美国与韩国队列中健康眼及青光眼伴或不伴轴性高度近视眼的Bruch膜开口检测准确性
横断面研究纳入美国与韩国队列的健康眼及青光眼眼,按眼轴长度(AL)分为非高度近视组(AL<26 mm)和高度近视组(AL≥26 mm),由专家评估Spectralis光学相干断层扫描(OCT)径向扫描自动定位Bruch膜开口(BMO)的准确性。结果显示高度近视眼BMO定位不准确的风险是非高度近视眼的2.4倍,无论诊断如何;作者建议对高度近视眼的OCT图像仔细审阅并在可能时校正BMO位置后再用于青光眼临床管理。
要点
- 健康眼中,伴与不伴高度近视者分别有9.1%与1.7%的BMO不可辨认,准确分割者分别占54.5%与87.6%。
- 青光眼眼中,伴与不伴高度近视者分别有15.0%与3.2%的BMO不可辨认,无需手动校正即准确分割者分别占30.0%与58.7%。
- 高度近视、较大AL、较大BMO倾斜角、较低BMO椭圆指数(更椭圆)及青光眼诊断是BMO定位不准确的显著预测因素。
核心结论
- 无论诊断如何,高度近视眼BMO定位不准确的可能性是非高度近视眼的2.4倍;作者建议仔细审阅高度近视眼的OCT图像,并在可能时于使用视神经乳头扫描结果进行青光眼临床管理前校正BMO位置。
研究设计与样本
- 横断面研究,纳入美国研究与韩国诊所人群的健康眼和青光眼眼,按AL分为非高度近视组(AL<26 mm)和高度近视组(AL≥26 mm),由专家审阅者评估Spectralis OCT径向扫描自动BMO定位的准确性。
主要发现
- 共纳入438只非高度近视眼(263名受试者)和113只高度近视眼(81名受试者)。
- 健康眼中,伴与不伴高度近视者分别有9.1%与1.7%的BMO不可辨认,准确分割者分别占54.5%与87.6%。
- 健康眼中,BMO不可辨认者分别有36.4%与10.7%可手动校正(P = .017)。
- 青光眼眼中,伴与不伴高度近视者分别有15.0%与3.2%的BMO不可辨认,手动校正者分别占55.0%与38.2%,无需手动校正即准确分割者分别占30.0%与58.7%(P = .005)。
- 多变量logistic回归分析中,高度近视、较大AL、较大BMO倾斜角、较低BMO椭圆指数(更椭圆)及青光眼诊断是BMO定位不准确的显著预测因素。
临床怎么用
- 作者建议,高度近视眼的OCT图像应仔细审阅,并在可能时于使用视神经乳头扫描结果进行青光眼临床管理前校正BMO位置。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To determine the predictors of Bruch membrane opening (BMO) location accuracy and the visibility of the BMO location in glaucoma and healthy individuals with and without axial high myopia. DESIGN: Cross-sectional study. METHODS: Healthy eyes and eyes with glaucoma from an American study and a Korean clinic population were classified into 2 groups: those with no axial high myopia (axial length [AL] <26 mm) and those with axial high myopia (AL ≥26 mm). The accuracy of the automated BMO location on optic nerve head Spectralis optical coherence tomography radial scans was assessed by expert reviewers. RESULTS: Four hundred thirty-eight non-highly myopic eyes (263 subjects) and 113 highly myopic eyes (81 subjects) were included. In healthy eyes with and without axial high myopia, 9.1% and 1.7% had indiscernible BMOs while 54.5% and 87.6% were accurately segmented, respectively. More than a third (36.4%) and 10.7% of eyes with indiscernible BMOs were manually correctable (respectively, P = .017). In eyes with glaucoma with and without high myopia, 15.0% and 3.2% had indiscernible BMOs, 55.0% and 38.2% were manually corrected, and 30.0% and 58.7% were accurately segmented without the need for manual correction (respectively, P = .005). Having axial high myopia, a larger AL, a larger BMO tilt angle, a lower BMO ovality index (more oval), and a glaucoma diagnosis were significant predictors of BMO location inaccuracy in multivariable logistic regression analysis. CONCLUSIONS: As BMO location inaccuracy was 2.4 times more likely in eyes with high axial myopia regardless of diagnosis, optical coherence tomography images of high myopes should be reviewed carefully, and when possible, BMO location should be corrected before using optic nerve head scan results for the clinical management of glaucoma.
引用
- Rezapour J, Proudfoot JA, Bowd C, et al. Bruch Membrane Opening Detection Accuracy in Healthy Eyes and Eyes With Glaucoma With and Without Axial High Myopia in an American and Korean Cohort. Am J Ophthalmol. 2021. PMID: 34902327. doi:10.1016/j.ajo.2021.11.030
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。