魏斯环是测量视盘周围视网膜神经纤维层厚度的重要混杂因素:短标题:魏斯环对RNFL厚度测量的影响
这项回顾性横断面研究将205只眼分为对照组(131只眼)和光学相干断层扫描眼底图像上可见魏斯环的WR组(74只眼),比较两组平均及分区视盘周围视网膜神经纤维层(pRNFL)厚度。WR组平均pRNFL厚度为94.6±10.8µm,低于对照组的97.2±6.7µm(P=.042);WR组下方分区厚度为112.1±23.2µm,显著薄于对照组的125.5±13.3µm(P<.001)。作者认为魏斯环会导致测量误差,可能是分析pRNFL变化尤其是青光眼患者时的重要混杂因素。
要点
- WR组平均pRNFL厚度94.6±10.8µm,低于对照组97.2±6.7µm(P=.042)。
- WR组下方分区pRNFL厚度112.1±23.2µm,显著薄于对照组125.5±13.3µm(P<.001)。
- 魏斯环位于下方分区40只眼(54.1%),下方分区测量变异系数10.90%、组内相关系数0.409,可靠性低。
核心结论
- 有魏斯环的眼平均及下方pRNFL厚度较正常对照薄,这可能是魏斯环导致的测量误差。
- 魏斯环可能是分析pRNFL变化的重要混杂因素,尤其在青光眼患者中。
研究设计与样本
- 回顾性横断面研究,共纳入205只眼,对照组131只眼,光学相干断层扫描眼底图像上可见魏斯环的WR组74只眼。
主要发现
- 对照组与WR组平均pRNFL厚度分别为97.2±6.7µm和94.6±10.8µm(P=.042)。
- WR组下方分区pRNFL厚度为112.1±23.2µm,显著薄于对照组的125.5±13.3µm(P<.001)。
- 魏斯环位于上方分区10只眼(13.5%)、颞侧分区7只眼(9.5%)、下方分区40只眼(54.1%)、鼻侧分区17只眼(23.0%)。
- 下方分区测量的变异系数为10.90%,组内相关系数为0.409,提示测量可靠性低。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To identify how the Weiss ring affects the measurement of mean and sectoral peripapillary retinal nerve fiber layer (pRNFL) thicknesses. DESIGN: Retrospective, cross-sectional study. METHODS: Subjects were divided into two groups: controls (control group) and subjects in which a Weiss ring was visible on optical coherence tomography fundus images (WR group). Mean and sectoral pRNFL thicknesses were compared between the two groups. RESULTS: A total of 205 eyes were enrolled: 131 eyes in the control group and 74 eyes in the WR group. The mean pRNFL thicknesses of the control group and WR group were 97.2 ± 6.7 µm and 94.6 ± 10.8 µm, respectively (P = .042). In sectoral thickness, the inferior sector of the WR group was 112.1 ± 23.2 µm, which was significantly thinner than that of the control group (125.5 ± 13.3 µm; P < .001). The Weiss ring was located in 10 eyes (13.5%) in the superior sector, 7 eyes (9.5%) in the temporal sector, 40 eyes (54.1%) in the inferior sector, and 17 eyes (23.0%) in the nasal sector. In analyses of reproducibility, the coefficient of variation and intraclass coefficient of the inferior sector measurement were 10.90% and 0.409, respectively, indicating low reliability of the measurement. CONCLUSIONS: Eyes with a Weiss ring showed thinner mean and inferior pRNFL thicknesses than normal controls, which would be a measurement error caused by the Weiss ring. This could be a major confounding factor for analyses of pRNFL changes, especially in glaucoma patients.
引用
- Lee MW, Kim MS, Yu HY, et al. The Weiss Ring, a Major Confounding Factor For Measurements of Peripapillary Retinal Nerve Fiber Layer Thickness: Short title: The Impact of the Weiss Ring On RNFL Thickness Measurements. Am J Ophthalmol. 2022. PMID: 35038416. doi:10.1016/j.ajo.2022.01.001
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。