文献速递心脏核医学放射性药物与辐射防护

18F-FDG PET/CT 诊断心脏植入式电子装置感染的敏感性与局限性

18F-FDG positron emission tomography/computed tomography of cardiac implantable electronic device infections.
PubMed 文献 · J Nucl Cardiol · 2020-07-31

结果显示近期植入(≤8周)者视觉阳性率高但仅部分确诊感染,远期植入者以TBR≥1.8为界,诊断敏感度90%、特异度73%、PPV 75%、NPV 89%。结论认为该方法敏感但不特异。

要点

  • 近期(≤8周)植入者PET/CT视觉阳性常见,但仅部分为真感染。
  • 远期植入者以TBR≥1.8为界,敏感度90%、特异度73%、PPV 75%、NPV 89%。

核心结论

  • 18F-FDG PET/CT诊断CIED感染敏感但不特异,近期植入者假阳性多见。

研究设计与样本

  • 最终诊断基于临床与细菌学数据,图像行视觉及半定量(SUVmax、TBR)评估。

主要发现

  • 近期(≤8周)植入的9例患者视觉均阳性,但仅4例确诊CIED感染。
  • 远期植入的21例中真阳性9例,假阳性3例(14.3%);对照组假阳性3例(30%)。

关键数值与推荐

  • 以TBR≥1.8为界,远期植入患者诊断CIED感染的敏感度90%、特异度73%、PPV 75%、NPV 89%。

临床怎么用

  • 对远期植入疑似CIED感染者,可结合TBR≥1.8辅助诊断,但需警惕假阳性。
  • 近期(≤8周)植入者PET/CT阳性需谨慎解读,避免仅凭视觉阳性诊断感染。

局限性

  • 对照组为肿瘤患者,可能存在混杂因素影响SUVmax与TBR解读。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: The diagnosis of cardiac implantable electronic device (CIED) infection is challenging because of its variable presentations. We studied the value of 2-[18F]fluoro-2-deoxy-D-glucose (18F-FDG) positron emission tomography/computed tomography (PET/CT) in the detection of CIED infection. METHODS AND RESULTS: Thirty patients with suspected CIED infection underwent 18F-FDG-PET/CT. The control group was ten patients with asymptomatic CIED who underwent cancer-related 18F-FDG-PET/CT. 18F-FDG-PET/CT was evaluated visually, semiquantitatively as maximum standardized uptake value (SUVmax) and target-to-background ratio (TBR). Final diagnosis of CIED infection was based on clinical and bacteriological data. 18F-FDG-PET/CT was visually positive in all 9 patients with recent (≤ 8 weeks) implantation of CIED, but only 4 had confirmed CIED infection. 18F-FDG-PET/CT was true positive in 9 out of 21 cases with remote implantation of CIED and false positive in 3 (14.3%) cases. 18F-FDG-PET/CT was also false positive in 3 (30%) cases of control group. The SUVmax of the pocket area was significantly higher in patients with CIED infection than in the control group (4.8 ± 2.4 vs 2.0 ± .8, P < .001). By using the cut-off value of TBR ≥ 1.8, sensitivity of 18F-FDG-PET/CT for the diagnosis of CIED infection in patients with remote implantation was 90% and specificity 73%, PPV 75%, and NPV 89%. CONCLUSIONS: 18F-FDG-PET/CT is a sensitive but nonspecific method in the diagnosis of CIED infection.

来源:J Nucl Cardiol,查看原文。版权归原作者所有。

引用

  1. Salomäki SP, Saraste A, Kemppainen J, et al. 18F-FDG positron emission tomography/computed tomography of cardiac implantable electronic device infections. J Nucl Cardiol. 2020. PMID: 32737839. doi:10.1007/s12350-020-02256-4

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。