文献速递角膜与眼表疾病

优化人工智能整合Scheimpflug角膜断层扫描与生物力学数据以增强扩张检测

Optimized Artificial Intelligence for Enhanced Ectasia Detection Using Scheimpflug-Based Corneal Tomography and Biomechanical Data.
PubMed 文献 · Am J Ophthalmol · 2022-12-19

多中心横断面病例对照回顾性研究纳入3886只未手术眼,评估TBIv1并开发优化AI算法TBIv2。TBIv2对VAE-NT检测的AUC为0.945,显著高于TBIv1的0.899;对临床扩张AUC均为0.999。AI整合断层扫描与生物力学可提高扩张检测准确性。

要点

  • TBIv2对VAE-NT的AUC为0.945,显著高于TBIv1的0.899(DeLong P<.0001)。
  • TBIv2对临床扩张的AUC为0.999,与TBIv1相似(DeLong P=.818)。
  • 部分VAE患者可能为真性单侧扩张。

核心结论

  • AI优化整合Scheimpflug角膜断层扫描与生物力学评估可提高扩张检测准确性,表征VAE-NT组的扩张易感性。
  • 部分VAE患者可能为真性单侧扩张。

研究设计与样本

  • 多中心横断面病例对照回顾性研究,纳入3412例患者的3886只未手术眼,行Pentacam和Corvis ST检查。

主要发现

  • TBIv1区分临床扩张(KC和VAE-E)的AUC为0.999(敏感性98.5%,特异性98.6%,切点0.5),区分VAE-NT的AUC为0.899(敏感性76%,特异性89.1%,切点0.29)。
  • TBIv2检测VAE-NT的AUC为0.945(敏感性84.4%,特异性90.1%,切点0.43;DeLong P<.0001),临床扩张AUC为0.999(敏感性98.7%,特异性99.2%,切点0.8;DeLong P=.818)。
  • 考虑所有病例,TBIv2的AUC为0.985,高于TBIv1的0.974(DeLong P<.0001)。

关键数值与推荐

  • TBIv2采用随机森林算法,使用18个特征、156棵树、10折交叉验证开发。
  • TBIv1对临床扩张切点0.5,对VAE-NT切点0.29;TBIv2对VAE-NT切点0.43,对临床扩张切点0.8。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To optimize artificial intelligence (AI) algorithms to integrate Scheimpflug-based corneal tomography and biomechanics to enhance ectasia detection. DESIGN: Multicenter cross-sectional case-control retrospective study. METHODS: A total of 3886 unoperated eyes from 3412 patients had Pentacam and Corvis ST (Oculus Optikgeräte GmbH) examinations. The database included 1 eye randomly selected from 1680 normal patients (N) and from 1181 "bilateral" keratoconus (KC) patients, along with 551 normal topography eyes from patients with very asymmetric ectasia (VAE-NT), and their 474 unoperated ectatic (VAE-E) eyes. The current TBIv1 (tomographic-biomechanical index) was tested, and an optimized AI algorithm was developed for augmenting accuracy. RESULTS: The area under the receiver operating characteristic curve (AUC) of the TBIv1 for discriminating clinical ectasia (KC and VAE-E) was 0.999 (98.5% sensitivity; 98.6% specificity [cutoff: 0.5]), and for VAE-NT, 0.899 (76% sensitivity; 89.1% specificity [cutoff: 0.29]). A novel random forest algorithm (TBIv2), developed with 18 features in 156 trees using 10-fold cross-validation, had a significantly higher AUC (0.945; DeLong, P < .0001) for detecting VAE-NT (84.4% sensitivity and 90.1% specificity; cutoff: 0.43; DeLong, P < .0001) and a similar AUC for clinical ectasia (0.999; DeLong, P = .818; 98.7% sensitivity; 99.2% specificity [cutoff: 0.8]). Considering all cases, the TBIv2 had a higher AUC (0.985) than TBIv1 (0.974; DeLong, P < .0001). CONCLUSIONS: AI optimization to integrate Scheimpflug-based corneal tomography and biomechanical assessments augments accuracy for ectasia detection, characterizing ectasia susceptibility in the diverse VAE-NT group. Some patients with VAE may have true unilateral ectasia. Machine learning considering additional data, including epithelial thickness or other parameters from multimodal refractive imaging, will continuously enhance accuracy. NOTE: Publication of this article is sponsored by the American Ophthalmological Society.

来源:Am J Ophthalmol,查看原文。版权归原作者所有。

引用

  1. Ambrósio R, Machado AP, Leão E, et al. Optimized Artificial Intelligence for Enhanced Ectasia Detection Using Scheimpflug-Based Corneal Tomography and Biomechanical Data. Am J Ophthalmol. 2022. PMID: 36549584. doi:10.1016/j.ajo.2022.12.016

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。