基于生成对抗网络的深度学习方法实现全身[18F]FDG-PET独立衰减校正
本研究开发并评估了一种基于生成对抗网络(GAN)的深度学习方法,仅利用非衰减校正PET数据生成伪CT用于全身[18F]FDG-PET衰减校正。100组数据用于训练,25例PET/CT用于测试。定性分析显示大多数解剖区域无主要SUV偏差,定量分析显示所有器官平均SUV偏差为-0.8±8.6%,病灶平均偏差为0.9±9.2%。结论认为独立衰减校正是可行的,但需进一步临床验证。
要点
- GAN可基于非衰减校正PET生成伪CT用于全身[18F]FDG-PET衰减校正。
- 测试集25例中,所有器官SUV平均偏差-0.8±8.6%,病灶平均偏差0.9±9.2%。
- 视觉可检测偏差主要位于膈肌和肺边界区域。
核心结论
- 使用所提出的深度学习方法对全身[18F]FDG-PET进行独立衰减校正是可行的,可获得令人满意的PET定量准确性。
- 在临床常规应用前需进一步临床验证。
研究设计与样本
- 研究开发、实现并评估了一种基于生成对抗网络(GAN)的深度学习方法,用于全身[18F]FDG-PET的独立衰减校正。
- 使用100组全身非衰减校正PET数据(PETNAC)及对应CT数据训练GAN以生成伪CT(CTGAN)。
- 使用25例PET/CT检查作为测试数据集(未纳入训练),分别用采集的CT和CTGAN进行衰减校正,得到PETAC和PETGAN。
主要发现
- 定性分析显示,PETGAN在大多数解剖区域无主要SUV偏差,视觉可检测偏差主要位于膈肌和肺边界。
- 定量分析显示,PETGAN所有器官SUV平均百分比偏差为-0.8±8.6%(范围[-30.7%, +27.1%])。
- 病灶SUV平均偏差为0.9±9.2%(范围[-19.6%, +29.2%])。
局限性
- 原文结论指出在临床常规应用前需进一步临床验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Attenuation correction (AC) of PET data is usually performed using a second imaging for the generation of attenuation maps. In certain situations however-when CT- or MR-derived attenuation maps are corrupted or CT acquisition solely for the purpose of AC shall be avoided-it would be of value to have the possibility of obtaining attenuation maps only based on PET information. The purpose of this study was to thus develop, implement, and evaluate a deep learning-based method for whole body [18F]FDG-PET AC which is independent of other imaging modalities for acquiring the attenuation map. METHODS: The proposed method is investigated on whole body [18F]FDG-PET data using a Generative Adversarial Networks (GAN) deep learning framework. It is trained to generate pseudo CT images (CTGAN) based on paired training data of non-attenuation corrected PET data (PETNAC) and corresponding CT data. Generated pseudo CTs are then used for subsequent PET AC. One hundred data sets of whole body PETNAC and corresponding CT were used for training. Twenty-five PET/CT examinations were used as test data sets (not included in training). On these test data sets, AC of PET was performed using the acquired CT as well as CTGAN resulting in the corresponding PET data sets PETAC and PETGAN. CTGAN and PETGAN were evaluated qualitatively by visual inspection and by visual analysis of color-coded difference maps. Quantitative analysis was performed by comparison of organ and lesion SUVs between PETAC and PETGAN. RESULTS: Qualitative analysis revealed no major SUV deviations on PETGAN for most anatomic regions; visually detectable deviations were mainly observed along the diaphragm and the lung border. Quantitative analysis revealed mean percent deviations of SUVs on PETGAN of - 0.8 ± 8.6% over all organs (range [- 30.7%, + 27.1%]). Mean lesion SUVs showed a mean deviation of 0.9 ± 9.2% (range [- 19.6%, + 29.2%]). CONCLUSION: Independent AC of whole body [18F]FDG-PET is feasible using the proposed deep learning approach yielding satisfactory PET quantification accuracy. Further clinical validation is necessary prior to implementation in clinical routine applications.
引用
- Armanious K, Hepp T, Küstner T, et al. Independent attenuation correction of whole body [18F]FDG-PET using a deep learning approach with Generative Adversarial Networks. EJNMMI Res. 2020. PMID: 32449036. doi:10.1186/s13550-020-00644-y
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。