文献速递神经核医学放射性药物与辐射防护

脑PET/CT研究对比恢复性能的协调标准

Harmonisation of PET/CT contrast recovery performance for brain studies.
PubMed 文献 · Eur J Nucl Med Mol Imaging · 2021-01-31

结果发现PSF建模与否存在系统差异,并提出了确保图像可比性的RC和GMWMr上下限,建议据此选择重建设置。

要点

  • PSF建模与非建模重建间存在系统差异,需分别制定协调标准。
  • 基于图像衍生放射性活度归一化,提出了PSFON和PSFOFF下的RC与GMWMr限值。
  • 建议以RCGMe和GMWMre作为选择重建设置的依据,以实现跨扫描仪可比性。

核心结论

  • 为实现跨扫描仪可比性,建议基于RCGMe和GMWMre选择重建设置,并前瞻性验证所提出的协调标准。

研究设计与样本

  • 扫描数据采用多种重建设置重建,图像刚性配准至模板(体素大小1.17×1.17×2.00 mm3),并定义多个感兴趣体积(VOI)。
  • 导出腐蚀(下标e)和非腐蚀灰质(GM)与白质(WM)VOI、模体中部冷点(VOIcold)及Hammers图谱VOI的恢复系数(RC)和灰白质比值(GMWMr)。
  • 评估左右半球差异及与参考图像的体素级差异。

主要发现

  • 使用与未使用点扩散函数建模(PSFON和PSFOFF)的重建之间存在系统差异。
  • 基于图像衍生放射性活度归一化,确保图像可比性的上下限如下:PSFON时,腐蚀VOI的RCGMe=[0.97-1.01]、GMWMre=[3.51-3.91],非腐蚀VOI的RCGM=[0.78-0.83]、GMWMr=[1.77-2.06];PSFOFF时,腐蚀VOI的RCGMe=[0.92-0.99]、GMWMre=[3.14-3.68],非腐蚀VOI的RCGM=[0.75-0.81]、GMWMr=[1.72-1.95]。

关键数值与推荐

  • PSFON重建:腐蚀VOI的RCGMe范围为[0.97-1.01],GMWMre范围为[3.51-3.91];非腐蚀VOI的RCGM范围为[0.78-0.83],GMWMr范围为[1.77-2.06]。
  • PSFOFF重建:腐蚀VOI的RCGMe范围为[0.92-0.99],GMWMre范围为[3.14-3.68];非腐蚀VOI的RCGM范围为[0.75-0.81],GMWMr范围为[1.72-1.95]。

临床怎么用

  • 在脑PET/CT研究中,应根据RCGMe和GMWMre等指标选择重建设置,以确保跨扫描仪图像质量的可比性。

局限性

  • 所提出的协调标准需前瞻性测试以验证和/或优化。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: In order to achieve comparability of image quality, harmonisation of PET system performance is imperative. In this study, prototype harmonisation criteria for PET brain studies were developed. METHODS: Twelve clinical PET/CT systems (4 GE, 4 Philips, 4 Siemens, including SiPM-based "digital" systems) were used to acquire 30-min PET scans of a Hoffman 3D Brain phantom filled with ~ 33 kBq·mL-1 [18F]FDG. Scan data were reconstructed using various reconstruction settings. The images were rigidly coregistered to a template (voxel size 1.17 × 1.17 × 2.00 mm3) onto which several volumes of interest (VOIs) were defined. Recovery coefficients (RC) and grey matter to white matter ratios (GMWMr) were derived for eroded (denoted in the text by subscript e) and non-eroded grey (GM) and white (WM) matter VOIs as well as a mid-phantom cold spot (VOIcold) and VOIs from the Hammers atlas. In addition, left-right hemisphere differences and voxel-by-voxel differences compared to a reference image were assessed. RESULTS: Systematic differences were observed for reconstructions with and without point-spread-function modelling (PSFON and PSFOFF, respectively). Normalising to image-derived activity, upper and lower limits ensuring image comparability were as follows: for PSFON, RCGMe = [0.97-1.01] and GMWMre = [3.51-3.91] for eroded VOI and RCGM = [0.78-0.83] and GMWMr = [1.77-2.06] for non-eroded VOI, and for PSFOFF, RCGMe = [0.92-0.99] and GMWMre = [3.14-3.68] for eroded VOI and RCGM = [0.75-0.81] and GMWMr = [1.72-1.95] for non-eroded VOI. CONCLUSIONS: To achieve inter-scanner comparability, we propose selecting reconstruction settings based on RCGMe and GMWMre as specified in "Results". These proposed standards should be tested prospectively to validate and/or refine the harmonisation criteria.

来源:Eur J Nucl Med Mol Imaging,查看原文。版权归原作者所有。

引用

  1. Verwer EE, Golla SSV, Kaalep A, et al. Harmonisation of PET/CT contrast recovery performance for brain studies. Eur J Nucl Med Mol Imaging. 2021. PMID: 33517517. doi:10.1007/s00259-021-05201-w

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。