联影uPMR790一体化PET/MR系统的NEMA NU2-2012性能测量
本研究按NEMA标准评估联影uPMR790一体化PET/MR系统的PET性能,在MR空闲与同时运行2点DIXON序列两种条件下测量空间分辨率、灵敏度、散射分数、计数率、图像质量及TOF分辨率等指标,并与同数字探测器技术的PET/CT临床图像比较。结果显示MR序列同时运行对PET性能无显著影响,整体性能与同类先进PET/MR系统相当,临床图像质量与对比PET/CT相似。
要点
- MR同时运行对PET性能无显著影响。
- NECR峰值129.2 kcps(MR空闲)与129.5 kcps(MR运行)。
- TOF分辨率535 ps(MR空闲)与540 ps(MR运行)。
核心结论
- 新型一体化PET/MR系统的PET性能不受MR序列(2点DIXON)同时运行的显著影响。
- 测量结果与同类先进PET/MR系统性能相当。
研究设计与样本
- 按NEMA标准测量PET系统性能,分别在MR空闲与MR运行条件下进行。
- 评估指标包括空间分辨率、灵敏度、散射分数、计数率性能、计数丢失与随机校正精度、图像质量和TOF分辨率。
- 在PET/MR系统采集临床扫描,并与同数字探测器技术的PET/CT图像比较。
主要发现
- 在1 cm径向偏移处,轴向、径向和切向空间分辨率(FWHM)分别为2.72 mm(2.73 mm)、2.86 mm(2.85 mm)和2.81 mm(2.82 mm),括号内为MR运行条件。
- NECR峰值在14.7 kBq mL-1(14.2 kBq mL-1)时为129.2 kcps(129.5 kcps)。
- NECR峰值处散射分数为37.9%(36.5%),NECR以下最大断层误差为4.1%(4.5%)。
- 对比恢复系数从10 mm热球的51.8%(52.3%)到37 mm冷球的87.3%(87.2%)。
- 在5.3 kBq mL-1时TOF分辨率为535 ps(540 ps),点源测量TOF为474 ps(485 ps)。
临床怎么用
- MR序列同时采集不显著影响PET定量性能,支持PET/MR同步采集的可行性。
- 高空间分辨率和长轴向覆盖的临床获益仍需在特定临床成像应用中进一步评估。
局限性
- 高空间分辨率和长轴向覆盖的临床获益尚待进一步评估。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: In this paper, we aimed to evaluate the positron emission tomography (PET) performance of, to the best of our knowledge, the third commercially available whole-body integrated PET/magnetic resonance (MR) system. METHODS: The PET system performance was measured following the NEMA standards with and without simultaneous MR operation. PET spatial resolution, sensitivity, scatter fraction, count-rate performance, accuracy of count losses and random corrections, image quality, and time-of-flight (TOF) resolution were quantitatively evaluated. Clinical scans were acquired at the PET/MR system and compared with images acquired at a PET/CT with the same digital detector technology. RESULTS: Measurement results of essential PET performance were reported in the form of MR idle (MR pulsing). The axial, radial, and tangential spatial resolutions were measured as 2.72 mm (2.73 mm), 2.86 mm (2.85 mm), and 2.81 mm (2.82 mm) FWHM, respectively, at 1 cm radial offset. The NECR peak was measured as 129.2 kcps (129.5 kcps) at 14.7 kBq mL-1 (14.2 kBq mL-1). The scatter fraction at NECR peak was 37.9% (36.5%), and the maximum slice error below NECR was 4.1% (4.5%). Contrast recovery coefficients ranged from 51.8% (52.3%) for 10 mm hot sphere to 87.3% (87.2%) for 37 mm cold sphere. TOF resolution at 5.3 kBq mL-1 was measured at 535 ps (540 ps). With point source, TOF was measured to be 474 ps (485 ps). Clinical scans revealed similar image quality from the PET/MR and the comparative PET/CT system. CONCLUSION: The PET performance of the newly introduced integrated PET/MR system is not significantly affected by the simultaneous operation of an MR sequence (2-point DIXON sequence). Measurement results demonstrate comparable performance with other state-of-the-art PET/MR systems. The clinical benefits of high spatial resolution and long axial coverage remain to be further evaluated in specific clinical imaging applications.