文献速递青光眼

基于深度学习的前节光学相干断层扫描图像中巩膜突定位及前房角测量的可重复性

Reproducibility of deep learning based scleral spur localisation and anterior chamber angle measurements from anterior segment optical coherence tomography images.
PubMed 文献 · Br J Ophthalmol · 2022-01-28

这项多中心横断面研究纳入287只眼(116只健康开放角眼和171只原发性房角关闭疾病眼)的5166幅前节OCT图像,应用深度学习模型自动定位巩膜突并与手动定位比较前房角宽度可重复性。深度学习定位与手动定位巩膜突的平均欧几里得距离在正常组为66.50±20.54 µm、在原发性房角关闭疾病组为84.78±28.33 µm;AOD750和TISA750的可重复性系数在深度学习定位下分别为0.049 mm和0.026 mm²,在手动定位下分别为0.058 mm和0.030 mm²。结论认为深度学习定位巩膜突测量前房角可获得与手动定位相当的可重复性。

要点

  • 深度学习自动定位巩膜突与手动定位的平均欧几里得距离在正常组为66.50±20.54 µm、在原发性房角关闭疾病组为84.78±28.33 µm。
  • AOD750和TISA750的可重复性系数在深度学习定位下分别为0.049 mm和0.026 mm²,在手动定位下分别为0.058 mm和0.030 mm²。
  • 深度学习定位巩膜突测量前房角可获得与手动定位相当的可重复性。

核心结论

  • 深度学习定位巩膜突测量前房角可获得与手动定位相当的可重复性。

研究设计与样本

  • 多中心横断面研究,测试数据集包含来自4家眼科诊所的287名受试者287只眼的5166幅前节OCT图像,其中健康开放角眼116只、原发性房角关闭疾病眼171只。
  • 每只眼在同一就诊日接受两次扫频源前节OCT(CASIA2,Tomey,日本名古屋)成像,每名患者随机选取一只眼进行前房角测量。
  • 采用欧几里得距离评估深度学习定位巩膜突与手动定位巩膜突之间的一致性,使用CASIA2内置软件计算750 µm的房角开放距离(AOD750)和750 µm的小梁虹膜间隙面积(TISA750),并测量前房角宽度的可重复性。

主要发现

  • 正常组平均年龄为60.8±12.3岁(范围30-85岁),原发性房角关闭疾病组平均年龄为63.4±10.6岁(范围40-91岁)。
  • 深度学习定位与手动定位巩膜突的平均欧几里得距离在正常组为66.50±20.54 µm,在原发性房角关闭疾病组为84.78±28.33 µm。
  • 深度学习定位与手动定位之间AOD750的95%一致性界限跨度为0.064 mm,TISA750的95%一致性界限跨度为0.034 mm²。
  • AOD750和TISA750的可重复性系数在深度学习定位下分别为0.049 mm和0.026 mm²,在手动定位下分别为0.058 mm和0.030 mm²。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIMS: To apply a deep learning model for automatic localisation of the scleral spur (SS) in anterior segment optical coherence tomography (AS-OCT) images and compare the reproducibility of anterior chamber angle (ACA) width between deep learning located SS (DLLSS) and manually plotted SS (MPSS). METHODS: In this multicentre, cross-sectional study, a test dataset comprising 5166 AS-OCT images from 287 eyes (116 healthy eyes with open angles and 171 eyes with primary angle-closure disease (PACD)) of 287 subjects were recruited from four ophthalmology clinics. Each eye was imaged twice by a swept-source AS-OCT (CASIA2, Tomey, Nagoya, Japan) in the same visit and one eye of each patient was randomly selected for measurements of ACA. The agreement between DLLSS and MPSS was assessed using the Euclidean distance (ED). The angle opening distance (AOD) of 750 µm (AOD750) and trabecular-iris space area (TISA) of 750 µm (TISA750) were calculated using the CASIA2 embedded software. The repeatability of ACA width was measured. RESULTS: The mean age was 60.8±12.3 years (range: 30-85 years) for the normal group and 63.4±10.6 years (range: 40-91 years) for the PACD group. The mean difference in ED for SS localisation between DLLSS and MPSS was 66.50±20.54 µm and 84.78±28.33 µm for the normal group and the PACD group, respectively. The span of 95% limits of agreement between DLLSS and MPSS was 0.064 mm for AOD750 and 0.034 mm2 for TISA750. The respective repeatability coefficients of AOD750 and TISA750 were 0.049 mm and 0.026 mm2 for DLLSS, and 0.058 mm and 0.030 mm2 for MPSS. CONCLUSION: DLLSS achieved comparable repeatability compared with MPSS for measurement of ACA.

来源:Br J Ophthalmol,查看原文。版权归原作者所有。

引用

  1. Liu P, Higashita R, Guo PY, et al. Reproducibility of deep learning based scleral spur localisation and anterior chamber angle measurements from anterior segment optical coherence tomography images. Br J Ophthalmol. 2022. PMID: 35091438. doi:10.1136/bjophthalmol-2021-319798

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。