全身PET/CT中CT呼吸运动时PET与CT图像配准的改善
研究在28例癌症患者的40个肺部或上腹部≤3 cm病灶中采集CT与PET呼吸波形,比较PET/static、PET/ex与PET/matched三种重建方法的配准。视觉判读良好配准率分别为13/21、13/23、18/21,平均PET/CT距离分别为3.5、5.1、2.8 mm。结论认为在PET和CT中均采集呼吸波形有望改善PET与CT图像配准。
要点
- PET/matched视觉良好配准率为18/21,平均PET/CT距离2.8 mm,优于PET/static的13/21和3.5 mm。
- 使用PET/ex或PET/matched时SUVmax平均较PET/static高21%,SUVpeak高12%。
- 肝脏图像噪声PET/ex较PET/static高15%、PET/matched高40%,仍远低于门控PET。
核心结论
- 在PET和CT中均采集呼吸波形有望改善PET与CT图像配准,所提出的方法经测试显示配准改善。
研究设计与样本
- 在28例癌症患者的28次PET/CT扫描中采集CT与PET呼吸波形,涉及肺部或上腹部40个≤3 cm的病灶。
- PET列表模式数据采用PET/static、PET/ex和PET/matched三种方法重建并比较。
主要发现
- 视觉判读良好配准率PET/static为13/21、PET/ex为13/23、PET/matched为18/21,平均PET/CT距离分别为3.5、5.1、2.8 mm。
- 与PET/ex视觉比较,良好配准率在PET/static中10例有1例增加,在PET/matched中10例有7例增加。
- 使用PET/ex或PET/matched时SUVmax平均较PET/static高21%,SUVpeak高12%。
- 肝脏图像噪声PET/ex较PET/static高15%,PET/matched较PET/static高40%,即噪声远低于门控PET。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Respiratory motion during the CT and PET parts of a PET/CT scan leads to imperfect alignment of anatomic features seen by the 2 modalities. In this work, we concentrate on the effects of motion during CT. We propose a novel approach for improving the alignment. Methods: Respiratory waveform data were gathered during the CT and PET parts of 28 PET/CT scans of cancer patients with 40 lesions up to 3 cm in size in the lung or upper abdomen. PET list-mode data were reconstructed by 3 reconstruction methods: PET/static (the standard method with no motion correction); PET/ex (a method that calculates a range of expiratory amplitudes from the lowest one to the highest one); and PET/matched (a novel method that uses both waveforms). The 3 methods were compared. The distance between tumor positions in PET and CT were characterized in visual interpretation by physicians as well as quantitatively. Tumor SUVs (SUVmax and SUVpeak) were determined relative to SUV based on the static method. Image noise was evaluated in the liver and compared with PET/static. Results: In visual interpretation, the rate of good alignment was 13 of 21, 13 of 23, and 18 of 21 for the PET/static, PET/ex, and PET/matched methods, respectively, and the mean PET/CT distances were 3.5, 5.1, and 2.8 mm. In visual comparison with PET/ex, the rate of good alignment was increased in 1 of 10 and 7 of 10 cases for PET/static and PET/matched, respectively. SUVmax was on average 21% higher than PET/static when either PET/ex or PET/matched was used. SUVpeak was 12% higher. Image noise in the liver was 15% higher than PET/static for the PET/ex method, and 40% higher for PET/matched; that is, noise was much lower than in gated PET. Conclusion: Acquiring respiratory waveforms both in PET (as in the current state of the art) and in CT (an unusual key step in this approach) has the potential to improve the alignment of PET and CT images. A proposed method for using this information was tested. Improved alignment was demonstrated.