SiPM PET-CT系统静息-负荷铷-82心肌灌注显像中运动所致图像退化的发生率
该研究利用数据驱动运动校正(DDMC)算法重建300幅负荷和300幅静息铷-82相对灌注图像,与未校正图像对比评估运动所致视觉差异。结果显示运动退化见于58%负荷图像和33%静息图像,2%的研究因运动致图像不可诊断,DDMC心脏识别与追踪全部成功。
要点
- 运动退化见于58%负荷和33%静息铷-82相对灌注图像。
- 2%的研究因运动导致图像不可诊断。
- Dwell Fraction(DF)指标可有效预测图像退化。
核心结论
- 运动退化存在于近半数相对灌注图像中,DDMC算法可稳健地预测、评估和校正图像退化。
研究设计与样本
- 研究纳入300幅负荷和300幅静息静态图像,分别用DDMC和未校正(NMC)重建,共600对图像被赋予视觉差异评分(VDS),并记录不可诊断图像数量,从数据中衍生Dwell Fraction(DF)以量化运动并预测图像退化。
主要发现
- 运动退化(VDS=1或2)在58%的负荷图像和33%的静息图像中明显。
- 7幅NMC图像不可诊断,来自6项研究,因运动导致不可诊断研究的率为2%。
- DF指标能有效预测图像退化。
- DDMC心脏识别和追踪在所有图像中均成功。
临床怎么用
- 在铷-82心肌灌注显像中应关注运动所致图像退化,可利用DDMC算法预测、评估和校正运动伪影。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Motion of the heart is known to affect image quality in cardiac PET. The prevalence of motion blurring in routine cardiac PET is not fully appreciated due to challenges identifying subtle motion artefacts. This study utilizes a recent prototype Data-Driven Motion Correction (DDMC) algorithm to generate corrected images that are compared with non-corrected images to identify visual differences in relative rubidium-82 perfusion images due to motion. METHODS: 300 stress and 300 rest static images were reconstructed with DDMC and without correction (NMC). The 600 DDMC/NMC image pairs were assigned Visual Difference Score (VDS). The number of non-diagnostic images were noted. A "Dwell Fraction" (DF) was derived from the data to quantify motion and predict image degradation. RESULTS: Motion degradation (VDS = 1 or 2) was evident in 58% of stress images and 33% of rest images. Seven NMC images were non-diagnostic-these originated from six studies giving a 2% rate of non-diagnostic studies due to motion. The DF metric was able to effectively predict image degradation. The DDMC heart identification and tracking was successful in all images. CONCLUSION: Motion degradation is present in almost half of all relative perfusion images. The DDMC algorithm is a robust tool for predicting, assessing and correcting image degradation.
引用
- Armstrong IS, Memmott MJ, Hayden C, et al. The prevalence of image degradation due to motion in rest-stress rubidium-82 imaging on a SiPM PET-CT system. J Nucl Cardiol. 2021. PMID: 33608851. doi:10.1007/s12350-021-02531-y
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。