非接触式眼压测量期间泪膜微气溶胶的无创可视化
本研究采用前瞻性实验与健康眼研究,用无创实时观察装置探究商用非接触式眼压计测量时气溶胶的产生。实验模型眼和20名健康志愿者均证实测量过程中有气溶胶生成,其数量受眼压、吹气压力和测量前是否滴眼药水影响。
要点
- 非接触式眼压计测量时确实从泪膜产生气溶胶。
- 在实验研究中,较低眼压(5 mm Hg)比较高压(25 mm Hg)产生更多气溶胶。
- 吹气压力与气溶胶颗粒数呈显著正相关,滴眼药水可增加气溶胶生成。
核心结论
- 非接触式眼压计测量会从泪膜产生明显气溶胶,其数量由眼压、吹气压力及测量前是否使用眼药水决定。
研究设计与样本
- 前瞻性实验研究与健康眼研究;实验部分使用模型假眼,人类研究纳入20名健康志愿者对照受试者。
主要发现
- 实验与人类研究均证实非接触式眼压计测量过程中有气溶胶生成。
- 实验研究中,吹气压力设为20 mm Hg和40 mm Hg时,低眼压(5 mm Hg)比高眼压(25 mm Hg)产生显著更多气溶胶(20 mm Hg时P=0.0159;40 mm Hg时P=0.0079)。
- 在高眼压眼和低眼压眼中,吹气压力与气溶胶颗粒数均呈显著正相关(P<0.001)。
- 吹气压力为40 mm Hg时,使用眼药水比不使用眼药水产生的气溶胶量显著更高(P=0.047)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To investigate the aerosol generation by a noninvasive real-time observation device and assess the conditions relating to aerosolization during intraocular pressure (IOP) measurements using a commercial noncontact tonometer (NCT). STUDY DESIGN: Prospective experimental and healthy eye studies. METHODS: In an initial experimental study, we devised a model mannequin eye to investigate how air puff pressure and IOP of the eye affected aerosol generation. In the human study including 20 healthy volunteer control subjects, the number of tear aerosol particles generated at 20 and 40 mm Hg air puff pressures with and without eye drop was investigated. The recorded aerosol visualization video was analyzed and the number of aerosol particles generated in 5 seconds after IOP measurement was measured. RESULTS: The experimental and human studies confirmed the aerosol generation during NCT measurements. In the experimental study, when the air puff pressures were set at 20 and 40 mm Hg, a lower IOP (5 mm Hg) generated significantly more aerosols than a higher IOP (25 mm Hg) (20 mm Hg, P = .0159; 40 mm Hg, P = .0079). There was also a significant positive correlation between the air puff pressure and the number of aerosol particles in both high- and low-IOP eyes (P < .001). At an air puff pressure of 40 mm Hg, the amount of aerosol generated was significantly higher with eye drop than without eye drop (P = .047). CONCLUSIONS: NCT generates significant aerosolization from the tear film, the amount of which is determined by the IOP and the air puff pressure and the presence of eye drop use before the measurements.
引用
- Kojima T, Dogru M, Yazu H, et al. Noninvasive Visualization of the Tear Film Microaerosol During Noncontact Tonometry Measurements. Am J Ophthalmol. 2022. PMID: 35469788. doi:10.1016/j.ajo.2022.03.031
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。