数字Biograph Vision上缩短扫描时间的[68Ga]Ga-PSMA PET/CT图像评估

Evaluation of [68Ga]Ga-PSMA PET/CT images acquired with a reduced scan time duration in prostate cancer patients using the digital biograph vision.
PubMed 文献 · EJNMMI Res · 2021-02-28

在20例前列腺癌患者中,比较数字PET/CT缩短扫描时间(平均4.6分钟 vs 参考16.7分钟)结合四种迭代重建算法的诊断效能。缩短协议下98%区域(91%受累区域)正确分类,SUVmax平均绝对偏差约9%,但约10%患者可能遗漏小病灶,5%患者因此降期。

要点

  • 扫描时间缩短至约1/3.5时,区域检出率仍达98%,SUVmax平均偏差约9%。
  • 小病灶(≤4mm淋巴结)可能遗漏,导致约10%患者漏诊、5%降期。
  • 转移性疾病患者可考虑缩短扫描时间或减少注射活度。

核心结论

  • 扫描时间缩短3.5倍时,区域检出(98%区域)和图像定量(平均偏差≤10%)高度一致。
  • 但小病灶可能遗漏,约10%患者漏诊,5%患者降期(T1N0M0而非T1N1M0)。
  • 转移性疾病患者可考虑缩短扫描时间或减少[68Ga]Ga-PSMA-11注射活度。

研究设计与样本

  • 20例前列腺癌患者(11例生化复发、5例初始分期、4例转移性疾病)序贯接受数字Siemens Biograph Vision的[68Ga]Ga-PSMA-11 PET/CT。
  • PET数据采用连续床运动模式采集,参考采集协议平均扫描时间16.7分钟,缩短采集协议平均4.6分钟。
  • 应用四种迭代重建算法:单独TOF或联合PSF校正,各含2或4次迭代。

主要发现

  • 缩短采集协议下,无论何种重建算法,98%区域(91%受累区域)被正确分类。
  • 2个淋巴结病灶(各≤4mm)未被识别,导致1/20例降期。
  • 每种重建算法的SUVmax(SUVpeak)平均绝对百分比偏差约9%(6%)。
  • PSF+TOF图像平均噪声从13%增至21%(4次迭代),从10%增至15%(2次迭代)。

关键数值与推荐

  • 诊断效能指标包括:区域检出率、肿瘤SUVmax和SUVpeak、基于肝脏活度分布的图像噪声。
  • 参考采集协议平均扫描时间16.7分钟,缩短采集协议平均4.6分钟(约3.5倍缩短)。

临床怎么用

  • 在转移性疾病患者中,可考虑缩短扫描时间或减少[68Ga]Ga-PSMA-11注射活度,因遗漏小病灶不影响患者管理。

局限性

  • 样本量小且异质性大。
  • 缺乏随访数据。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIM: [68Ga]Ga-PSMA-11 PET/CT allows for a superior detection of prostate cancer tissue, especially in the context of a low tumor burden. Digital PET/CT bears the potential of reducing scan time duration/administered tracer activity due to, for instance, its higher sensitivity and improved time coincidence resolution. It might thereby expand [68Ga]Ga-PSMA-11 PET/CT that is currently limited by 68Ge/68Ga-generator yield. Our aim was to clinically evaluate the influence of a reduced scan time duration in combination with different image reconstruction algorithms on the diagnostic performance. METHODS: Twenty prostate cancer patients (11 for biochemical recurrence, 5 for initial staging, 4 for metastatic disease) sequentially underwent [68Ga]Ga-PSMA-11 PET/CT on a digital Siemens Biograph Vision. PET data were collected in continuous-bed-motion mode with a mean scan time duration of 16.7 min (reference acquisition protocol) and 4.6 min (reduced acquisition protocol). Four iterative reconstruction algorithms were applied using a time-of-flight (TOF) approach alone or combined with point-spread-function (PSF) correction, each with 2 or 4 iterations. To evaluate the diagnostic performance, the following metrics were chosen: (a) per-region detectability, (b) the tumor maximum and peak standardized uptake values (SUVmax and SUVpeak), and (c) image noise using the liver's activity distribution. RESULTS: Overall, 98% of regions (91% of affected regions) were correctly classified in the reduced acquisition protocol independent of the image reconstruction algorithm. Two nodal lesions (each ≤ 4 mm) were not identified (leading to downstaging in 1/20 cases). Mean absolute percentage deviation of SUVmax (SUVpeak) was approximately 9% (6%) for each reconstruction algorithm. The mean image noise increased from 13 to 21% (4 iterations) and from 10 to 15% (2 iterations) for PSF + TOF and TOF images. CONCLUSIONS: High agreement at 3.5-fold reduction of scan time in terms of per-region detection (98% of regions) and image quantification (mean deviation ≤ 10%) was demonstrated; however, small lesions can be missed in about 10% of patients leading to downstaging (T1N0M0 instead of T1N1M0) in 5% of patients. Our results suggest that a reduction of scan time duration or administered [68Ga]Ga-PSMA-11 activities can be considered in metastatic patients, where missing small lesions would not impact patient management. Limitations include the small and heterogeneous sample size and the lack of follow-up.