18F-FDG PET/CT诊断心脏植入电子装置感染的准确性
该前瞻性研究纳入63例疑似心脏植入电子装置(CIED)感染患者,评估18F-FDG PET/CT鉴别囊袋感染(PI)与导线感染(CIED-IE)的诊断效能。结果显示,PET/CT对CIED-IE敏感性仅38.5%、特异性98.0%;对PI敏感性72.2%、特异性95.6%。阳性PET/CT高度提示CIED-IE,但阴性结果不能排除诊断。
要点
- PET/CT诊断CIED-IE特异性高(98.0%)但敏感性低(38.5%)。
- PET/CT诊断囊袋感染敏感性72.2%、特异性95.6%。
- 临床高度怀疑CIED-IE时,阴性PET/CT需谨慎解读。
核心结论
- 18F-FDG PET/CT对疑似CIED感染的诊断效能因感染部位而异,对CIED-IE敏感性低、特异性高,阴性结果不能排除诊断。
研究设计与样本
- 2013至2018年前瞻性纳入63例疑似CIED感染住院患者,均行PET/CT检查,并行装置及囊袋组织培养。
主要发现
- 63例中14例为孤立性囊袋感染,13例为CIED-IE。
- 以导线心内部分放射性摄取判断CIED-IE,敏感性38.5%、特异性98.0%,阳性似然比19.2、阴性似然比0.6。
- 诊断囊袋感染敏感性72.2%、特异性95.6%。
关键数值与推荐
- CIED-IE诊断标准:PET/CT显示导线心内部分放射性摄取。
- 囊袋感染诊断标准:PET/CT显示囊袋部位放射性摄取。
临床怎么用
- 阳性PET/CT高度提示CIED-IE,可支持诊断;但阴性结果不能排除CIED-IE,尤其临床高度怀疑时。
- PET/CT对囊袋感染诊断敏感性较高,有助于识别囊袋感染。
局限性
- 样本量较小(n=63),CIED-IE病例数有限(13例)。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Utility of 18F-FDG PET/CT in diagnosing infective endocarditis (IE) associated with cardiac implantable electronic devices (CIEDs) is not well established. Current ESC guidelines recommend the use of FDG-PET imaging in patients with CIEDs and positive blood cultures, but the number of studies evaluating the diagnostic performance of FDG-PET imaging in these patients remain limited. Our objective was to assess the diagnostic yield of 18F-FDG PET/CT in patients with suspected CIED infections, differentiating between pocket infection (PI) and lead infection (CIED-IE). METHODS AND RESULTS: From 2013 to 2018, all patients (n = 63) admitted to a hospital with suspected CIED infection were prospectively recruited, undergoing a diagnostic work-up including a PET/CT. Explanted devices and material from the pocket were cultured. 14 cases corresponded to isolated PI and 13 were categorized as CIED-IE. Considering radionuclide uptake in the intracardiac portion of the lead, sensitivity and specificity of PET/CT for CIED-IE were 38.5% and 98.0%, respectively. Positive (19.2) and negative (0.6) likelihood ratio values, suggest that a positive PET/CT is much more probable to correspond to a patient with CIED-IE, whereas it is not possible to exclude this diagnosis when negative. For PI, sensitivity and specificity were 72.2% and 95.6%, respectively. CONCLUSIONS: The yield of 18F-FDG PET/CT for suspected CIED infections differs depending on the site of infection. Due to very high specificity but poor sensitivity, negative studies must be interpreted with caution if the suspicion of CIED-IE is high.
引用
- Jerónimo A, Olmos C, Vilacosta I, et al. Accuracy of 18F-FDG PET/CT in patients with the suspicion of cardiac implantable electronic device infections. J Nucl Cardiol. 2020. PMID: 32748277. doi:10.1007/s12350-020-02285-z
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。