衰减校正在CZT-SPECT测量心肌血流量中的影响
研究纳入104例患者,在CZT-SPECT上按一日法行负荷/静息Tc-99m-tetrofosmin采集,比较衰减校正(AC)与无衰减校正(NAC)对心肌血流量(MBF)和心肌血流储备(MFR)的影响。AC后负荷与静息MBF均显著降低(P<0.001),但区域及整体MFR在AC与NAC间无显著差异(P>0.25)。
要点
- AC后负荷与静息MBF均显著低于NAC(P<0.001)。
- 区域及整体MFR在AC与NAC间无显著差异(P>0.25)。
- 以2为阈值,86例(83%)患者整体LV MFR分类在AC与否时保持一致。
核心结论
- AC校正对区域及整体左心室MFR测量均无显著影响。
研究设计与样本
- 研究纳入104例患者,在碲锌镉(CZT)针孔心脏相机上以列表模式采集SPECT数据,采用负荷(250±17 MBq)/静息(511±23 MBq)一日法Tc-99m-tetrofosmin方案。
- 低剂量CT在另一台SPECT/CT相机上以相同体位采集,所有分析使用Corridor4DM完成。
主要发现
- 应用AC时负荷与静息MBF均显著降低(P<0.001)。
- 区域及整体MFR在AC与NAC测量间无显著差异(P>0.25)。
- 整体左心室MFR均值在NAC为2.43±0.87、AC为2.33±0.89。
- 以2为阈值,86例(83%)患者整体左心室MFR分类在是否应用AC时均保持正常或异常不变。
- 其余18例患者NAC与AC值的平均差异为0.3。
关键数值与推荐
- 整体左心室MFR阈值取2时,86例(83%)患者分类不变。
- 其余18例患者NAC与AC值的平均差异为0.3。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: Most of cardiac dedicated CZT-SPECT systems are not equipped with CT, whereas PET systems are. We evaluated the impact of AC correction on CZT-SPECT myocardial blood flow (MBF) and myocardial flow reserve (MFR) measurements. METHODS: 104 patients were included. SPECT data were acquired on cadmium zinc telluride (CZT)-based pinhole cardiac camera in listmode using a stress (250 ± 17 MBq)/rest (511 ± 23 MBq) 1-day Tc-99m-tetrofosmin protocol. Low-dose CT was acquired on another SPECT/CT camera in the same position. All analysis was performed using Corridor4DM. RESULTS: Stress and rest MBF were significantly lower when AC was applied (P < 0.001). For regional and global MFR, there was no significant difference between AC and NAC measurements (P > 0.25 at least). Mean global LV MFR was 2.43 ± 0.87 and 2.33 ± 0.89, respectively, for NAC and AC measurements. Using a threshold of 2, 86 patients (83%) remained classified as normal and abnormal regarding global LV MFR whether AC was applied or not. Mean difference between NAC and AC values for the 18 other patients was 0.3. CONCLUSION: AC correction does not significantly affect MFR measurement both in regional and global LV analyses.
引用
- Bailly M, Thibault F, Courtehoux M, et al. Impact of attenuation correction for CZT-SPECT measurement of myocardial blood flow. J Nucl Cardiol. 2020. PMID: 32080802. doi:10.1007/s12350-020-02075-7
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。