健康年轻志愿者静息/腺苷负荷铷-82 PET/CT心肌血流测量的重测重复性与软件可重复性:运动校正与否的对比
40名健康志愿者完成两次铷-82 PET/CT静息与腺苷负荷扫描,用syngo.MBF、QGS、4DM(含运动校正)分析心肌血流储备(MFR)。各软件MFR重测重复性不一,4DM(无论是否运动校正)测得的MFR高于syngo.MBF和QGS(均P<0.0001),syngo.MBF与QGS一致性高(P=0.83)。
要点
- 各软件MFR重测重复性存在差异,4DM的MFR均值差最小(0.08,P=0.50)。
- 4DM(含或不含运动校正)测得的MFR均高于syngo.MBF和QGS(均P<0.0001)。
- syngo.MBF与QGS之间MFR一致性高(P=0.83)。
核心结论
- 不同软件测得的MFR可重复性存在差异,syngo.MBF与QGS之间MFR一致性最高。
研究设计与样本
- 40名健康志愿者(50%为男性,年龄24±4岁)完成两次铷-82 PET/CT静息与腺苷负荷扫描。
- 使用syngo.MBF(Siemens)、QGS(Cedars-Sinai)和Corridor4DM(4DM)三种软件分析,其中4DM具备运动校正功能。
主要发现
- syngo.MBF的MFR均值差(95%CI)为0.26(-0.03至0.54),P=0.07,一致性限(LoA)为-1.43至1.95。
- QGS的MFR均值差(95%CI)为0.19(-0.08至0.46),P=0.15,LoA为-1.38至1.76。
- 4DM的MFR均值差(95%CI)为0.08(-0.17至0.34),P=0.50,LoA为-1.37至1.53。
- 4DM运动校正的MFR均值差(95%CI)为0.17(-0.17至0.51),P=0.32,LoA为-1.89至2.22。
- 4DM(含或不含运动校正)测得的MFR高于syngo.MBF和QGS(均P<0.0001);syngo.MBF与QGS之间一致性高(P=0.83)。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Myocardial flow reserve (MFR) assessment with cardiac positron emission computed tomography (PET/CT) is well established, and quantification relies on commercial software packages. However, for reliable use, repeatability and reproducibility are important. The aim of this study was therefore to investigate and compare between scans and software packages the repeatability and reproducibility of 82Rb-PET/CT estimated MFR. METHODS AND RESULTS: Forty healthy volunteers completed two 82Rb-PET/CT rest and adenosine stress scans. syngo.MBF (Siemens), quantitative-gated SPECT (QGS) (Cedars-Sinai), and Corridor4DM (4DM) were used for analyses. Motion correction was available for 4DM. Fifty percent were men and age was 24 ± 4 years (mean ± SD). Repeatability of MFR varied between scans. syngo.MBF: mean difference (95% CI) 0.26 (- 0.03 to 0.54), P = 0.07, limits of agreement (LoA): - 1.43 to 1.95; QGS: 0.19 (- 0.08 to 0.46), P = 0.15, LoA: - 1.38 to 1.76; 4DM: 0.08 (- 0.17 to 0.34), P = 0.50, LoA: - 1.37 to 1.53; and 4DM motion corrected: 0.17 (- 0.17 to 0.51), P = 0.32, LoA: - 1.89 to 2.22. MFR was higher using 4DM +/- motion correction compared with syngo.MBF and QGS (all P < 0.0001). Concordance between syngo.MBF and QGS was high (P = 0.83). CONCLUSIONS: Reproducibility of MFR varied for the different software. The highest concordance between MFRs was found between syngo.MBF and QGS.
引用
- Byrne C, Kjaer A, Olsen NE, et al. Test-retest repeatability and software reproducibility of myocardial flow measurements using rest/adenosine stress Rubidium-82 PET/CT with and without motion correction in healthy young volunteers. J Nucl Cardiol. 2020. PMID: 32390111. doi:10.1007/s12350-020-02140-1
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。