一种用于动态SPECT的新型心脏体模
研究开发了一种能模拟左心室腔和心肌放射性示踪剂浓度随时间变化并具有搏动样运动的机械心脏体模,同时建立数学模型描述两区域浓度随时间函数并计算体模理论心肌血流量。在GE Discovery NM 530c SPECT系统上进行体模实验,用已知注入放射性活度验证SPECT报告的心肌血流量值,从而确定该系统的准确性。
要点
- 新型机械心脏体模可模拟左心室腔和心肌的生理性放射性示踪剂变化及搏动样运动。
- 体模数学模型可描述左心室和心肌浓度随时间函数,用于实验规划和模拟生理样时间-活性曲线。
- 在GE Discovery NM 530c SPECT系统上,用已知注入放射性活度验证了SPECT报告的心肌血流量值。
核心结论
- 开发了一种用于发射断层成像的新型动态心脏体模,能产生模拟生理性曲线(未来可能模拟病理性曲线)的广泛时间-活性曲线。
- 体模实验用已知注入放射性活度验证了SPECT报告的心肌血流量值,从而确定GE Discovery 530c SPECT系统的准确性。
研究设计与样本
- 方法学研究:开发机械心脏体模并建立其数学模型及净滞留模型,在GE Discovery NM 530c SPECT系统上进行体模实验。
主要发现
- 体模能模拟左心室腔和心肌的生理性放射性示踪剂变化,同时进行搏动样运动。
- 数学模型可描述左心室和心肌的放射性示踪剂浓度随时间函数,用于实验规划和准确模拟生理样时间-活性曲线。
- 净滞留模型仅从测量量计算体模的理论(即预期)心肌血流量,用于验证SPECT系统报告的心肌血流量。
- 体模实验用已知(测量的)注入放射性示踪剂活度验证了SPECT报告的心肌血流量值。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: In recent years, with the advance of myocardial blood flow (MBF) measurement capability in dynamic single photon emission computerized tomography (SPECT) systems, significant effort has been devoted to validation of the new capability. Unfortunately, the mechanical phantoms available for the validation process lack essential features-they either have a constant radiotracer concentration or they have rigid (static) walls unable to simulate cardiac beating. METHODS AND RESULTS: We have developed a mechanical cardiac phantom that is able to mimic physiological radiotracer variation in the left ventricle (LV) cavity and in the myocardium (M), while performing beating-like motion. We have also developed a mathematical model of the phantom, allowing a description of the radiotracer concentrations in both regions (LV, M) as a function of time, which served as a tool for experiment planning and to accurately mimic physiological-like time-activity curves (TACs). A net retention model for the phantom was also developed, which served to compute the theoretical (i.e., expected) MBF of the phantom from measured quantities only, and thus validate the MBF reported by the SPECT system. In this paper, phantom experiments were performed on a GE Discovery NM 530c SPECT system. CONCLUSIONS: A novel dynamic cardiac phantom for emission tomography has been developed. The new phantom is capable of producing a wide range of TACs that can mimic physiological (and potentially in the future, pathological) curves, similar to those observed in dynamic SPECT systems. SPECT-reported MBF values were validated against known (measured) activity of the injected radiotracer from phantom experiments, which allowed to determine the accuracy of the GE Discovery 530c SPECT system.
引用
- Krakovich A, Zaretsky U, Moalem I, et al. A new cardiac phantom for dynamic SPECT. J Nucl Cardiol. 2020. PMID: 31997101. doi:10.1007/s12350-020-02028-0
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。