新型疏水性丙烯酸酯人工晶状体平台的旋转稳定性、偏心与倾斜
该前瞻性干预性病例系列在维也纳医科大学眼科纳入130眼(68例)年龄相关性白内障患者,评估Nanex(NC1-SP)人工晶状体植入后的旋转稳定性、偏心与倾斜。结果显示中位旋转1.9°,10眼(9.71%)旋转超过5°,旋转与前囊纤维化严重程度正相关,术前术后偏心与倾斜无显著变化。
要点
- 从手术结束到术后6个月,所有眼的中位人工晶状体旋转为1.9°。
- 10眼(9.71%)旋转超过5°,2眼(1.94%)旋转超过10°。
- 人工晶状体旋转与前囊纤维化严重程度正相关(Spearman r=0.321,P=.002)。
核心结论
- 所研究的人工晶状体表现出良好的旋转稳定性和较低的偏心与倾斜值。
- 前囊纤维化发展导致术后1周后人工晶状体旋转倾向更高。
研究设计与样本
- 前瞻性干预性病例系列,在维也纳医科大学眼科开展,纳入130眼(68例)年龄相关性白内障患者,行Nanex(NC1-SP)人工晶状体植入。
主要发现
- 所有眼从手术结束到术后6个月的中位人工晶状体旋转为1.9°(四分位距0.1°-37.5°)。
- 10眼(9.71%)旋转超过5°,2眼(1.94%)旋转超过10°。
- 人工晶状体旋转与眼轴长度无相关(Spearman r=-0.042,P=.46),与晶状体厚度无相关(Spearman r=0.134,P=.19),与晶状体赤道直径无相关(Spearman r=0.101,P=.325)。
- 人工晶状体旋转与前囊纤维化严重程度正相关(Spearman r=0.321,P=.002)。
- 术前偏心(0.2±0.12 mm)与倾斜(5.7±1.6°)术后无显著变化(术后分别为0.22±0.12 mm和5.62±1.49°)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To evaluate rotational stability, decentration and tilt of the monofocal intraocular lens (IOL) Nanex (NC1-SP; HOYA Surgical Optics). DESIGN: Prospective interventional case series. METHODS: The study was performed at the Department of Ophthalmology, Medical University of Vienna. The study population comprised 130 eyes of 68 patients with age-related cataract who underwent cataract surgery with implantation of a Nanex IOL. Baseline image for rotational stability evaluation was obtained at the end of surgery (EoS) and compared to retroillumination images taken at 1 hour, 1 week, 1 month, and 6 months after surgery. Axis alignment was assessed using nonmovable landmarks on the sclera and the optic-haptic junctions of the IOL. Anterior segment OCT images were performed to evaluate decentration and tilt. The main outcome measure was absolute rotation from EoS to 6 months postoperatively. RESULTS: Median IOL rotation of all eyes from EoS to 6 months was 1.9° (interquartile range 0.1°-37.5°). Ten eyes (9.71%) rotated more than 5° and 2 eyes (1.94%), more than 10°. IOL rotation did not correlate with axial eye length (Spearman r = -0.042, P = .46), crystalline lens thickness (Spearman r = 0.134, P = .19), and crystalline lens equatorial diameter (Spearman r = 0.101, P = .325). IOL rotation positively correlated with anterior fibrosis severity (Spearman r = 0.321, P = .002). Preoperative decentration (0.2 ± 0.12 mm) and tilt (5.7 ± 1.6°) did not change significantly after surgery (0.22 ± 0.12 mm and 5.62 ± 1.49°, respectively). CONCLUSION: The investigated IOL presented good rotational stability and low decentration and tilt values. Nevertheless, anterior capsule fibrosis development led to a higher tendency of IOL rotation after 1 week.
引用
- Röggla V, Schartmüller D, Schwarzenbacher L, et al. Rotational Stability, Decentration, and Tilt of a New Hydrophobic Acrylic Intraocular Lens Platform. Am J Ophthalmol. 2023. PMID: 36754132. doi:10.1016/j.ajo.2023.01.026
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。