文献速递心脏核医学

在CZT相机上使用超低剂量铊-201 SPECT评估左心室容积和射血分数:与磁共振成像的比较

Assessment of left ventricular volumes and ejection fraction using ultra-low-dose thallium-201 SPECT on a CZT camera: a comparison with magnetic resonance imaging.
PubMed 文献 · J Nucl Cardiol · 2020-05-14

研究纳入45例因心肌灌注或存活心肌成像转诊的患者,分别用CZT-SPECT超低剂量方案(0.5 Mbq 201-Tl/kg体重)和1.5 T磁共振评估左心室舒张末期容积、收缩末期容积、射血分数和心肌质量。结果显示射血分数、收缩末期容积、舒张末期容积相关性很好,心肌质量相关性中等;与MRI相比,SPECT系统性高估收缩末期容积和心肌质量、低估射血分数(P ≤ .001),舒张末期容积无差异。

要点

  • 超低剂量铊-201 CZT-SPECT与MRI的射血分数、收缩末期容积、舒张末期容积相关性很好,Pearson系数分别为0.86、0.95、0.91。
  • 心肌质量相关性中等,r = 0.57。
  • 与MRI相比,SPECT系统性高估收缩末期容积和心肌质量、低估射血分数(P ≤ .001),舒张末期容积无差异。

核心结论

  • 超低剂量CZT-SPECT评估的左心室容积和射血分数与MRI所得值相关性很好,心肌质量相关性中等。

研究设计与样本

  • 研究纳入45例因心肌灌注或存活心肌成像转诊的患者,分别接受CZT-SPECT和1.5 T磁共振扫描。
  • CZT-SPECT采用超低剂量方案,201-Tl按每公斤体重0.5 Mbq给药。
  • 两种技术均评估收缩末期容积、舒张末期容积、左心室射血分数和心肌质量。

主要发现

  • 射血分数、收缩末期容积、舒张末期容积在CZT-SPECT与心脏磁共振MRI之间相关性很好,Pearson系数分别为0.86、0.95、0.91。
  • 心肌质量相关性中等,r = 0.57。
  • 与MRI相比,SPECT系统性高估收缩末期容积和心肌质量、低估射血分数,P ≤ .001。
  • 舒张末期容积估算在两种方法间无差异。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Cadmium-Zinc-Telluride (CZT) technology allows use of low activities of radiopharmaceuticals. The aim was to verify the values of left ventricular volume parameters, obtained via ultra-low-dose thallium Single Photon Emission Computed Tomography (SPECT) using a CZT camera. METHODS AND RESULTS: Forty-five patients referred for an assessment of myocardial perfusion or viability imaging were examined using CZT-SPECT and 1.5 T magnetic resonance (MRI) scanner. The ultra-low-dose protocol with 0.5 Mbq 201-Tl per kg of body weight was used. The values of end-systolic (ESV) and end-diastolic volumes (EDV), left ventricular ejection fraction (EF) and myocardial mass (MM) were assessed using both techniques. A very good correlation was found between the EF, ESV, and EDV values assessed with CZT-SPECT and cardiac magnetic resonance MRI; the Pearson coefficients were 0.86, 0.95, and 0.91, respectively. A moderate correlation was found for myocardial mass, r = 0.57. Compared to MRI, SPECT systematically overestimated ESV and MM, while it underestimates the EF, with P ≤ .001 in all cases. There was no difference in EDV estimation. CONCLUSIONS: Left ventricular volumes and ejection fraction assessed via ultra-low-dose CZT-SPECT showed very good correlation with the values obtained by MRI. A moderate correlation was found for myocardial mass.

来源:J Nucl Cardiol,查看原文。版权归原作者所有。

引用

  1. Sala M, Kincl V, Kamínek M, et al. Assessment of left ventricular volumes and ejection fraction using ultra-low-dose thallium-201 SPECT on a CZT camera: a comparison with magnetic resonance imaging. J Nucl Cardiol. 2020. PMID: 32410056. doi:10.1007/s12350-020-02161-w

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。