闭角型青光眼眼中测量眼部生物参数的最佳前节OCT图像数量与方向:中美眼研究
从中美眼研究中选取房角关闭眼,以每眼4条子午线图像计算平均房角开放距离(AOD500),并按AOD500四分位数每组随机选10眼行32图像详细分析,比较1、2、4、8或16张图像与32张图像的测量均值和范围。结果显示单张图像偏差较大,2张图像均值偏差降至0.08%–4.21%,8张图像除前房深度和前房面积外范围偏差降至<7.00%;沿25°-205°子午线放置首张图像在≤4张图像时更接近测量范围。
要点
- 单张AS-OCT图像测量偏差可达0.44%–19.31%,2张图像均值偏差降至0.08%–4.21%
- 8张图像时除前房深度和前房面积外,范围偏差降至<7.00%
- 分析≤4张图像时,首张图像沿25°-205°子午线放置更接近测量范围
核心结论
- 闭角型青光眼眼存在扇区解剖变异,现有AS-OCT成像惯例容易造成误判,修正的多图像方法能更好地反映生物测量值的均值和范围。
研究设计与样本
- 从中国美国人眼研究(Chinese American Eye Study)中选取静态房角镜下>3个象限不可见色素性小梁网的房角关闭受试者,每眼用4条子午线(0°-180°、45°-225°、90°-270°、135°-315°)图像计算平均AOD500。
- 按AOD500四分位数每组随机选10眼(共40眼,来自335只房角关闭眼)行32图像详细分析,涵盖AOD500、虹膜曲率(IC)、前房深度(ACD)、晶状体拱高(LV)和前房面积(ACA)等10项生物参数。
- 将1、2、4、8或16张图像的测量均值与范围同32张图像值进行比较。
主要发现
- 单张图像时各参数测量值与32张图像均值的偏差为0.44%–19.31%,2张图像时降至0.08%–4.21%。
- 单张图像时测量范围与32张图像范围的偏差为54.67%–88.94%,8张图像时除ACD和ACA外各参数降至<7.00%。
- 分析≤4张图像时,将首张分析图像沿25°-205°子午线放置能更好地近似测量范围。
关键数值与推荐
- 房角关闭定义为静态房角镜下>3个象限不可见色素性小梁网。
- 从335只房角关闭眼中选取40眼(AOD500四分位数每组10眼)行32图像分析。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To assess the optimal number and orientation of anterior segment optical coherence tomography (AS-OCT) images for accurately measuring ocular biometric parameters in angle closure eyes. METHODS: Subjects with angle closure, defined as >3 quadrants of non-visible pigmented trabecular meshwork on static gonioscopy, were selected from the Chinese American Eye Study. Mean angle opening distance (AOD500) was calculated using four images (0°-180°, 45°-225°, 90°-270° and 135°-315° meridians) from one eye per subject. Ten eyes from each quartile of AOD500 measurements were randomly selected for detailed 32-image analysis of 10 biometric parameters, including AOD500, iris curvature (IC), anterior chamber depth (ACD), lens vault (LV), and anterior chamber area (ACA). Mean and range of measurements from 1, 2, 4, 8 or 16 images were compared with 32-image values for all parameters. RESULTS: 40 out of 335 eyes with angle closure were selected for 32-image analysis. Deviation from the 32-image mean was between 0.44% and 19.31% with one image, decreasing to 0.08% to 4.21% with two images for all parameters. Deviation from the 32-image range of measurements was between 54.67% to 88.94% with one image, decreasing to <7.00% with eight images for all parameters except ACD and ACA. Orienting the first image analysed along the 25°-205° meridian better approximated the range of measurements when four or fewer images were analysed. CONCLUSIONS: Sectoral anatomical variations in angle closure eyes are easily misrepresented based on current AS-OCT imaging conventions. A revised multi-image approach can better capture the mean and range of biometric measurements.