在CZT相机上使用超低剂量铊-201 SPECT评估左心室容积和射血分数:与磁共振成像的比较
研究纳入45例因心肌灌注或存活心肌成像转诊的患者,分别用CZT-SPECT超低剂量方案(0.5 Mbq 201-Tl/kg体重)和1.5 T磁共振评估左心室舒张末期容积、收缩末期容积、射血分数和心肌质量。结果显示射血分数、收缩末期容积、舒张末期容积相关性很好,心肌质量相关性中等;与MRI相比,SPECT系统性高估收缩末期容积和心肌质量、低估射血分数(P ≤ .001),舒张末期容积无差异。
要点
- 超低剂量铊-201 CZT-SPECT与MRI的射血分数、收缩末期容积、舒张末期容积相关性很好,Pearson系数分别为0.86、0.95、0.91。
- 心肌质量相关性中等,r = 0.57。
- 与MRI相比,SPECT系统性高估收缩末期容积和心肌质量、低估射血分数(P ≤ .001),舒张末期容积无差异。
核心结论
- 超低剂量CZT-SPECT评估的左心室容积和射血分数与MRI所得值相关性很好,心肌质量相关性中等。
研究设计与样本
- 研究纳入45例因心肌灌注或存活心肌成像转诊的患者,分别接受CZT-SPECT和1.5 T磁共振扫描。
- CZT-SPECT采用超低剂量方案,201-Tl按每公斤体重0.5 Mbq给药。
- 两种技术均评估收缩末期容积、舒张末期容积、左心室射血分数和心肌质量。
主要发现
- 射血分数、收缩末期容积、舒张末期容积在CZT-SPECT与心脏磁共振MRI之间相关性很好,Pearson系数分别为0.86、0.95、0.91。
- 心肌质量相关性中等,r = 0.57。
- 与MRI相比,SPECT系统性高估收缩末期容积和心肌质量、低估射血分数,P ≤ .001。
- 舒张末期容积估算在两种方法间无差异。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Cadmium-Zinc-Telluride (CZT) technology allows use of low activities of radiopharmaceuticals. The aim was to verify the values of left ventricular volume parameters, obtained via ultra-low-dose thallium Single Photon Emission Computed Tomography (SPECT) using a CZT camera. METHODS AND RESULTS: Forty-five patients referred for an assessment of myocardial perfusion or viability imaging were examined using CZT-SPECT and 1.5 T magnetic resonance (MRI) scanner. The ultra-low-dose protocol with 0.5 Mbq 201-Tl per kg of body weight was used. The values of end-systolic (ESV) and end-diastolic volumes (EDV), left ventricular ejection fraction (EF) and myocardial mass (MM) were assessed using both techniques. A very good correlation was found between the EF, ESV, and EDV values assessed with CZT-SPECT and cardiac magnetic resonance MRI; the Pearson coefficients were 0.86, 0.95, and 0.91, respectively. A moderate correlation was found for myocardial mass, r = 0.57. Compared to MRI, SPECT systematically overestimated ESV and MM, while it underestimates the EF, with P ≤ .001 in all cases. There was no difference in EDV estimation. CONCLUSIONS: Left ventricular volumes and ejection fraction assessed via ultra-low-dose CZT-SPECT showed very good correlation with the values obtained by MRI. A moderate correlation was found for myocardial mass.