18F-FDG PET/CT 对左心室辅助装置患者装置相关感染诊断价值的系统评价与荟萃分析

The value of 18F-FDG PET/CT for the diagnosis of device-related infections in patients with a left ventricular assist device: a systematic review and meta-analysis.
PubMed 文献 · Eur J Nucl Med Mol Imaging · 2020-06-27

该系统评价与荟萃分析纳入10项研究(其中8项进入荟萃分析,共256次FDG-PET/CT检查、230例患者),评估18F-FDG PET/CT诊断左心室辅助装置相关感染的效能。汇总敏感度为0.95(95%CI 0.89-0.97),汇总特异度为0.91(95%CI 0.54-0.99);泵/囊袋感染与驱动线感染的敏感度分别为0.97和0.96。作者结论认为FDG-PET/CT是有价值的评估工具,敏感度高、特异度较高但存在变异,需标准化检查流程与判定标准。

要点

  • 汇总敏感度0.95(95%CI 0.89-0.97),汇总特异度0.91(95%CI 0.54-0.99)。
  • 泵/囊袋感染:敏感度0.97(95%CI 0.69-1.00),特异度0.93(95%CI 0.64-0.99)。
  • 驱动线感染:敏感度0.96(95%CI 0.88-0.99),特异度0.99(95%CI 0.13-1.00)。

核心结论

  • FDG-PET/CT是评估左心室辅助装置患者装置相关感染的有价值工具,敏感度高、特异度较高但存在变异。
  • 需要标准化FDG-PET/CT检查流程以及疑似装置相关左心室辅助装置感染的判定标准,以实现FDG-PET/CT扫描报告的一致性。

研究设计与样本

  • 系统评价与荟萃分析,检索多个数据库自建库至2019年12月31日的文献。
  • 纳入研究FDG-PET/CT对疑似左心室辅助装置感染患者诊断效能的研究。
  • 采用QUADAS-2进行偏倚风险评估后,以每次检查为单位进行汇总荟萃分析。
  • 系统评价共纳入10项研究,其中8项符合汇总荟萃分析条件。
  • 荟萃分析共纳入230例患者的256次FDG-PET/CT扫描,检查泵/囊袋和/或驱动线感染。

主要发现

  • FDG-PET/CT诊断装置相关感染的汇总敏感度为0.95(95%CI 0.89-0.97),汇总特异度为0.91(95%CI 0.54-0.99)。
  • 泵/囊袋感染的敏感度为0.97(95%CI 0.69-1.00),特异度为0.93(95%CI 0.64-0.99)。
  • 驱动线感染的敏感度为0.96(95%CI 0.88-0.99),特异度为0.99(95%CI 0.13-1.00)。
  • 各研究间特异度存在显著异质性,主要由扫描流程差异引起。
  • 所有纳入研究均缺乏预先定义的疑似左心室辅助装置和/或驱动线感染判定标准。

关键数值与推荐

  • FDG-PET/CT诊断装置相关感染:汇总敏感度0.95(95%CI 0.89-0.97),汇总特异度0.91(95%CI 0.54-0.99)。
  • FDG-PET/CT诊断泵/囊袋感染:敏感度0.97(95%CI 0.69-1.00),特异度0.93(95%CI 0.64-0.99)。
  • FDG-PET/CT诊断驱动线感染:敏感度0.96(95%CI 0.88-0.99),特异度0.99(95%CI 0.13-1.00)。

临床怎么用

  • 作者结论认为FDG-PET/CT是评估左心室辅助装置患者装置相关感染的有价值工具。
  • 作者指出需要标准化FDG-PET/CT检查流程和疑似装置相关左心室辅助装置感染的判定标准,以支持扫描报告的一致性。

局限性

  • 各研究间特异度存在显著异质性,主要由扫描流程差异引起。
  • 所有纳入研究均缺乏预先定义的疑似左心室辅助装置和/或驱动线感染判定标准。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Left ventricular assist devices (LVADs) are increasingly used for the treatment of advanced heart failure. LVADs improve quality of life and decrease mortality, but the driveline carries substantial risk for major infections. These device-related LVAD and driveline infections are difficult to diagnose with conventional imaging. We reviewed and analysed the current literature on the additive value of 18F-fluorodeoxyglucose positron emission tomography combined with computed tomography (FDG-PET/CT) imaging for the diagnosis of LVAD-related infections." MATERIALS/METHODS: We performed a systematic literature review using several databases from their inception until the 31st of December, 2019. Studies investigating the diagnostic performance of FDG-PET/CT in patients with suspected LVAD infection were retrieved. After a bias risk assessment using QUADAS-2, a study-aggregate meta-analysis was performed on a per examination-based analysis. RESULTS: A total of 10 studies were included in the systematic review, eight of which were also eligible for study-aggregate meta-analysis. For the meta-analysis, a total of 256 FDG-PET/CT scans, examining pump/pocket and/or driveline infection, were acquired in 230 patients. Pooled sensitivity of FDG-PET/CT was 0.95 (95% confidence interval (CI) 0.89-0.97) and pooled specificity was 0.91 (95% CI 0.54-0.99) for the diagnosis of device-related infection. For pump/pocket infection, sensitivity and specificity of FDG-PET/CT were 0.97 (95%CI 0.69-1.00) and 0.93 (95%CI 0.64-0.99), respectively. For driveline infection, sensitivity and specificity were 0.96 (95%CI 0.88-0.99) and 0.99 (95%CI 0.13-1.00) respectively. Significant heterogeneity existed across studies for specificity, mostly caused by differences in scan procedures. Predefined criteria for suspicion of LVAD and/or driveline infection were lacking in all included studies. CONCLUSIONS: FDG-PET/CT is a valuable tool for assessment of device-related infection in LVAD patients, with high sensitivity and high, albeit variable, specificity. Standardization of FDG-PET/CT procedures and criteria for suspected device-related LVAD infections are needed for consistent reporting of FDG-PET/CT scans.