呼吸平均CT与标准CT衰减图对混合PET/CT冠状动脉粥样硬化病变18F-氟化钠摄取校正的比较
本研究纳入23例在混合PET/CT系统接受18F-氟化钠冠状动脉PET的患者,比较呼吸平均CT衰减校正(RACTAC)与标准单时相CT衰减校正(CTAC)对34个冠状动脉病变定量指标的影响。结果显示两种方法在最大靶本比(TBRMAX)和冠状动脉微钙化活动(CMA)上一致性好且可重复,TBRMAX中位数CTAC=1.65 vs RACTAC=1.63(p=0.55),CMA中位数CTAC=0.10 vs RACTAC=0.15(p=0.55)。
要点
- 呼吸平均CT衰减校正与标准单时相CT衰减校正对冠状动脉18F-氟化钠摄取的定量结果相似。
- TBRMAX中位数CTAC=1.65 vs RACTAC=1.63(p=0.55),可重复性系数0.18。
- CMA中位数CTAC=0.10 vs RACTAC=0.15(p=0.55),可重复性系数0.17。
核心结论
- 呼吸平均与标准单时相衰减校正图提供相似且可重复的冠状动脉18F-氟化钠摄取定量方法。
研究设计与样本
- 本研究纳入23例在混合PET/CT系统接受18F-氟化钠冠状动脉PET的患者,所有患者均采集自由呼吸下的标准单时相CTAC和4D cine-CT扫描。
- 从cine-CT采集数据中通过平均5秒内获取的所有图像获得RACTAC图,并比较使用CTAC或RACTAC的PET重建。
- 采用最大靶本比(TBRMAX)和冠状动脉微钙化活动(CMA)评估RACTAC的定量影响,使用可重复性系数和Bland-Altman图分析统计差异。
主要发现
- 在23例患者中,使用CTAC和RACTAC重建评估了34个冠状动脉病变。
- TBRMAX在CTAC与RACTAC之间一致性良好:CTAC中位数1.65(四分位距1.23至2.38)vs RACTAC中位数1.63(四分位距1.23至2.33),p=0.55,可重复性系数0.18。
- CMA在CTAC与RACTAC之间一致性良好:CTAC中位数0.10(四分位距0至1.0)vs RACTAC中位数0.15(四分位距0至1.03),p=0.55,可重复性系数0.17。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: To evaluate the impact of respiratory-averaged computed tomography attenuation correction (RACTAC) compared to standard single-phase computed tomography attenuation correction (CTAC) map, on the quantitative measures of coronary atherosclerotic lesions of 18F-sodium fluoride (18F-NaF) uptake in hybrid positron emission tomography and computed tomography (PET/CT). METHODS: This study comprised 23 patients who underwent 18F-NaF coronary PET in a hybrid PET/CT system. All patients had a standard single-phase CTAC obtained during free-breathing and a 4D cine-CT scan. From the cine-CT acquisition, RACTAC maps were obtained by averaging all images acquired over 5 seconds. PET reconstructions using either CTAC or RACTAC were compared. The quantitative impact of employing RACTAC was assessed using maximum target-to-background (TBRMAX) and coronary microcalcification activity (CMA). Statistical differences were analyzed using reproducibility coefficients and Bland-Altman plots. RESULTS: In 23 patients, we evaluated 34 coronary lesions using CTAC and RACTAC reconstructions. There was good agreement between CTAC and RACTAC for TBRMAX (median [Interquartile range]): CTAC = 1.65 [1.23 to 2.38], RACTAC = 1.63 [1.23 to 2.33], p = 0.55), with coefficient of reproducibility of 0.18, and CMA: CTAC = 0.10 [0 to 1.0], RACTAC = 0.15 [0 to 1.03], p = 0.55 with coefficient of reproducibility of 0.17 CONCLUSION: Respiratory-averaged and standard single-phase attenuation correction maps provide similar and reproducible methods of quantifying coronary 18F-NaF uptake on PET/CT.