长轴视野PET/CT延迟显像[68Ga]Ga-PSMA-11用于复发性前列腺癌诊断的可行性
该回顾性病例系列纳入10例复发性前列腺癌患者,在注射192±36 MBq [68Ga]Ga-PSMA-11后1小时行标准显像、4小时行延迟显像,均使用长轴视野(LAFOV)PET/CT(106 cm轴视野)。延迟显像主观图像质量仅轻度降低(4.2/5 vs 5/5),但肿瘤背景比显著改善(4小时3.41 vs 1小时1.93,p<0.001),信噪比改善接近显著性(33.02 vs 24.80,p=0.062)。结论认为在LAFOV系统上延迟显像可行且可能更优。
要点
- LAFOV PET/CT使[68Ga]Ga-PSMA-11延迟至4小时显像可行,总采集时间与常规轴视野扫描相当。
- 延迟显像肿瘤背景比显著提高(3.41 vs 1.93,p<0.001),信噪比有改善趋势(p=0.062)。
- 主观图像质量仅轻度下降(4.2/5 vs 5/5),支持在LAFOV系统上采用较晚采集时间。
核心结论
- 在LAFOV PET/CT上,[68Ga]Ga-PSMA-11延迟至4小时采集可改善肿瘤背景比和信噪比,主观图像质量仅轻度受损。
- 数据提示在LAFOV系统上使用较晚的[68Ga]Ga-PSMA-11采集时间可能更可取。
研究设计与样本
- 回顾性病例系列,纳入10例复发性前列腺癌患者。
- 注射192±36 MBq [68Ga]Ga-PSMA-11后1小时行标准显像,4小时行延迟显像。
- 使用LAFOV PET/CT(106 cm轴视野),1小时单床位采集6分钟,4小时采集16分钟。
- 由两名经验丰富的核医学医师共识阅片,评估图像质量(5分Likert量表)、SUV、肿瘤背景比和靶病灶信噪比。
主要发现
- 主观图像质量5分Likert评分:4小时为4.2/5,1小时为5/5。
- 肿瘤背景比在4小时显著改善(3.41 vs 1.93,p<0.001)。
- 信噪比在4小时有改善但仅为边缘显著性(33.02 vs 24.80,p=0.062)。
- 延迟显像总采集时间与标准轴视野扫描仪的常规检查相当。
临床怎么用
- 在配备LAFOV PET/CT的核医学科,可考虑对复发性前列腺癌患者行[68Ga]Ga-PSMA-11 4小时延迟显像以改善肿瘤背景比。
- 延迟显像虽可提高对比度,但需注意主观图像质量轻度下降,判读时应结合临床。
局限性
- 回顾性病例系列,样本量仅10例。
- 未报告SUV具体数值、病灶检出率及诊断准确性等数据。
- 信噪比改善仅为边缘显著性(p=0.062)。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: While acquisition of images in [68 Ga]Ga-PSMA-11 following longer uptake times can improve lesion uptake and contrast, resultant imaging quality and count statistics are limited by the isotope's half-life (68 min). Here, we present a series of cases demonstrating that when performed using a long axial field-of-view (LAFOV) PET/CT system, late imaging is feasible and can even provide improved image quality compared to regular acquisitions. METHODS: In this retrospective case series, we report our initial experiences with 10 patients who underwent standard imaging at 1 h p.i. following administration of 192 ± 36 MBq [68 Ga]Ga-PSMA-11 with additional late imaging performed at 4 h p.i. Images were acquired in a single bed position for 6 min at 1 h p.i. and 16 min p.i. at 4 h p.i. using a LAFOV scanner (106 cm axial FOV). Two experienced nuclear medicine physicians reviewed all scans in consensus and evaluated overall image quality (5-point Likert scale), lesion uptake in terms of standardised uptake values (SUV), tumour to background ratio (TBR) and target-lesion signal to background noise (SNR). RESULTS: Subjective image quality as rated on a 5-point Likert scale was only modestly lower for late acquisitions (4.2/5 at 4 h p.i.; 5/5 1 h p.i.), TBR was significantly improved (4 h: 3.41 vs 1 h: 1.93, p < 0.001) and SNR was improved with borderline significance (4 h: 33.02 vs 1 h: 24.80, p = 0.062) at later imaging. Images were obtained with total acquisition times comparable to routine examinations on standard axial FOV scanners. CONCLUSION: Late acquisition in tandem with a LAFOV PET/CT resulted in improvements in TBR and SNR and was associated with only modest impairment in subjective visual imaging quality. These data show that later acquisition times for [68 Ga]Ga-PSMA-11 may be preferable when performed on LAFOV systems.
引用
- Alberts I, Prenosil G, Mingels C, et al. Feasibility of late acquisition [68Ga]Ga-PSMA-11 PET/CT using a long axial field-of-view PET/CT scanner for the diagnosis of recurrent prostate cancer-first clinical experiences. Eur J Nucl Med Mol Imaging. 2021. PMID: 34155538. doi:10.1007/s00259-021-05438-5
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。