不同[99mTc]Tc-DPD SPECT/CT定量重建技术在有症状髋膝关节假体中的准确性比较
该前瞻性研究纳入30例患者共33个有症状髋膝关节假体,比较xSPECT-Quant与xSPECT-Bone(联合或不联合iMAR)的SUVmax定量诊断假体松动的准确性。以术中诊断或至少1年临床随访为标准,xSPECT-Bone-iMAR准确性最高(93.9%),显著优于xSPECT-Quant-iMAR(81.8%)和不联合iMAR的xSPECT-Quant(77.4%)。
要点
- xSPECT-Bone-iMAR定量诊断假体松动的准确性达93.9%,为所有方法中最高。
- 松动假体周围平均SUVmax显著高于稳定假体,各定量方法均如此。
- 定量方法准确性数值上高于视觉判读(84.8%和81.5%),但差异未达显著。
核心结论
- xSPECT-Bone-iMAR定量在区分松动与稳定假体方面表现最佳。
- 不同定量方法总体准确性较高,支持[99mTc]Tc-DPD定量作为生物标志物的潜力。
研究设计与样本
- 前瞻性研究纳入30例连续患者共33个有症状髋膝关节假体,行[99mTc]Tc-DPD SPECT/CT。
- 以术中诊断或至少1年临床结局作为标准,计算SUVmax的截断值与准确性。
主要发现
- 10个假体诊断为松动,23个为稳定。
- 松动假体周围平均SUVmax显著高于稳定假体(P=0.0025-0.0001,各定量方法均如此)。
关键数值与推荐
- xSPECT-Bone-iMAR准确性93.9%(95%CI 79.6-100%),显著高于xSPECT-Quant-iMAR的81.8%(67.5-96.1%,P=0.04)。
- xSPECT-Bone-iMAR准确性显著高于不联合iMAR的xSPECT-Quant的77.4%(62.4-92.4%,P=0.02)。
- 视觉判读准确性为高年资医师84.8%(70.6-99.1%)、受训者81.5%(67.5-96.1%),与定量方法差异未达显著。
临床怎么用
- 对疑似髋膝关节假体松动的患者,可优先考虑采用xSPECT-Bone-iMAR进行[99mTc]Tc-DPD SPECT/CT定量评估。
- 定量SUVmax可作为辅助视觉判读的客观指标,但需结合临床综合判断。
局限性
- 样本量较小(30例患者33个假体),需在更大样本中进一步前瞻性验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: There is a need for better diagnostic tools that identify loose total hip and knee arthroplasties. Here, we present the accuracy of different 99mTc-dicarboxypropandiphosphate ([99mTc]Tc-DPD) SPECT/CT quantification tools for the detection of loose prostheses in patients with painful hip and knee arthroplasties. METHODS: Quantitative reconstruction of mineral phase SPECT data was performed using Siemens xSPECT-Quant and xSPECT-Bone, with and without metal artefact reduction (iMAR) of CT-data. Quantitative data (SUVmax values) were compared to intraoperative diagnosis or clinical outcome after at least 1 year as standard of comparison. Cut-off values and accuracies were calculated using receiver operator characteristics. Accuracy of uptake quantification was compared to the accuracy of visual SPECT/CT readings, blinded for the quantitative data and clinical outcome. RESULTS: In this prospective study, 30 consecutive patients with 33 symptomatic hip and knee prostheses underwent [99mTc]Tc-DPD SPECT/CT. Ten arthroplasties were diagnosed loose and 23 stable. Mean-SUVmax was significantly higher around loose prostheses compared to stable prostheses, regardless of the quantification method (P = 0.0025-0.0001). Quantification with xSPECT-Bone-iMAR showed the highest accuracy (93.9% [95% CI 79.6-100%]) which was significantly higher compared to xSPECT-Quant-iMAR (81.8% [67.5-96.1%], P = 0.04) and xSPECT-Quant without iMAR (77.4% [62.4-92.4%], P = 0.02). Accuracies of clinical reading were non-significantly lower compared to quantitative measures (84.8% [70.6-99.1%] (senior) and 81.5% [67.5-96.1%] (trainee)). CONCLUSION: Quantification with [99mTc]Tc-DPD xSPECT-Bone-iMAR discriminates best between loose and stable prostheses of all evaluated methods. The overall high accuracy of different quantitative measures underlines the potential of [99mTc]Tc-DPD-quantification as a biomarker and demands further prospective evaluation in a larger number of prosthesis.