文献速递肿瘤 PET/CT

Deep-JASC:使用深度残差网络在全身18F-FDG PET中联合进行衰减和散射校正

Deep-JASC: joint attenuation and scatter correction in whole-body 18F-FDG PET using a deep residual network.
PubMed 文献 · Eur J Nucl Med Mol Imaging · 2020-05-15

本研究构建了二维和三维深度残差网络,输入未校正图像,端到端地直接生成全身18F-FDG PET的衰减和散射校正图像。共纳入1150例临床全身18F-FDG PET/CT研究,其中900、100和150例分别随机分为训练、验证和独立验证集。以CT衰减和散射校正(CTAC)PET图像为参考,所提方法生成的图像质量与CTAC相当,但观察到对噪声的脆弱性,导致伪热/冷斑和/或器官边界定义不清。

要点

  • Deep-JASC可直接从未校正图像生成衰减和散射校正图像,无需CT或MRI解剖图像。
  • 在独立验证集的150例患者中,DL-2DS的体素相对误差为-1.72±4.22%,DL-3DS为3.75±6.91%,DL-3DP为-3.08±5.64%。
  • DL-2DS方法整体性能优于其他两种3D方法,但观察到对噪声的脆弱性,可导致伪热/冷斑和/或器官边界定义不清。

核心结论

  • 本研究证明了使用深度残差网络通过仅发射数据直接对全身18F-FDG PET图像进行衰减和散射校正的可行性。
  • 所提方法无需CT和MRI等解剖图像即可实现准确的衰减和散射校正,该技术适用于独立PET或PET/MRI系统的临床环境。
  • Deep-JASC显示出有前景的定量准确性,但观察到对噪声的脆弱性,导致生成的PET图像中出现伪热/冷斑和/或器官边界定义不清。

研究设计与样本

  • 构建了二维和三维深度残差网络,分别使用2D连续切片(DL-2DS)、3D切片(DL-3DS)和3D图像块(DL-3DP)作为输入,以端到端方式对未校正全身图像进行联合衰减和散射校正。
  • 纳入1150例临床全身18F-FDG PET/CT研究,其中900、100和150例患者分别随机分为训练集、验证集和独立验证集。

主要发现

  • 在独立外部验证集的150例患者中,DL-2DS、DL-3DS和DL-3DP的体素相对误差(%)分别为-1.72±4.22%、3.75±6.91%和-3.08±5.64。
  • 总体而言,DL-2DS方法相比其他两种3D方法表现出更优的性能。
  • Deep-JASC与CTAC图像之间的RMSE值在脑部区域最高,在颈部区域最低。
  • 对于DL-2DS,最大ARE出现在胸部(15.16±3.96%)和肝/肺(11.18±3.23%)区域;DL-3DS和DL-3DP在胸部区域表现略好,ARE分别为11.16±3.42%和11.69±2.71%(p值<0.05)。
  • 联合直方图分析得到DL-2DS、DL-3DS和DL-3DP方法的相关系数分别为0.985、0.980和0.981。

局限性

  • 观察到Deep-JASC对噪声的脆弱性,导致生成的PET图像中出现伪热/冷斑和/或器官边界定义不清。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: We demonstrate the feasibility of direct generation of attenuation and scatter-corrected images from uncorrected images (PET-nonASC) using deep residual networks in whole-body 18F-FDG PET imaging. METHODS: Two- and three-dimensional deep residual networks using 2D successive slices (DL-2DS), 3D slices (DL-3DS) and 3D patches (DL-3DP) as input were constructed to perform joint attenuation and scatter correction on uncorrected whole-body images in an end-to-end fashion. We included 1150 clinical whole-body 18F-FDG PET/CT studies, among which 900, 100 and 150 patients were randomly partitioned into training, validation and independent validation sets, respectively. The images generated by the proposed approach were assessed using various evaluation metrics, including the root-mean-squared-error (RMSE) and absolute relative error (ARE %) using CT-based attenuation and scatter-corrected (CTAC) PET images as reference. PET image quantification variability was also assessed through voxel-wise standardized uptake value (SUV) bias calculation in different regions of the body (head, neck, chest, liver-lung, abdomen and pelvis). RESULTS: Our proposed attenuation and scatter correction (Deep-JASC) algorithm provided good image quality, comparable with those produced by CTAC. Across the 150 patients of the independent external validation set, the voxel-wise REs (%) were - 1.72 ± 4.22%, 3.75 ± 6.91% and - 3.08 ± 5.64 for DL-2DS, DL-3DS and DL-3DP, respectively. Overall, the DL-2DS approach led to superior performance compared with the other two 3D approaches. The brain and neck regions had the highest and lowest RMSE values between Deep-JASC and CTAC images, respectively. However, the largest ARE was observed in the chest (15.16 ± 3.96%) and liver/lung (11.18 ± 3.23%) regions for DL-2DS. DL-3DS and DL-3DP performed slightly better in the chest region, leading to AREs of 11.16 ± 3.42% and 11.69 ± 2.71%, respectively (p value < 0.05). The joint histogram analysis resulted in correlation coefficients of 0.985, 0.980 and 0.981 for DL-2DS, DL-3DS and DL-3DP approaches, respectively. CONCLUSION: This work demonstrated the feasibility of direct attenuation and scatter correction of whole-body 18F-FDG PET images using emission-only data via a deep residual network. The proposed approach achieved accurate attenuation and scatter correction without the need for anatomical images, such as CT and MRI. The technique is applicable in a clinical setting on standalone PET or PET/MRI systems. Nevertheless, Deep-JASC showing promising quantitative accuracy, vulnerability to noise was observed, leading to pseudo hot/cold spots and/or poor organ boundary definition in the resulting PET images.

来源:Eur J Nucl Med Mol Imaging,查看原文。版权归原作者所有。

引用

  1. Shiri I, Arabi H, Geramifar P, et al. Deep-JASC: joint attenuation and scatter correction in whole-body 18F-FDG PET using a deep residual network. Eur J Nucl Med Mol Imaging. 2020. PMID: 32415552. doi:10.1007/s00259-020-04852-5

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。