文献速递心脏核医学

PET-CT错配对心脏15O-水PET定量准确性的影响

Effect of PET-CT misalignment on the quantitative accuracy of cardiac 15O-water PET.
PubMed 文献 · J Nucl Cardiol · 2020-11-04

结果显示MBF仅受轻微影响,而MBFt受影响较大;整体水平上MBF平均绝对偏差为1.7%±1.4%,MBFt为5.4%±3.2%。结论认为基于示踪剂洗脱的15O-水PET MBF定量准确性仅受PET与CT错配的轻微影响。

要点

  • 15O-水PET的MBF定量受PET-CT错配影响较小,整体平均绝对偏差1.7%±1.4%。
  • MBFt对错配更敏感,整体平均绝对偏差5.4%±3.2%,LCX在X+20时达-19.3%±9.6%。
  • LAD和LCX所受影响一般大于RCA。

核心结论

  • 基于示踪剂洗脱的15O-水PET MBF定量准确性仅受PET与CT错配的轻微影响。

研究设计与样本

  • 纳入10例临床患者,因缺血评估和MBF测定接受15O-水PET检查。

主要发现

  • MBF仅受错配轻微影响,而MBFt受影响较大。
  • 整体水平上,MBF平均绝对偏差为1.7%±1.4%,MBFt为5.4%±3.2%。
  • MBF最大偏差为-4.8%±5.8%(LCX,X+20),MBFt最大偏差为-19.3%±9.6%(LCX,X+20)。
  • LAD和LCX所受影响一般大于RCA。

临床怎么用

  • 解读15O-水PET的MBF定量时,PET-CT错配的影响可能较小,但MBFt指标需更谨慎对待。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Quantification of myocardial blood flow (MBF) with PET requires accurate attenuation correction, which is performed using a separate CT. Misalignment between PET and CT scans has been reported to be a common problem. The purpose of the present study was to assess the effect of PET CT misalignment on the quantitative accuracy of cardiac 15O-water PET. METHODS: Ten clinical patients referred for evaluation of ischemia and assessment of MBF with 15O-water were included in the study. Eleven different misalignments between PET and CT were induced in 6 different directions with 10 and 20 mm amplitudes: caudal (+Z), cranial (- Z), lateral (±X), anterior (+Y), and anterior combined with cranial (+ Y and - Z). Blood flow was quantified from rates of washout (MBF) and uptake (transmural MBF, MBFt) for the whole left ventricle and the three coronary territories. The results from all misalignments were compared to the original scan without misalignment. RESULTS: MBF was only minorly affected by misalignments, but larger effects were seen in MBFt. On the global level, average absolute deviation across all misalignments for MBF was 1.7% ± 1.4% and for MBFt 5.4% ± 3.2 Largest deviation for MBF was - 4.8% ± 5.8% (LCX, X + 20) and for MBFt - 19.3% ± 9.6% (LCX, X + 20). In general, larger effects were seen in LAD and LCX compared to in RCA. CONCLUSION: The quantitative accuracy of MBF from 15O-water PET, based on the washout of the tracer, is only to a minor extent affected by misalignment between PET and CT.

来源:J Nucl Cardiol,查看原文。版权归原作者所有。

引用

  1. Nordström J, Harms HJ, Kero T, et al. Effect of PET-CT misalignment on the quantitative accuracy of cardiac 15O-water PET. J Nucl Cardiol. 2020. PMID: 33146863. doi:10.1007/s12350-020-02408-6

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。