镉锌碲相机与常规相机心肌灌注显像后的辐射暴露比较
本研究纳入244例疑似冠心病患者,将接受CZT-SPECT心肌灌注显像者按性别、年龄和体质指数1:1匹配至接受Anger-SPECT者,比较90天随访期内MPI及后续所有心脏检查的总辐射暴露。CZT-SPECT组90天总辐射剂量显著低于Anger-SPECT组(6.4±4.8 vs 9.5±4.9 mSv,P<.001);校正混杂因素后CZT-SPECT仍与较低总辐射剂量相关,但该关联在肥胖个体中显著减弱(交互作用P<0.032)。
要点
- CZT-SPECT组90天总辐射剂量为6.4±4.8 mSv,低于Anger-SPECT组的9.5±4.9 mSv(P<.001)。
- 校正混杂因素后CZT-SPECT仍与较低总辐射剂量相关。
- 该获益在肥胖个体中显著减弱(交互作用P<0.032)。
核心结论
- CZT-SPECT与较Anger-SPECT更低的总辐射剂量相关,但这一获益在肥胖个体中可能减弱。
研究设计与样本
- 研究评估244例疑似冠心病患者,接受CZT-SPECT者按性别、年龄和体质指数1:1匹配接受Anger-SPECT心肌灌注显像者。
- 结局为心肌灌注显像研究的辐射暴露加上90天随访期间所有后续心脏检查的辐射暴露。
主要发现
- CZT-SPECT组90天总辐射剂量为6.4±4.8 mSv,Anger-SPECT组为9.5±4.9 mSv,CZT-SPECT组显著低于Anger-SPECT组(P<.001)。
- 校正潜在混杂因素后,CZT-SPECT仍与较低总辐射剂量相关。
- 肥胖个体中该关联显著减弱:体质指数<30者Beta系数为-3.73±0.86,体质指数≥30 Kg/m2者为-2.30±0.92,交互作用P<0.032。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Patient exposure to radiation during the management of coronary heart disease (CHD) can be reduced with more efficient technologies in nuclear medicine, such as the Cadmium-Zinc-Telluride (CZT) gamma-camera for myocardial perfusion imaging (MPI) studies. However, it has been suggested that CZT has lower specificity, which might lead to more downstream radiological procedures, particularly among obese individuals. METHODS AND RESULTS: We evaluated 244 patients with suspected CHD who underwent CZT-SPECT and matched 1:1 according to sex, age, and body mass index (BMI) with those undergoing MPI study with the Anger gamma-camera (Anger-SPECT). The outcome was the total radiation exposure from the MPI study added to the radiation exposure from all subsequent cardiac examinations during a 90-day follow-up. The total radiation dose after 90 days was significantly lower in the CZT-SPECT group (6.4 ± 4.8 vs 9.5±4.9 mSv, P < .001). After adjusting for potential confounders, CZT-SPECT remained associated with lower total radiation dose, but it significantly attenuated among obese individuals (Beta coefficient - 3.73 ± 0.86 for BMI < 30 vs - 2.30 ± 0.92 for BMI ≥ 30 Kg/m2, P for interaction < 0.032). CONCLUSIONS: CZT-SPECT was associated with lower total radiation doses compared to Anger-SPECT, albeit this benefit may be attenuated in obese individuals.
引用
- Baumgarten R, Cerci RJ, de Nadai Costa A, et al. Radiation exposure after myocardial perfusion imaging with Cadmium-Zinc-Telluride camera versus conventional camera. J Nucl Cardiol. 2020. PMID: 32410061. doi:10.1007/s12350-020-02146-9
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。