文献速递角膜与眼表疾病

一种针对中国人群优化的新型Corvis生物力学指数用于圆锥角膜检测

Detection of Keratoconus With a New Corvis Biomechanical Index Optimized for Chinese Populations.
PubMed 文献 · Am J Ophthalmol · 2023-04-13

在外部验证集中,cCBI的曲线下面积为0.985,敏感度95.5%、特异度93.4%,优于原CBI(曲线下面积0.978,敏感度97.7%、特异度68.1%),受试者工作特征曲线差异有统计学意义(De Long P = .0009)。

要点

  • cCBI在验证集中曲线下面积0.985,特异度93.4%,敏感度95.5%。
  • 原CBI在同数据集曲线下面积0.978,特异度68.1%,敏感度97.7%。
  • cCBI与原CBI的受试者工作特征曲线差异有统计学意义(De Long P = .0009)。

核心结论

  • 针对中国患者的cCBI在区分健康眼与圆锥角膜眼方面较CBI有统计学显著更优的表现。
  • 外部验证数据集的存在确认了这一发现,并提示可在日常临床实践中使用cCBI辅助中国族群患者的圆锥角膜诊断。

研究设计与样本

  • 回顾性、多中心临床效度增强研究,患者来自中国北京、沈阳、广州、上海、温州、重庆和天津的7家诊所。
  • 以数据库1(7家诊所中的6家)为开发数据集,采用逻辑回归优化CBI常数,构建新指数cCBI。
  • CBI的因子(A1Velocity、ARTh、Stiffness Parameter-A、DARatio2mm和Inverse Integrated Radius)及截断值保持不变(0.5)。
  • cCBI确定后在数据库2(7家诊所中的1家)上进行验证。

主要发现

  • 在数据库2中,cCBI的曲线下面积为0.985,特异度93.4%,敏感度95.5%。
  • 在同一数据集中,原CBI的曲线下面积为0.978,特异度68.1%,敏感度97.7%。
  • cCBI与原CBI的受试者工作特征曲线差异有统计学意义(De Long P = .0009)。

临床怎么用

  • 作者建议在日常临床实践中使用cCBI辅助中国族群患者的圆锥角膜诊断。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: The aim of this study was to introduce an optimized version of the Corvis Biomechanical Index for Chinese populations (cCBI). DESIGN: Retrospective, multicenter clinical validity enhancement study. METHODS: Patients were included from 7 clinics in Beijing, Shenyang, Guangzhou, Shanghai, Wenzhou, Chongqing, and Tianjin, China. Logistic regression was used to optimize the values of the constants of the CBI, based on database 1 as the development dataset (6 of 7 clinics), to create a new version of the index named cCBI. The factors of the CBI (A1Velocity, ARTh, Stiffness Parameter-A, DARatio2mm, and Inverse Integrated Radius) and the cutoff value were kept the same (0.5). With the formation of cCBI determined, it was validated on database 2 (1 of the 7 clinics). RESULTS: Two thousand four hundred seventy-three patients (healthy and keratoconus) were included. In database 2, the area under the curve of the cCBI was 0.985 with 93.4% specificity and 95.5% sensitivity. In the same dataset, the original CBI produced an area under the curve of 0.978 with 68.1% specificity and 97.7% sensitivity. There was a statistically significant difference between the receiver operating characteristic curve of cCBI and CBI (De Long P = .0009) CONCLUSION: The new cCBI for Chinese patients was shown to be statistically significantly better when compared with CBI to separate healthy from keratoconic eyes. The presence of an external validation dataset confirms this finding and suggests the use of cCBI in everyday clinical practice to aid in the diagnosis of keratoconus in patients who are of Chinese ethnicity.

来源:Am J Ophthalmol,查看原文。版权归原作者所有。

引用

  1. Vinciguerra R, Ambrosio R, Wang Y, et al. Detection of Keratoconus With a New Corvis Biomechanical Index Optimized for Chinese Populations. Am J Ophthalmol. 2023. PMID: 37059320. doi:10.1016/j.ajo.2023.04.002

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。