使用家用设备自我监测新生血管性年龄相关性黄斑变性的自动化光学相干断层扫描图像分析的定量评估

QUANTITATIVE ASSESSMENT OF AUTOMATED OPTICAL COHERENCE TOMOGRAPHY IMAGE ANALYSIS USING A HOME-BASED DEVICE FOR SELF-MONITORING NEOVASCULAR AGE-RELATED MACULAR DEGENERATION.
PubMed 文献 · Retina · 2022-11-18

这项前瞻性观察研究纳入了接受抗血管内皮生长因子治疗的新生血管性年龄相关性黄斑变性患者,使用原型家用光学相干断层扫描设备采集70例患者的136次扫描,用自动化分析软件分割视网膜液体并与人工分级比较。结果显示各类型液体检测的受试者工作特征曲线下面积在每次B扫描为0.95至0.98、按体积为0.997至0.998,作者认为自动化分析与人工分级一致性极好。

要点

  • 每次B扫描检测视网膜内液、视网膜下液和视网膜色素上皮下液的曲线下面积分别为0.95、0.97和0.98。
  • 按体积计算的曲线下面积分别为0.997、0.998和0.998。
  • 平均Dice相似系数分别为0.64、0.73和0.74,决定系数分别为0.81、0.93和0.97。

核心结论

  • 家用光学相干断层扫描设备图像经自动化分析软件评估后与人工分级显示极好的一致性。

研究设计与样本

  • 前瞻性观察研究,纳入接受抗血管内皮生长因子治疗的患者,使用原型家用光学相干断层扫描设备采集70例患者的136次扫描。

主要发现

  • 每次B扫描检测视网膜内液、视网膜下液和视网膜色素上皮下液的受试者工作特征曲线下面积分别为0.95、0.97和0.98。
  • 按体积计算,视网膜内液、视网膜下液和视网膜色素上皮下液的曲线下面积分别为0.997、0.998和0.998。
  • 所有B扫描的平均Dice相似系数对视网膜内液、视网膜下液和视网膜色素上皮下液分别为0.64、0.73和0.74。
  • 决定系数对视网膜内液、视网膜下液和视网膜色素上皮下液分别为0.81、0.93和0.97。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To evaluate a prototype home optical coherence tomography device and automated analysis software for detection and quantification of retinal fluid relative to manual human grading in a cohort of patients with neovascular age-related macular degeneration. METHODS: Patients undergoing anti-vascular endothelial growth factor therapy were enrolled in this prospective observational study. In 136 optical coherence tomography scans from 70 patients using the prototype home optical coherence tomography device, fluid segmentation was performed using automated analysis software and compared with manual gradings across all retinal fluid types using receiver-operating characteristic curves. The Dice similarity coefficient was used to assess the accuracy of segmentations, and correlation of fluid areas quantified end point agreement. RESULTS: Fluid detection per B-scan had area under the receiver-operating characteristic curves of 0.95, 0.97, and 0.98 for intraretinal fluid, subretinal fluid, and subretinal pigment epithelium fluid, respectively. On a per volume basis, the values for intraretinal fluid, subretinal fluid, and subretinal pigment epithelium fluid were 0.997, 0.998, and 0.998, respectively. The average Dice similarity coefficient values across all B-scans were 0.64, 0.73, and 0.74, and the coefficients of determination were 0.81, 0.93, and 0.97 for intraretinal fluid, subretinal fluid, and subretinal pigment epithelium fluid, respectively. CONCLUSION: Home optical coherence tomography device images assessed using the automated analysis software showed excellent agreement to manual human grading.