增强深度学习模型提高PSMA PET-MRI前列腺显像中基于MR的PET衰减校正定量准确性
研究纳入28例前列腺癌患者,用18例数据训练增强生成对抗网络(GANs),在MRI与CT配准中扰动非线性形变场以扩充训练集,并在10例患者验证。结果显示μ-mapGAN+aug在骨盆区域衰减校正μ图误差最低为1.9%,重建PET图像误差最低为2.2%,均优于标准基于MR的方法。
要点
- 增强GANs训练使基于MR的衰减校正μ图精度较标准方法提高4.5%。
- 骨盆区域μ图误差最低为μ-mapGAN+aug的1.9%,低于μ-mapdixon的6.4%。
- 重建PET图像误差最低为PETGAN+aug的2.2%,低于PETdixon的10.3%。
核心结论
- 在训练GAN时扰动MRI与CT配准中的非线性形变场以增强训练数据集,可提高估计μ图的准确性并改善PET定量,优于现有技术。
研究设计与样本
- 研究采集28例前列腺癌患者的CT及对应PET-MR图像,其中18例患者数据(2160层,后增强至270000层)用于训练GANs,其余用于验证。
- 在10例患者上进行定量验证,计算骨与软组织区域衰减校正μ图及重建PET图像的平均相对绝对误差。
主要发现
- 基于深度学习的MR方法生成的伪CT衰减校正μ图精度较标准基于MR的技术高4.5%。
- 骨盆区域衰减校正μ图误差最低为μ-mapGAN+aug的1.9%,对比μ-mapdixon的6.4%、μ-mapdixon+bone的5.9%、μ-mapU-Net的2.1%和μ-mapU-Net+aug的2.0%。
- 重建PET图像误差最低为PETGAN+aug的2.2%,对比PETdixon的10.3%、PETdixon+bone的8.7%、PETU-Net的2.6%和PETU-Net+aug的2.4%。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: Estimation of accurate attenuation maps for whole-body positron emission tomography (PET) imaging in simultaneous PET-MRI systems is a challenging problem as it affects the quantitative nature of the modality. In this study, we aimed to improve the accuracy of estimated attenuation maps from MRI Dixon contrast images by training an augmented generative adversarial network (GANs) in a supervised manner. We augmented the GANs by perturbing the non-linear deformation field during image registration between MRI and the ground truth CT images. METHODS: We acquired the CT and the corresponding PET-MR images for a cohort of 28 prostate cancer patients. Data from 18 patients (2160 slices and later augmented to 270,000 slices) was used for training the GANs and others for validation. We calculated the error in bone and soft tissue regions for the AC μ-maps and the reconstructed PET images. RESULTS: For quantitative analysis, we use the average relative absolute errors and validate the proposed technique on 10 patients. The DL-based MR methods generated the pseudo-CT AC μ-maps with an accuracy of 4.5% more than standard MR-based techniques. Particularly, the proposed method demonstrates improved accuracy in the pelvic regions without affecting the uptake values. The lowest error of the AC μ-map in the pelvic region was 1.9% for μ-mapGAN + aug compared with 6.4% for μ-mapdixon, 5.9% for μ-mapdixon + bone, 2.1% for μ-mapU-Net and 2.0% for μ-mapU-Net + aug. For the reconstructed PET images, the lowest error was 2.2% for PETGAN + aug compared with 10.3% for PETdixon, 8.7% for PETdixon + bone, 2.6% for PETU-Net and 2.4% for PETU-Net + aug.. CONCLUSION: The proposed technique to augment the training datasets for training of the GAN results in improved accuracy of the estimated μ-map and consequently the PET quantification compared to the state of the art.
引用
- Pozaruk A, Pawar K, Li S, et al. Augmented deep learning model for improved quantitative accuracy of MR-based PET attenuation correction in PSMA PET-MRI prostate imaging. Eur J Nucl Med Mol Imaging. 2020. PMID: 32394162. doi:10.1007/s00259-020-04816-9
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。