文献速递乳腺癌

锥光束乳腺CT与乳腺X线摄影在致密乳腺恶性病变诊断效能的比较

[Comparison of the diagnostic efficiency in breast malignancy between cone beam breast CT and mammography in dense breast].
PubMed 文献 · Zhonghua Zhong Liu Za Zhi · 2018-01-01

该研究纳入160例患者共165个乳腺,比较锥光束乳腺CT(CBBCT)与乳腺X线摄影对致密类乳腺病灶的诊断效能。结果显示,在全部乳腺及致密类乳腺中,CBBCT的AUC均高于乳腺X线摄影,以BI-RADS 4b为截断点时CBBCT灵敏度更高、特异度相近,提示CBBCT对致密类乳腺恶性肿瘤诊断有较高灵敏度。

要点

  • 在全部165个乳腺中,CBBCT的AUC为0.959,高于乳腺X线摄影的0.923。
  • 在141个致密类乳腺中,CBBCT的AUC为0.973,高于乳腺X线摄影的0.919。
  • 以BI-RADS 4b为截断点,CBBCT在致密类乳腺的灵敏度为83.1%,高于乳腺X线摄影的69.0%。

核心结论

  • 锥光束乳腺CT对乳腺恶性肿瘤诊断表现出较高灵敏度和特异度,有望为致密类乳腺恶性肿瘤诊断提供更多有效信息。

研究设计与样本

  • 研究纳入2012年5月至2014年8月同时接受乳腺X线摄影和锥光束乳腺CT检查的160例患者共165个乳腺,结合病理结果比较两种影像方法的诊断效能。

主要发现

  • 165个乳腺中非致密类乳腺24个、致密类乳腺141个;两种检查诊断结果一致者109个、不一致者56个。
  • 全部165个乳腺中,乳腺X线摄影AUC为0.923(95%CI:0.878~0.967,P<0.05),CBBCT AUC为0.959(95%CI:0.926~0.992,P<0.05)。
  • 全部165个乳腺以BI-RADS 4b为截断点,乳腺X线摄影灵敏度和特异度分别为70.0%和98.7%,CBBCT分别为83.3%和97.3%。
  • 141个致密类乳腺中,乳腺X线摄影AUC为0.919(95%CI:0.868~0.969,P<0.05),CBBCT AUC为0.973(95%CI:0.947~0.999,P<0.05)。
  • 141个致密类乳腺以BI-RADS 4b为截断点,乳腺X线摄影灵敏度和特异度分别为69.0%和98.6%,CBBCT分别为83.1%和98.6%。

关键数值与推荐

  • 以BI-RADS 4b为截断点评估诊断效能。
  • 全部乳腺中CBBCT灵敏度83.3%、特异度97.3%,乳腺X线摄影灵敏度70.0%、特异度98.7%。
  • 致密类乳腺中CBBCT灵敏度83.1%、特异度98.6%,乳腺X线摄影灵敏度69.0%、特异度98.6%。

临床怎么用

  • 对致密类乳腺,可考虑使用锥光束乳腺CT辅助诊断,以提高对恶性肿瘤的检出灵敏度。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Objective: To compare the diagnostic efficiency of lesion in dense breast between cone beam breast computer tomography (CBBCT) and mammography. Methods: From May 2012 to August 2014, 160 patients with 165 breasts who underwent mammography and CBBCT examinations were included in this study. The diagnostic results of CBBCT and mammography were reviewed and compared with pathological results. Results: In the 165 breast, 24 were dense breasts and 141 were dense breasts. The diagnostic results were similar in 109 lesions, but different in 56 lesions. According to the analysis of the 165 breasts using receiver operation characteristics (ROC) curve, the area under the ROC curves (AUC) of CBBCT and mammography were 0.923 (95%CI: 0.878-0.967, P<0.05) and 0.959 (95%CI: 0.926-0.992, P<0.05), respectively. With Breast Imaging-Reporting and Data System (BI-RADS) 4b as the critical value, the sensitivity and specificity were 70.0% and 98.7% using mammography, and 83.3% and 97.3% using CBBCT, respectively. The AUC of CBBCT and mammography of the 141 dense breasts was 0.919(95%CI: 0.868-0.969, P<0.05) and 0.973(95%CI: 0.947-0.999, P<0.05), respectively. With BI-RADS 4b as the critical value, the sensitivity and specificity were 69.0% and 98.6% by mammography, and 83.1% and 98.6% by CBBCT, respectively. Conclusions: CBBCT showed higher sensitivity and specificity in the diagnosis of breast malignant tumors compared to mammography. It is expected to be applied to breast cancer detection in the future, especially in dense breast. 目的: 比较锥光束乳腺CT与乳腺X线摄影对致密类乳腺中病灶的诊断效能。 方法: 收集同时接受乳腺X线摄影和锥光束乳腺CT检查的160例患者(165个乳腺)的影像和病理资料,比较乳腺X线摄影和锥光束乳腺CT的诊断结果,结合病理诊断结果评价两种影像检查方法在全部乳腺及致密类乳腺中的诊断效能。 结果: 165个乳腺中,非致密类乳腺24个,致密类乳腺141个。乳腺X线摄影和锥光束乳腺CT诊断结果一致者109个,不一致者56个。受试者工作特征(ROC)曲线分析显示,对于全部165个乳腺,乳腺X线摄影诊断的曲线下面积(AUC)为0.923(95%CI为0.878~0.967,P<0.05),锥光束乳腺CT诊断的AUC为0.959(95%CI为0.926~0.992,P<0.05)。以BI-RADS 4b为截断点,乳腺X线摄影诊断乳腺癌的灵敏度和特异度分别为70.0%和98.7%,锥光束乳腺CT的灵敏度和特异度分别为83.3%和97.3%。对于141个致密类乳腺,乳腺X线摄影诊断的AUC为0.919(95%CI为0.868~0.969,P<0.05),锥光束乳腺CT诊断的AUC为0.973(95%CI为0.947~0.999,P<0.05)。以BI-RADS 4b为截断点,乳腺X线摄影诊断致密类乳腺乳腺癌的灵敏度和特异度分别为69.0%和98.6%,锥光束乳腺CT的灵敏度和特异度分别为83.1%和98.6%。 结论: 锥光束乳腺CT对乳腺恶性肿瘤的诊断表现出了较高的灵敏度和特异度,有望为乳腺恶性肿瘤尤其是致密类乳腺的恶性肿瘤诊断提供更多有效信息。.

来源:Zhonghua Zhong Liu Za Zhi,查看原文。版权归原作者所有。

引用

  1. Liu AD, Ma Y, Yin L, et al. [Comparison of the diagnostic efficiency in breast malignancy between cone beam breast CT and mammography in dense breast]. Zhonghua Zhong Liu Za Zhi. 2018. PMID: 30139031. doi:10.3760/cma.j.issn.0253-3766.2018.08.008