钙化性玻璃膜疣妨碍扫频源光学相干断层扫描血管成像检测其下毛细血管层
本研究在正常眼、软性玻璃膜疣眼和钙化性玻璃膜疣眼中测量毛细血管层血流缺损,发现钙化性玻璃膜疣因高散射产生伪影信号,无法可靠量化其下毛细血管层血流缺损,需识别并排除钙化性玻璃膜疣。
要点
- 钙化性玻璃膜疣的高散射特性产生散射尾伪影,模拟毛细血管层信号但无血流信息。
- 钙化性玻璃膜疣下无法可靠测量毛细血管层血流缺损,需识别并排除。
- 软性玻璃膜疣下可用既定补偿策略测量毛细血管层血流缺损。
核心结论
- 钙化性玻璃膜疣的高散射特性导致其下出现散射尾,产生可量化的光学相干断层扫描血管成像信号错觉,但该信号不含评估毛细血管层血流缺损所需的血管成像信息,因此无法可靠测量钙化性玻璃膜疣下毛细血管层血流缺损,必须识别并排除钙化性玻璃膜疣。
研究设计与样本
- 研究纳入正常眼(n=30)、软性玻璃膜疣的年龄相关性黄斑变性眼(n=30)和钙化性玻璃膜疣的年龄相关性黄斑变性眼(n=30),测量毛细血管层血流缺损,并生成基于结构en face光学相干断层扫描图像和对应B扫描的软性玻璃膜疣与钙化性玻璃膜疣掩膜,计算Dice相似系数,另进行体模实验模拟钙化性玻璃膜疣的光散射影响。
主要发现
- 毛细血管层血流缺损面积测量与钙化性玻璃膜疣面积高度相关,但与软性玻璃膜疣不相关。
- 钙化性玻璃膜疣下检测到的光学相干断层扫描信号并非来自定义的毛细血管层,而是钙化性玻璃膜疣多重散射特性所致的伪影。
- 体模实验显示,类似钙化性玻璃膜疣内容物的高散射材料产生伪影散射尾,在毛细血管层结构层虚假生成信号,但无法检测其下血流。
- 钙化性玻璃膜疣产生伪影散射尾并阻止光线穿透至脉络膜,导致en face低透射缺陷和无法检测毛细血管层内血流。
- 钙化性玻璃膜疣消退后,毛细血管层灌注变为可检测。
临床怎么用
- 作者认为,毛细血管层血流缺损无法在钙化性玻璃膜疣下可靠测量,必须识别钙化性玻璃膜疣并将其从黄斑毛细血管层血流缺损测量中排除。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: In age-related macular degeneration (AMD), choriocapillaris flow deficits (CCFDs) under soft drusen can be measured using established compensation strategies. This study investigated whether CCFDs can be quantified under calcified drusen (CaD). METHODS: CCFDs were measured in normal eyes (n = 30) and AMD eyes with soft drusen (n = 30) or CaD (n = 30). CCFD density masks were generated to highlight regions with higher CCFDs. Masks were also generated for soft drusen and CaD based on both structural en face OCT images and corresponding B-scans. Dice similarity coefficients were calculated between the CCFD density masks and both the soft drusen and CaD masks. A phantom experiment was conducted to simulate the impact of light scattering that arises from CaD. RESULTS: Area measurements of CCFDs were highly correlated with those of CaD but not soft drusen, suggesting an association between CaD and underlying CCFDs. However, unlike soft drusen, the detected optical coherence tomography (OCT) signals underlying CaD did not arise from the defined CC layer but were artifacts caused by the multiple scattering property of CaD. Phantom experiments showed that the presence of highly scattering material similar to the contents of CaD caused an artifactual scattering tail that falsely generated a signal in the CC structural layer but the underlying flow could not be detected. Similarly, CaD also caused an artifactual scattering tail and prevented the penetration of light into the choroid, resulting in en face hypotransmission defects and an inability to detect blood flow within the choriocapillaris. Upon resolution of the CaD, the CC perfusion became detectable. CONCLUSIONS: The high scattering property of CaD leads to a scattering tail under these drusen that gives the illusion of a quantifiable optical coherence tomography angiography signal, but this signal does not contain the angiographic information required to assess CCFDs. For this reason, CCFDs cannot be reliably measured under CaD, and CaD must be identified and excluded from macular CCFD measurements.