铷-82发生器用于PET灌注成像的产额与效率:两种临床系统的比较

Rubidium-82 generator yield and efficiency for PET perfusion imaging: Comparison of two clinical systems.
PubMed 文献 · J Nucl Cardiol · 2020-05-20

该研究回顾了2年内9台CardioGen-82®和9台RUBY-FILL®发生器的515份每日质量保证报告,比较两种FDA批准的82Sr/82Rb发生器的82Rb产额与生产效率。结果显示CardioGen-82®每日QA报告的82Rb活度低于RUBY-FILL®(2.3±0.2 vs 3.0±0.2 GBq,P<0.001),经统一刻度转换后CardioGen-82®效率仍略低(88±4% vs 95±4%,P<0.001),两系统效率在临床使用期间均改善。

要点

  • CardioGen-82®每日QA报告的82Rb活度低于RUBY-FILL®(2.3±0.2 vs 3.0±0.2 GBq,P<0.001)。
  • 剂量校准器「最大值」低估整合「总」活度约24%(50 mL·min-1条件下)。
  • 统一刻度后CardioGen-82®效率仍略低(88±4% vs 95±4%,P<0.001)。

核心结论

  • 使用CardioGen-82®每日QA程序会显著低估82Rb发生器产额。
  • 以整合总活度表示时,CardioGen-82®系统在整个临床使用期间估计的82Rb生产效率比RUBY-FILL®低约7%。

研究设计与样本

  • 回顾性分析2年内9台CardioGen-82®和9台RUBY-FILL®发生器的515份连续每日质量保证报告。
  • 在RUBY-FILL®系统上以不同流速进行系列测试洗脱,确定经验校正因子,将CardioGen-82®每日QA的82Rb活度值转换为RUBY-FILL®等效值。
  • 测量并比较转换至共同刻度后的发生器产额(82Rb)与生产效率(82Rb产额/82Sr母体活度)。

主要发现

  • 临床使用开始时,每日QA系统报告的82Rb活度CardioGen-82®低于RUBY-FILL®(2.3±0.2 vs 3.0±0.2 GBq,P<0.001),尽管82Sr活度相似。
  • 剂量校准器「最大值」(CardioGen-82®)在50 mL·min-1下低估整合「总」(RUBY-FILL®)活度约24%。
  • 转换为共同测量刻度(整合总活度)后,CardioGen-82®效率平均仍略低于RUBY-FILL®系统(88±4% vs 95±4%,P<0.001)。
  • 两种系统在各自临床使用期间82Rb生产效率均改善。

关键数值与推荐

  • CardioGen-82®每日QA值对应剂量校准器「最大值」:50 mL洗脱、50 mL·min-1。
  • RUBY-FILL®等效值对应整合「总」:35 mL洗脱、20 mL·min-1。
  • 临床使用开始时每日QA报告82Rb活度:CardioGen-82® 2.3±0.2 GBq vs RUBY-FILL® 3.0±0.2 GBq,P<0.001。
  • 转换至共同刻度后生产效率:CardioGen-82® 88±4% vs RUBY-FILL® 95±4%,P<0.001。
  • CardioGen-82®系统整个临床使用期间估计82Rb生产效率比RUBY-FILL®低约7%。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

INTRODUCTION: Strontium-82/Rubidium-82 (82Sr/82Rb) generators are used widely for positron emission tomography (PET) imaging of myocardial perfusion. In this study, the 82Rb isotope yield and production efficiency of two FDA-approved 82Sr/82Rb generators were compared. METHODS: N = 515 sequential daily quality assurance (QA) reports from 9 CardioGen-82® and 9 RUBY-FILL® generators were reviewed over a period of 2 years. A series of test elutions was performed at different flow-rates on the RUBY-FILL® system to determine an empirical correction-factor used to convert CardioGen-82® daily QA values of 82Rb activity (dose-calibrator 'maximum' of 50 mL elution at 50 mL·min-1) to RUBY-FILL® equivalent values (integrated 'total' of 35 mL elution at 20 mL·min-1). The generator yield (82Rb) and production efficiency (82Rb yield/82Sr parent activity) were measured and compared after this conversion to a common scale. RESULTS: At the start of clinical use, the system reported 82Rb activity from daily QA was lower for CardioGen-82® vs RUBY-FILL® (2.3 ± 0.2 vs 3.0 ± 0.2 GBq, P < 0.001) despite having similar 82Sr activity. Dose-calibrator 'maximum' (CardioGen-82®) values were found to under-estimate the integrated 'total' (RUBY-FILL®) activity by ~ 24% at 50 mL·min-1. When these data were used to convert the CardioGen-82 values to a common measurement scale (integrated total activity) the CardioGen-82® efficiency remained slightly lower than the RUBY-FILL® system on average (88 ± 4% vs 95 ± 4%, P < 0.001). The efficiency of 82Rb production improved for both systems over the respective periods of clinical use. CONCLUSIONS: 82Rb generator yield was significantly under-estimated using the CardioGen-82® vs RUBY-FILL® daily QA procedure. When generator yield was expressed as the integrated total activity for both systems, the estimated 82Rb production efficiency of the CardioGen-82® system was ~ 7% lower than RUBY-FILL® over the full period of clinical use.