深度学习去噪提升标准剂量SPECT心肌灌注显像灌注缺损检测准确性
该研究在190例临床SPECT-MPI数据上,以混合研究为金标准,通过ROC分析比较经与未经深度学习网络处理的图像对灌注缺损的检测准确性。结果显示,深度学习去噪可显著提高AUC,并在较少高斯后滤波时兼顾空间分辨率。
要点
- 深度学习去噪使标准剂量SPECT-MPI的灌注缺损检测AUC从0.80提升至0.88。
- 在较少高斯后滤波条件下,AUC从0.78提升至0.86且空间分辨率更优。
- 深度学习去噪优于传统重建,可改善标准剂量采集的缺损检测。
核心结论
- 深度学习去噪可有效改善标准剂量SPECT-MPI图像中异常缺损的检测,优于传统重建。
研究设计与样本
- 研究对190例临床SPECT-MPI数据重建图像进行ROC分析,比较经与未经深度学习网络处理的图像,以混合研究为灌注缺损检测金标准。
主要发现
- 与最优高斯后滤波(sigma=1.2体素)重建相比,进一步深度学习去噪使ROC曲线下面积(AUC)从0.80增至0.88(P值<10-4)。
- 与较少高斯后滤波(sigma=0.8体素)重建相比,深度学习去噪使AUC从0.78增至0.86(P值<10-4),并在重建中实现更好的空间分辨率。
关键数值与推荐
- 采用有序子集期望最大化(OSEM)重建,并校正衰减、分辨率和散射,结合三维高斯后滤波。
- 总灌注缺损(TPD)评分由定量灌注SPECT(QPS)软件计算,用于评估重建图像。
临床怎么用
- 在标准剂量SPECT-MPI临床采集时,可应用该深度学习去噪网络提升灌注缺损检测准确性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: We previously developed a deep-learning (DL) network for image denoising in SPECT-myocardial perfusion imaging (MPI). Here we investigate whether this DL network can be utilized for improving detection of perfusion defects in standard-dose clinical acquisitions. METHODS: To quantify perfusion-defect detection accuracy, we conducted a receiver-operating characteristic (ROC) analysis on reconstructed images with and without processing by the DL network using a set of clinical SPECT-MPI data from 190 subjects. For perfusion-defect detection hybrid studies were used as ground truth, which were created from clinically normal studies with simulated realistic lesions inserted. We considered ordered-subset expectation-maximization (OSEM) reconstruction with corrections for attenuation, resolution, and scatter and with 3D Gaussian post-filtering. Total perfusion deficit (TPD) scores, computed by Quantitative Perfusion SPECT (QPS) software, were used to evaluate the reconstructed images. RESULTS: Compared to reconstruction with optimal Gaussian post-filtering (sigma = 1.2 voxels), further DL denoising increased the area under the ROC curve (AUC) from 0.80 to 0.88 (P-value < 10-4). For reconstruction with less Gaussian post-filtering (sigma = 0.8 voxels), thus better spatial resolution, DL denoising increased the AUC value from 0.78 to 0.86 (P-value < 10-4) and achieved better spatial resolution in reconstruction. CONCLUSIONS: DL denoising can effectively improve the detection of abnormal defects in standard-dose SPECT-MPI images over conventional reconstruction.
引用
- Liu J, Yang Y, Wernick MN, et al. Improving detection accuracy of perfusion defect in standard dose SPECT-myocardial perfusion imaging by deep-learning denoising. J Nucl Cardiol. 2021. PMID: 34282538. doi:10.1007/s12350-021-02676-w
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。