Biograph Vision Quadra PET/CT系统按NEMA NU 2-2018标准的性能特征
本研究依据NEMA NU 2-2018标准评估Biograph Vision Quadra PET/CT的物理性能,并辅以能量分辨率和图像质量实验。该设备轴向视野106 cm,空间分辨率约3.3-3.8 mm,灵敏度在MRD 85和MRD 322下分别为83和176 cps/kBq,峰值NECR分别为1,613和2,956 kcps。临床18F-FDG图像中30秒帧可读且噪声可接受。
要点
- 空间分辨率径向、切向、轴向分别为3.3、3.4、3.8 mm。
- 灵敏度MRD 85为83 cps/kBq,MRD 322为176 cps/kBq。
- 峰值NECR在27.49 kBq/mL下MRD 85为1,613 kcps,MRD 322为2,956 kcps。
核心结论
- Biograph Vision Quadra的空间和时间分辨率与Biograph Vision 600相似,但因轴向视野延长而具有更高的灵敏度和NECR。
- 报告的空间分辨率、时间分辨率和灵敏度使其成为扩展轴向视野PET扫描仪类别中有竞争力的新设备。
研究设计与样本
- 依据NEMA NU 2-2018标准评估扫描仪物理性能,并附加能量分辨率表征实验。
- 图像质量以4:1和8:1的前景-背景比评估,并对临床18F-FDG PET研究采用不同帧时长重建。
- 所有实验采集使用最大环距离322晶体(MRD 322),图像重建可仅用最大环距离85晶体(MRD 85)。
主要发现
- 半高全宽空间分辨率径向3.3 mm、切向3.4 mm、轴向3.8 mm。
- 灵敏度MRD 85为83 cps/kBq,MRD 322为176 cps/kBq。
- 峰值NECR在27.49 kBq/mL下MRD 85为1,613 kcps,MRD 322为2,956 kcps。
- 峰值NECR处散射分数MRD 85为36%,MRD 322为37%;TOF分辨率MRD 85为228 ps,MRD 322为230 ps。
- 使用PSF-TOF 8迭代5子集重建时,图像对比恢复4:1为69.6%至86.9%,8:1为77.7%至92.6%。
关键数值与推荐
- NEMA NU 2-2018标准用于物理性能评估。
- 图像质量前景-背景比采用4:1和8:1。
- 临床图像30秒帧提供可读的病灶可检测性和可接受的噪声水平。
临床怎么用
- 该设备扩展轴向视野106 cm,可在1个床位覆盖主要器官。
- 临床18F-FDG PET中30秒帧即可获得可读病灶检测和可接受噪声。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Our purpose was to evaluate the performance of the Biograph Vision Quadra PET/CT system. This new system is based on the Biograph Vision 600, using the same silicon photomultiplier-based detectors with 3.2 × 3.2 × 20 mm lutetium-oxoorthosilicate crystals. The 32 detector rings of the Quadra provide a 4-fold larger axial field of view (AFOV) of 106 cm, enabling imaging of major organs in 1 bed position. Methods: The physical performance of the scanner was evaluated according to the National Electrical Manufacturers Association NU 2-2018 standard, with additional experiments to characterize energy resolution. Image quality was assessed with foreground-to-background ratios of 4:1 and 8:1. Additionally, a clinical 18F-FDG PET study was reconstructed with varying frame durations. In all experiments, data were acquired using the maximum ring distance of 322 crystals (MRD 322), whereas image reconstructions could be performed with a maximum ring distance of only 85 crystals (MRD 85). Results: The spatial resolution at full width at half maximum in the radial, tangential, and axial directions was 3.3, 3.4, and 3.8 mm, respectively. The sensitivity was 83 cps/kBq for MRD 85 and 176 cps/kBq for MRD 322. The noise-equivalent count rates (NECRs) at peak were 1,613 kcps for MRD 85 and 2,956 kcps for MRD 322, both at 27.49 kBq/mL. The respective scatter fractions at peak NECR equaled 36% and 37%. The time-of-flight resolution at peak NECR was 228 ps for MRD 85 and 230 ps for MRD 322. Image contrast recovery ranged from 69.6% to 86.9% for 4:1 contrast ratios and from 77.7% to 92.6% for 8:1 contrast ratios reconstructed using point-spread function time of flight with 8 iterations and 5 subsets. Thirty-second frames provided readable lesion detectability and acceptable noise levels in clinical images. Conclusion: The Biograph Vision Quadra PET/CT device has spatial and time resolution similar to those of the Biograph Vision 600 but exhibits improved sensitivity and NECR because of its extended AFOV. The reported spatial resolution, time resolution, and sensitivity make it a competitive new device in the class of PET scanners with an extended AFOV.
引用
- Prenosil GA, Sari H, Fürstner M, et al. Performance Characteristics of the Biograph Vision Quadra PET/CT System with a Long Axial Field of View Using the NEMA NU 2-2018 Standard. J Nucl Med. 2021. PMID: 34301780. doi:10.2967/jnumed.121.261972