROC 0.957,敏感性80.0%)。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To measure low perfusion area (LPA) and focal perfusion loss (FPL) in the macula using optical coherence tomography (OCT) angiography (OCTA) for glaucoma. DESIGN: Prospective, cross-sectional "case-control" comparison study. METHODS: A total of 60 patients with primary open-angle glaucoma (POAG) and 37 healthy participants were analyzed. AngioVue 6 × 6-mm high-definition (400 × 400 transverse pixels) macular OCTA scans were performed on one eye of each participant. Flow signal was calculated using the split-spectrum amplitude-decorrelation angiography algorithm. En face ganglion cell layer plexus (GCLP) and superficial vascular complex (SVC) images were generated. Using custom software, vessel density (VD) maps were obtained by computing the fraction of area occupied by flow pixels after low-pass filtering by local averaging 41 × 41 pixels. LPA was defined by local VD below 0.5 percentile over a contiguous area above 98.5 percentile of the healthy reference population. The FPL was the percentage VD loss (relative to normal mean) integrated over the LPA. RESULTS: Among patients with POAG, 30 had perimetric and 30 had preperimetric glaucoma. The LPAGCLP-VD was 0.16±0.38 mm2 in normal and 5.78±6.30 mm2 in glaucoma subjects (P < .001). The FPLGCLP-VD was 0.20%±0.47% in normal and 7.52%±8.84% in glaucoma subjects (P < .001). The perimetric glaucoma diagnostic accuracy, measured by the area under the receiver operating curve, was 0.993 for LPAGCLP-VD and 0.990 for FPLGCLP-VD. The sensitivities were, respectively, 96.7% and 93.3% at 95% specificity. The LPAGCLP-VD and FPLGCLP-VD had good repeatability (0.957 and 0.952 by intraclass correlation coefficient). Diagnostic accuracy was better than GCLP VD (AROC 0.950, sensitivity 83.3%) and OCT ganglion cell complex (GCC) thickness (AROC 0.927, sensitivity 80.0%) and GCC focal loss volume (AROC 0.957, sensitivity 80.0%). The LPAGCLP-VD and FPLGCLP-VD correlated well with central VF mean deviations (Pearson r = -0.716 and -0.705 respectively, both P < .001). CONCLUSION: Assessment of macular FPL using OCTA is useful in evaluating glaucomatous damage.