心脏SPECT影像组学特征的可重复性与再现性:多扫描仪体模研究

Cardiac SPECT radiomic features repeatability and reproducibility: A multi-scanner phantom study.
PubMed 文献 · J Nucl Cardiol · 2020-04-24

本研究用4台商用SPECT及SPECT/CT相机对静态心脏体模(185 MBq 99mTc)采集图像,变化采集与重建参数共重建5,063个横断面,提取87个影像组学特征,用变异系数评估可重复性与再现性。结果显示可重复性因特征而异,仅少数特征在所有成像设置下COV≤5%,矩阵大小对特征变异影响最大(82.8%特征COV>20%)。

要点

  • 仅IDMN、IDN、RP、ZE、DE五个特征在所有成像设置变化下COV≤5%。
  • LALGLE、SALGLE、LGLZE、SDLGLE对成像设置变化可重复性较差(COV>20%)。
  • 矩阵大小对特征变异影响最大,改变后82.8%的特征COV>20%。

核心结论

  • 不同成像设置下SPECT/CT心脏影像组学特征的可重复性与再现性因特征而异。
  • 不同图像采集与重建方案对影像组学特征的影响各不相同。
  • COV低的影像组学特征可作为未来临床研究的候选特征。

研究设计与样本

  • 使用4台商用SPECT及SPECT/CT相机对模拟典型心肌灌注成像的静态心脏体模采集图像,使用185 MBq的99mTc。
  • 研究不同图像采集与重建参数的影响,包括视图数、视图矩阵大小、衰减校正以及图像重建相关参数(算法、迭代次数、子集数、重建后滤波器类型及其滤波阶数和截止频率)。
  • 通过变化上述因素共重建5,063个横断面,并从这些图像中提取87个影像组学特征,包括一阶、二阶和高阶纹理。
  • 对每个影像组学特征在不同成像设置下测量变异系数(COV),以评估再现性和可重复性。

主要发现

  • 来自灰度共生矩阵(GLCM)的逆差矩归一化(IDMN)和逆差归一化(IDN)、来自灰度游程矩阵(GLRLM)的游程百分比(RP)、来自灰度区域大小矩阵(GLSZM)的区域熵(ZE)以及来自灰度依赖矩阵(GLDM)的依赖熵(DE)是仅有的在所有成像设置变化下表现出高再现性(COV≤5%)的特征。
  • 来自GLSZM的大面积低灰度强调(LALGLE)、小面积低灰度强调(SALGLE)和低灰度区域强调(LGLZE),以及来自GLDM的小依赖低灰度强调(SDLGLE)对成像设置变化再现性较差(COV>20%)。
  • GLRLM特征集具有最高比例的可重复特征(COV<5%,31.88%),GLDM特征集具有最低比例的可重复特征(COV>20%,54.2%)。
  • 矩阵大小对特征变异影响最大,改变矩阵大小时大多数特征不可重复,其中82.8%的特征COV>20%。

关键数值与推荐

  • 高再现性判定标准为COV≤5%,较差再现性判定标准为COV>20%。
  • GLRLM特征集中COV<5%的特征占31.88%,GLDM特征集中COV>20%的特征占54.2%。
  • 矩阵大小改变时82.8%的特征COV>20%。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: The aim of this work was to assess the robustness of cardiac SPECT radiomic features against changes in imaging settings, including acquisition, and reconstruction parameters. METHODS: Four commercial SPECT and SPECT/CT cameras were used to acquire images of a static cardiac phantom mimicking typical myorcardial perfusion imaging using 185 MBq of 99mTc. The effects of different image acquisition and reconstruction parameters, including number of views, view matrix size, attenuation correction, as well as image reconstruction related parameters (algorithm, number of iterations, number of subsets, type of post-reconstruction filter, and its associated parameters, including filter order and cut-off frequency) were studied. In total, 5,063 transverse views were reconstructed by varying the aforementioned factors. Eighty-seven radiomic features including first-, second-, and high-order textures were extracted from these images. To assess reproducibility and repeatability, the coefficient of variation (COV), as a widely adopted metric, was measured for each of the radiomic features over the different imaging settings. RESULTS: The Inverse Difference Moment Normalized (IDMN) and Inverse Difference Normalized (IDN) features from the Gray Level Co-occurrence Matrix (GLCM), Run Percentage (RP) from the Gray Level Co-occurrence Matrix (GLRLM), Zone Entropy (ZE) from the Gray Level Size Zone Matrix (GLSZM), and Dependence Entropy (DE) from the Gray Level Dependence Matrix (GLDM) feature sets were the only features that exhibited high reproducibility (COV ≤ 5%) against changes in all imaging settings. In addition, Large Area Low Gray Level Emphasis (LALGLE), Small Area Low Gray Level Emphasis (SALGLE) and Low Gray Level Zone Emphasis (LGLZE) from GLSZM, and Small Dependence Low Gray Level Emphasis (SDLGLE) from GLDM feature sets turned out to be less reproducible (COV > 20%) against changes in imaging settings. The GLRLM (31.88%) and GLDM feature set (54.2%) had the highest (COV < 5%) and lowest (COV > 20%) number of the reproducible features, respectively. Matrix size had the largest impact on feature variability as most of the features were not repeatable when matrix size was modified with 82.8% of them having a COV > 20%. CONCLUSION: The repeatability and reproducibility of SPECT/CT cardiac radiomic features under different imaging settings is feature-dependent. Different image acquisition and reconstruction protocols have variable effects on radiomic features. The radiomic features exhibiting low COV are potential candidates for future clinical studies.