文献速递心脏核医学

82Rb心脏PET的活度方案:对绝对心肌血流量和心肌灌注成像的影响

Activity regimes for 82Rb cardiac PET: Effects on absolute MBF and MPI.
PubMed 文献 · J Nucl Cardiol · 2020-07-16

本研究在40例患者中比较了推荐活度1110 MBq与降低活度740或370 MBq对82Rb PET绝对心肌血流量、心肌血流储备及相对灌注成像的影响。结果显示1110 MBq与740 MBq的绝对心肌血流量相关性高(R²>0.98),但1110/370的Bland-Altman一致性界限更大;降低至740 MBq显著减少了死时间校正因子,仍能产生可靠的静态和门控图像。

要点

  • 1110 MBq与740或370 MBq的绝对心肌血流量线性相关R²>0.98,但1110/370一致性界限更大。
  • 心肌血流储备线性相关较低(R²=0.96),1110/370的一致性界限尤其宽。
  • 总灌注缺损和射血分数在1110/740相关性好(R²分别为0.96和0.99),1110/370散点大。

核心结论

  • 将示踪剂活度降至740 MBq显著减少了死时间校正因子,同时仍能产生可靠的静态和门控图像。
  • 尽管1110 MBq时死时间较大,但与降低活度相比,绝对心肌血流量未观察到系统性偏倚。

研究设计与样本

  • 四十例患者在同一台基于光电倍增管的PET/CT(GE D690)上接受两次扫描:推荐活度1110 MBq与740或370 MBq之一。
  • 量化了心肌血流量、心肌血流储备、总灌注缺损和射血分数,并使用线性回归和Bland-Altman图比较结果。

主要发现

  • 1110 MBq与任一降低活度之间心肌血流量的线性相关R²>0.98。
  • 观察到小幅偏倚(±5%-9%),1110/740与1110/370的偏倚方向相反。
  • 1110/370的心肌血流量一致性界限更大。
  • 心肌血流储备的线性相关较低(R²=0.96),但一致性界限宽,尤其1110/370。
  • 总灌注缺损和射血分数在1110/740相关性好(R²分别为0.96和0.99),但1110/370观察到较大散点。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Selection of optimal tracer activity for 82Rb PET is based on a trade-off between necessary count-statistics in the late static phase and detector saturation in the early blood-pool phase. Administered tracer activity recommended in prescribing information differs substantially from recommendations in current literature. The present study examines the effect on both absolute myocardial blood flow (MBF), myocardial flow reserve (MFR) and relative myocardial perfusion imaging (MPI) of reducing dose. METHODS: Forty patients were scanned twice on a PMT-based PET/CT (GE D690): At recommended activity (1110 MBq) and at either 740 or 370 MBq. MBF, MFR, total perfusion deficit (TPD) and ejection fractions (EF) were quantified. Results were compared using linear regression and Bland-Altman plots. RESULTS: Linear correlation between MBF at 1110 MBq at either reduced activity had an R2 > 0.98. A small bias (± 5%-9%) was observed with opposite signs for 1110/740 and 1110/370. Limits of agreement for MBF were larger for 1110/370. MFR had a lower linear correlation (R2 = 0.96), but wide limits of agreement especially for 1110/370. TPD and EF correlated well at 1110/740 (R2 = 0.96 and 0.99, respectively), but large scatter was observed for 1110/370. CONCLUSION: Reduction of the tracer activity to 740 MBq, significantly reduced dead-time correction factors, while still producing reliable static and gated images. However, despite large dead-time at 1110 MBq, no systematic bias on absolute MBF was observed compared to reduced activities.

来源:J Nucl Cardiol,查看原文。版权归原作者所有。