基于负荷/静息心肌活性比值的PET心肌血流储备简化估计方法的内部验证

Internal validation of myocardial flow reserve PET imaging using stress/rest myocardial activity ratios with Rb-82 and N-13-ammonia.
PubMed 文献 · J Nucl Cardiol · 2021-01-03

本研究在分别使用82Rb和13NH3的连续患者中各250例中,利用负荷/静息心肌活性比值(MAR)通过逆幂函数简化估计心肌血流储备(MFREST),并与标准一室模型定量的MFR比较。两种示踪剂下MFREST与MFR均呈良好相关(13NH3 R=0.63,82Rb R=0.73),无偏倚,检测MFR<2的ROC曲线下面积分别为83.2±1.2%和90.4±0.7%。结论认为该简化方法无需动态成像或示踪剂动力学建模,可准确估计MFR。

要点

  • 负荷/静息心肌活性比值可简化估计MFR,与标准模型定量值相关性良好。
  • 简化MFR检测MFR<2的诊断性能良好,82Rb的AUC高于13NH3。
  • 该方法无需动态成像或示踪剂动力学建模,可用于常规质量保证。

核心结论

  • 使用82Rb和13NH3时,仅依赖负荷/静息活性比值的简化方法可较准确估计MFR。

研究设计与样本

  • 对连续患者分别使用82Rb或13NH3(各n=250)进行静息和双嘧达莫负荷动态PET成像。
  • 以标准一室模型定量的MFR为参考标准,用注射后2至6分钟左心室平均活性计算负荷/静息心肌活性比值(MAR),再通过逆幂函数得到简化MFR估计值(MFREST)。

主要发现

  • 13NH3组MFR与MFREST相关性良好(R=0.63),82Rb组结果相似(R=0.73)。
  • 两种示踪剂的MFREST值均无偏倚。
  • MFREST检测MFR<2的ROC曲线下面积:13NH3为83.2±1.2%,82Rb为90.4±0.7%。

临床怎么用

  • 该简化方法无需动态成像或示踪剂动力学建模,可用于PET MFR测量的常规质量保证。
  • 在因技术或后勤原因无法进行完整动态成像和示踪剂动力学建模的扫描仪上,可考虑使用该方法。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Myocardial flow reserve (MFR) measurement provides incremental diagnostic and prognostic information. The objective of the current study was to investigate the application of a simplified model for the estimation of MFR using only the stress/rest myocardial activity ratio (MAR) in patients undergoing rest-stress cardiac PET MPI. METHODS AND RESULTS: Rest and dipyridamole stress dynamic PET imaging was performed in consecutive patients using 82Rb or 13NH3 (n = 250 each). Reference standard MFR was quantified using a standard one-tissue compartment model. Stress/rest myocardial activity ratio (MAR) was calculated using the LV-mean activity from 2 to 6 minutes post-injection. Simplified estimates of MFR (MFREST) were then calculated using an inverse power function. For 13NH3, there was good correlation between MFR and MFREST values (R = 0.63), with similar results for 82Rb (R = 0.73). There was no bias in the MFREST values with either tracer. The overall diagnostic performance of MFREST for detection of MFR < 2 was good with ROC area under the curve (AUC) = 83.2 ± 1.2% for 13NH3 and AUC = 90.4 ± 0.7% for 82Rb. CONCLUSION: MFR was estimated with good accuracy using 82Rb and 13NH3 with a simplified method that relies only on stress/rest activity ratios. This novel approach does not require dynamic imaging or tracer kinetic modeling. It may be useful for routine quality assurance of PET MFR measurements, or in scanners where full dynamic imaging and tracer kinetic modeling is not feasible for technical or logistical reasons.