四种非球面人工晶状体植入术后偏心、倾斜及视觉质量的对比研究
前瞻性非随机对照研究纳入169例(224只眼)年龄相关性白内障患者,按植入的ZCB00、SN60WF、A1-UV、AO四种非球面人工晶状体分组。术后1天、1周及1个月SN60WF组偏心值大于其他三组,各组倾斜度无显著差异,AO组球面像差高于其他三组,OSI、MTF cut off和SR组间无显著差异。
要点
- SN60WF组术后1天及1周偏心值大于另外三组,术后1个月偏心值(0.259±0.101)mm大于ZCB00组(0.177±0.099)mm和AO组(0.163±0.122)mm。
- 术后各时间点四组间IOL倾斜度差异均无统计学意义,各组内术后不同时间偏心值与倾斜度差异均无统计学意义。
- 4 mm和6 mm瞳孔直径下AO组眼内及全眼球面像差均大于另外三组,四组间OSI、MTF cut off和SR差异均无统计学意义。
核心结论
- 四种非球面IOL植入晶状体囊袋后均有不同程度的偏心与倾斜,但差别未显示出临床意义。
- 术眼对轻度IOL偏心与倾斜有良好的耐受性,视觉质量良好。
研究设计与样本
- 前瞻性非随机对照研究,连续纳入2020年4至6月在川北医学院附属医院眼科拟行超声乳化白内障吸除IOL植入术的年龄相关性白内障患者。
- 按植入IOL类型分为ZCB00组、SN60WF组、A1-UV组和AO组,每组右眼与左眼各30只。
- 共招募181例(240只眼),169例(224只眼)完成随访(右眼114只,左眼110只),男性77例,女性92例,年龄(69±9)岁。
- 于术前及术后1天、1周和1个月用CASIA2扫描仪测量偏心值与倾斜度,并检查波前像差、OSI、MTF cut off和SR;双眼患者均取右眼数据进行组间比较。
主要发现
- 四组间性别分布、年龄、眼轴长度、术前晶状体偏心值和倾斜度及IOL度数差异均无统计学意义(均P>0.05)。
- 术后1天、1周及1个月四组间IOL偏心值差异均有统计学意义(F=7.11、6.12、4.66,均P<0.05);SN60WF组术后1天及1周偏心值均大于另外三组。
- 术后1个月SN60WF组偏心值(0.259±0.101)mm大于ZCB00组(0.177±0.099)mm和AO组(0.163±0.122)mm,差异均有统计学意义(SNK-q检验,均P<0.05)。
- 术后1个月ZCB00组、SN60WF组、A1-UV组和AO组IOL倾斜度分别为(4.806±1.129)°、(5.080±1.309)°、(4.586±1.338)°、(5.112±1.406)°,术后各时间点四组间倾斜度差异均无统计学意义(均P>0.05)。
- 术后1个月4 mm和6 mm瞳孔直径下四组间眼内球面像差(F=131.32、85.17)及全眼球面像差(F=46.64、47.55)差异均有统计学意义(均P<0.01),AO组均大于另外三组(SNK-q检验,均P<0.05)。
关键数值与推荐
- 各组内术后不同时间IOL偏心值及倾斜度差异均无统计学意义(均P>0.05)。
- 术后1个月四组间IOL偏心值与倾斜度在水平方向及垂直方向上差异均无统计学意义(均P>0.05),各组内右眼与左眼偏心值及倾斜度差异均无统计学意义(均P>0.05),且均有向鼻下方或向颞下方倾斜的趋势。
- 四组间OSI、MTF cut off和SR差异均无统计学意义(均P>0.05)。
临床怎么用
- 原文结论未据此给出具体临床推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Objective: To compare the difference of decentration and tilt among 4 kinds of aspherical intraocular lenses (IOLs), and to analyze their objective visual quality. Methods: Prospective non-randomized controlled clinical trial. Age-related cataract patients who planned to undergo phacoemulsification and IOL implantation in the Affiliated Hospital of North Sichuan Medical College from April to June in 2020 were divided into ZCB00 group, SN60WF group, A1-UV group, and AO group according to IOL types. Thirty right eyes and thirty left eyes were selected in each group. Before operation and at 1 day, 1 week and 1 month postoperatively, decentration and tilt values were measured by a new swept-source anterior-segment optical coherence tomography device (CASIA2), and wavefront aberrations, objective scatter index (OSI), modulation transfer function cut off frequency (MTF cut off) and Strehl ratio (SR) were also examined. Values of decentration, tilt and visual quality compared among 4 groups were took from the right eye. One-way analysis of variance was used for inter-group comparison, and repeated measurement one-way analysis of variance was used for intra-group comparison. Data between right eyes and left eyes from all the individuals were compared by independent sample t-test. Results: A total of 181 patients (240 eyes) were enrolled, and 169 patients (224 eyes) completed the follow-up (114 right eyes and 110 left eyes). There were 77 males and 92 females, with an age of (69±9) years. There was no significant difference in gender, age, axial length, decentration and tilt of crystalline lens and IOL power among 4 groups (all P>0.05). At 1 day, 1 week and 1 month postoperatively, there was significant difference in decentration value among 4 groups (F=7.11, 6.12, 4.66; all P<0.05). For further pairwise comparison, the decentration value of SN60WF group was higher than that of the other 3 groups at 1 day and 1 week postoperatively, and the decentration value of SN60WF group was (0.259±0.101) mm at 1 month postoperatively, which was higher than that of ZCB00 group (0.177±0.099) mm and AO group (0.163±0.122) mm, and the differences were statistically significant (using SNK-q test, both P<0.05). The IOL tilt value in the ZCB00 group, SN60WF group, A1-UV group, and AO group at 1 month postoperatively were (4.806±1.129)°, (5.080±1.309)°, (4.586±1.338)°, (5.112±1.406)°, respectively. No significant difference in tilt value among 4 groups was found at any time after surgery (all P>0.05). In each group, there was no significant difference in decentration and tilt value at different postoperative time points (all P>0.05). At 1 month postoperatively, there was no significant difference in decentration and tilt at horizontal and vertical directions respectively among 4 groups (all P>0.05), and in each group, there was no significant difference in decentration and tilt value between right eyes and left eyes (all P>0.05), and IOLs tended to tilt towards the inferonasal or inferotemporal direction in both eyes. With 4-mm and 6-mm pupil diameter, there was significant difference in internal (F=131.32, 85.17) and ocular (F=46.64, 47.55) spherical aberration among 4 groups (all P<0.01). For further pairwise comparison, the spherical aberration of AO group was higher than that of the other 3 groups, and the difference was statistically significant (using SNK-q test, all P<0.05). There was no significant difference in OSI, MTF cut off and SR among 4 groups (all P>0.05). Conclusions: The four types of IOLs show decentration and tilt in varying degrees after implantation in the capsular bag, but this difference do not lead to clinical significance. Human eyes have tolerance to mild decentration and tilt of aspheric IOLs, showing good visual quality. 目的: 比较4种非球面人工晶状体(IOL)植入术后IOL偏心和倾斜的差别及术眼的客观视觉质量。 方法: 前瞻性非随机对照研究。连续纳入2020年4至6月在川北医学院附属医院眼科拟行超声乳化白内障吸除IOL植入术的年龄相关性白内障患者,按植入的IOL类型(ZCB00、SN60WF、A1-UV、AO)分为4个组,每组右眼与左眼各30只。于术前及术后1 d、1周和1个月测量晶状体或IOL的偏心值与倾斜度(使用CASIA2扫描仪)及术眼波前像差、客观散射指数(OSI)、传递调制函数截止频率(MTF cut off)和斯特列尔比值(SR)。双眼患者均取右眼数据,组间比较采用单因素方差分析,组内比较采用单因素重复测量方差分析。所有纳入眼的眼别间数据比较采用独立样本t检验。 结果: 共招募181例(240只眼)患者,其中169例(224只眼)完成随访(右眼114只,左眼110只),男性77例,女性92例,年龄(69±9)岁。4个组间性别分布、年龄、眼轴长度、术前晶状体偏心值和倾斜度以及IOL度数差异均无统计学意义(均P>0.05)。术后1 d、1周及1个月,4个组间IOL偏心值差异均有统计学意义(F=7.11、6.12、4.66,均P<0.05);进一步两两比较,SN60WF组术后1 d及1周的IOL偏心值均大于另外3个组,术后1个月的偏心值为(0.259±0.101)mm,大于ZCB00组的(0.177±0.099)mm和AO组的(0.163±0.122)mm,差异均有统计学意义(SNK-q检验,均P<0.05)。术后1个月ZCB00组、SN60WF组、A1-UV组和AO组的IOL倾斜度分别为(4.806±1.129)°、(5.080±1.309)°、(4.586±1.338)°、(5.112±1.406)°,术后各时间点4个组间IOL倾斜度差异均无统计学意义(均P>0.05)。各组内术后不同时间的IOL偏心值及倾斜度差异均无统计学意义(均P>0.05)。术后1个月,4个组间IOL偏心值与倾斜度在水平方向及垂直方向上差异均无统计学意义(均P>0.05),各组内右眼与左眼的IOL偏心值及倾斜度差异均无统计学意义(均P>0.05),且均有向鼻下方或向颞下方倾斜的趋势。术后1个月,在4 mm和6 mm瞳孔直径下,4个组间眼内球面像差(F=131.32、85.17)及全眼球面像差(F=46.64、47.55)的差异均有统计学意义(均P<0.01);进一步两两比较,AO组均大于另外3个组,差异均有统计学意义(SNK-q检验,均P<0.05)。4个组间OSI、MTF cut off和SR差异均无统计学意义(均P>0.05)。 结论: 4种非球面IOL植入晶状体囊袋后均有不同程度的偏心与倾斜,但差别未显示出临床意义;术眼对轻度IOL偏心与倾斜有良好的耐受性,视觉质量良好。.