文献速递肿瘤 PET/CT放射性药物与辐射防护

全身连续床运动PET数据驱动呼吸门控算法的临床评估

Clinical Evaluation of a Data-Driven Respiratory Gating Algorithm for Whole-Body PET with Continuous Bed Motion.
PubMed 文献 · J Nucl Med · 2020-02-14

研究纳入56例疑似胸部或腹部恶性肿瘤患者,行18F-FDG全身连续床运动PET/CT,比较数据驱动门控(DDG)与压力带门控(BG)的信号相关性,并评估最优门控(OG)和弹性运动校正(EMOCO)重建。DDG与BG信号全局相关系数为0.48±0.11,在肾脏和肝脏区域峰值达0.89±0.07。运动校正重建的SUV测量值和病灶勾画质量均优于静态图像,DDG与BG之间无显著差异。

要点

  • DDG与BG信号全局相关系数0.48±0.11,肾脏和肝脏区域峰值0.89±0.07。
  • 运动校正重建的SUVmax中位数显著高于静态图像(P<0.001)。
  • BG与DDG在OG和EMOCO重建之间SUV及视觉勾画均无显著差异。

核心结论

  • 基于DDG的连续床运动PET运动补偿优于静态重建,质量与BG方法相当。
  • 新算法可能成为连续床运动PET系统的有价值替代方案。

研究设计与样本

  • 单中心临床队列研究,纳入56例疑似胸部或腹部恶性肿瘤患者。
  • 患者接受全身18F-FDG连续床运动PET/CT,同时使用DDG和BG。
  • 对两种门控信号进行相关性分析,并比较静态、OG和EMOCO重建。
  • 由3位专家独立评估不同重建的病灶勾画质量。

主要发现

  • BG与DDG信号全局相关系数为0.48±0.11,扫描肾脏和肝脏区域时峰值达0.89±0.07。
  • 共分析196个病灶。
  • BG-OG、DDG-OG、BG-EMOCO和DDG-EMOCO的SUV测量值均显著高于静态图像(P<0.001);SUVmax中位数:静态14.3±13.4,BG-EMOCO 19.8±15.7,DDG-EMOCO 20.5±15.6,BG-OG 19.6±17.1,DDG-OG 18.9±16.6。
  • BG-OG与DDG-OG之间、BG-EMOCO与DDG-EMOCO之间SUV均无显著差异。
  • BG-EMOCO和DDG-EMOCO的视觉病灶勾画显著优于静态重建(P<0.001);在EMOCO或OG重建中比较BG与DDG均无显著差异。

临床怎么用

  • 对于连续床运动PET系统,DDG运动补偿可获得与BG相当的重建质量,可能作为有价值的替代方案。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Respiratory gating is the standard to prevent respiration effects from degrading image quality in PET. Data-driven gating (DDG) using signals derived from PET raw data is a promising alternative to gating approaches requiring additional hardware (e.g., pressure-sensitive belt gating [BG]). However, continuous-bed-motion (CBM) scans require dedicated DDG approaches for axially extended PET, compared with DDG for conventional step-and-shoot scans. In this study, a CBM-capable DDG algorithm was investigated in a clinical cohort and compared with BG using optimally gated (OG) and fully motion-corrected (elastic motion correction [EMOCO]) reconstructions. Methods: Fifty-six patients with suspected malignancies in the thorax or abdomen underwent whole-body 18F-FDG CBM PET/CT using DDG and BG. Correlation analyses were performed on both gating signals. Besides static reconstructions, OG and EMOCO reconstructions were used for BG and DDG. The metabolic volume, SUVmax, and SUVmean of lesions were compared among the reconstructions. Additionally, the quality of lesion delineation in the different PET reconstructions was independently evaluated by 3 experts. Results: The global correlation coefficient between BG and DDG signals was 0.48 ± 0.11, peaking at 0.89 ± 0.07 when scanning the kidney and liver region. In total, 196 lesions were analyzed. SUV measurements were significantly higher in BG-OG, DDG-OG, BG-EMOCO, and DDG-EMOCO than in static images (P < 0.001; median SUVmax: static, 14.3 ± 13.4; BG-EMOCO, 19.8 ± 15.7; DDG-EMOCO, 20.5 ± 15.6; BG-OG, 19.6 ± 17.1; and DDG-OG, 18.9 ± 16.6). No significant differences between BG-OG and DDG-OG or between BG-EMOCO and DDG-EMOCO were found. Visual lesion delineation was significantly better in BG-EMOCO and DDG-EMOCO than in static reconstructions (P < 0.001); no significant difference was found when comparing BG and DDG for either EMOCO or OG reconstruction. Conclusion: DDG-based motion compensation of CBM PET acquisitions outperforms static reconstructions, delivering qualities comparable to BG approaches. The new algorithm may be a valuable alternative for CBM PET systems.