文献速递心脏核医学

使用碘-123间碘苄胍在IQ-SPECT生成的平面图像上计算心纵隔比的效用

The utility of heart-to-mediastinum ratio using a planar image created from IQ-SPECT with Iodine-123 meta-iodobenzylguanidine.
PubMed 文献 · J Nucl Cardiol · 2020-02-26

研究旨在验证从IQ-SPECT冠状面图像经射线求和处理生成平面等效图像(IQ-planar)计算心纵隔比(HMR)的效用。结果表明,未交叉校准的IQ-planar HMR在无衰减散射校正(NC)和衰减散射校正(ACSC)条件下均显著高于常规平面图像,但交叉校准后与平面图像计算的标准HMR相似,提示该方法可与常规平面HMR联合使用。

要点

  • IQ-planar图像通过叠加IQ-SPECT冠状面所有层面并经射线求和处理生成。
  • 未校准的IQ-planar HMR在NC和ACSC条件下均显著高于常规平面图像。
  • 交叉校准后的IQ-planar HMR与平面图像计算的标准HMR相似。

核心结论

  • 使用新射线求和处理方法得到的IQ-planar HMR可与常规平面HMR联合使用。

研究设计与样本

  • 获取IQ-SPECT的平面图像和横断面图像,包括衰减散射校正(ACSC)和无衰减散射校正(NC)条件。
  • 对IQ-SPECT冠状面图像应用多平面重建和射线求和处理,生成NC和ACSC条件下的IQ-planar图像。
  • 利用平面图像测量的HMR与数学计算HMR之间的线性回归校准HMR至标准化值。

主要发现

  • 交叉校准前后,平面与IQ-planar HMR的散点图和线性回归线在NC和ACSC条件下均显示系统性差异。
  • NC和ACSC条件下,IQ-planar HMR均显著高于常规平面图像。
  • 交叉校准后,NC和ACSC条件下的IQ-planar HMR与平面图像计算的标准HMR相似。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIMS: 123I-labeled meta-iodobenzylguanidine (MIBG) has used a planar image to measure the heart-to-mediastinum ratio (HMR). However, planar images are not available from IQ-SPECT with SMARTZOOM collimator due to its multi-focal collimation. Since we created the planar-equivalent (IQ-planar) images by adding all slices of the IQ-SPECT coronal image. The aim of this study was to demonstrate the utility of the new method for calculating HMR. METHODS: The planar image and transverse images of IQ-SPECT with attenuation and scatter corrections (ACSC) and without ACSC (NC) were obtained. Multi-planar reconstruction and ray-summation processing were applied to create IQ-planar images with NC and ACSC. Linear regression between the measured HMR from the planar image and the mathematically calculated HMR was used to calibrate HMR to standardized values. RESULTS: Scatterplots and linear regression lines between planar and IQ-planar HMRs before and after cross-calibration showed systematic differences in both NC and ACSC conditions. The IQ-planar HMR with NC and ACSC was significantly higher compared with that of the conventional planar image. However, the IQ-planar HMR with NC and ACSC after cross-calibration was similar to the standardized HMR calculated by planar image. CONCLUSION: The IQ-planar HMR using the new ray-summation processing method could be used along with the conventional planar HMR.

来源:J Nucl Cardiol,查看原文。版权归原作者所有。

引用

  1. Shibutani T, Nakajima K, Yoneyama H, et al. The utility of heart-to-mediastinum ratio using a planar image created from IQ-SPECT with Iodine-123 meta-iodobenzylguanidine. J Nucl Cardiol. 2020. PMID: 32103408. doi:10.1007/s12350-020-02081-9

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。