uEXPLORER全身PET/CT基于NEMA NU 2-2018的性能评估及长轴视野附加测试

Performance Evaluation of the uEXPLORER Total-Body PET/CT Scanner Based on NEMA NU 2-2018 with Additional Tests to Characterize PET Scanners with a Long Axial Field of View.
PubMed 文献 · J Nucl Med · 2020-10-02

该文对首台商用全身PET/CT系统uEXPLORER(轴向视野194 cm)按NEMA NU 2-2018进行物理性能评估,并新增基于人体平均身高(约165 cm)扩展体模的灵敏度和计数率测试,以及5个轴向位置的图像质量评估。结果显示系统灵敏度达174 kcps/MBq,全身成像峰值噪声等效计数率约2 Mcps,视野中心附近空间分辨率半高全宽≤3.0 mm,在低剂量、延迟成像和短扫描时间条件下均保持良好图像质量。

要点

  • uEXPLORER轴向视野194 cm,系统灵敏度174 kcps/MBq。
  • 全身成像峰值噪声等效计数率约2 Mcps。
  • 视野中心附近空间分辨率半高全宽≤3.0 mm。

核心结论

  • uEXPLORER提供超高灵敏度,在体模和人体成像中支持全视野优异的空间分辨率和图像质量。

研究设计与样本

  • 基于NEMA NU 2-2018协议评估灵敏度、计数率性能、飞行时间分辨率、空间分辨率和图像质量。
  • 使用基于世界平均人体身高(约165 cm)的扩展几何体模进行更代表全身成像的灵敏度和计数率测量。
  • 用NEMA图像质量体模在5个轴向位置及低剂量、延迟成像、短扫描时间条件下评估长轴视野内图像质量。

主要发现

  • 系统灵敏度为174 kcps/MBq。
  • 全身成像峰值噪声等效计数率约2 Mcps。
  • 人体成像空间分辨率在视野中心附近半高全宽≤3.0 mm。
  • NEMA图像质量体模评估和人体成像示例均显示全视野优异图像质量、优异对比度恢复和低噪声特性。
  • NEMA NU 2 70 cm体模与更代表全身成像的体模在灵敏度、计数率-活度关系和噪声等效计数率限值上存在差异,源于死时间和随机分数不同。

关键数值与推荐

  • 系统灵敏度174 kcps/MBq。
  • 全身成像峰值噪声等效计数率约2 Mcps。
  • 视野中心附近空间分辨率半高全宽≤3.0 mm。

临床怎么用

  • 该系统的超高灵敏度支持在低剂量、延迟成像和短扫描时间条件下获得全视野优异图像质量。
  • 对长轴视野PET系统进行性能评估时,除NEMA NU 2-2018外,宜采用基于人体身高的扩展体模以更准确表征全身成像性能。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

The world's first total-body PET scanner with an axial field of view (AFOV) of 194 cm is now in clinical and research use at our institution. The uEXPLORER PET/CT system is the first commercially available total-body PET scanner. Here we present a detailed physical characterization of this scanner based on National Electrical Manufacturers Association (NEMA) NU 2-2018 along with a new set of measurements devised to appropriately characterize the total-body AFOV. Methods: Sensitivity, count-rate performance, time-of-flight resolution, spatial resolution, and image quality were evaluated following the NEMA NU 2-2018 protocol. Additional measurements of sensitivity and count-rate capabilities more representative of total-body imaging were performed using extended-geometry phantoms based on the world-average human height (∼165 cm). Lastly, image quality throughout the long AFOV was assessed with the NEMA image quality (IQ) phantom imaged at 5 axial positions and over a range of expected total-body PET imaging conditions (low dose, delayed imaging, short scan duration). Results: Our performance evaluation demonstrated that the scanner provides a very high sensitivity of 174 kcps/MBq, a count-rate performance with a peak noise-equivalent count rate of approximately 2 Mcps for total-body imaging, and good spatial resolution capabilities for human imaging (≤3.0 mm in full width at half maximum near the center of the AFOV). Excellent IQ, excellent contrast recovery, and low noise properties were illustrated across the AFOV in both NEMA IQ phantom evaluations and human imaging examples. Conclusion: In addition to standard NEMA NU 2-2018 characterization, a new set of measurements based on extending NEMA NU 2-2018 phantoms and experiments was devised to characterize the physical performance of the first total-body PET system. The rationale for these extended measurements was evident from differences in sensitivity, count-rate-activity relationships, and noise-equivalent count-rate limits imposed by differences in dead time and randoms fraction between the NEMA NU 2 70-cm phantoms and the more representative total-body imaging phantoms. Overall, the uEXPLORER PET system provides ultra-high sensitivity that supports excellent spatial resolution and IQ throughout the field of view in both phantom and human imaging.