EARL认证常规与数字PET的SUV变异性研究
该前瞻性研究在50例疑似或确诊癌症患者的全身FDG-PET扫描中,头对头比较EARL认证的常规PET(cPET)与数字PET(dPET)的SUV变异性,共分析128个病灶。dPET的SUVmean、SUVpeak、SUVmax均高于cPET,平均ΔSUV为6-8%,仅少数病灶SUV差异超过30%,提示EARL标准化有效。
要点
- EARL认证的常规与数字PET之间平均SUV差异为6-8%,dPET值更高。
- ΔSUVmax(8%)显著高于ΔSUVmean(6%)和ΔSUVpeak(7%)。
- 个体病灶重复性系数为27%(ΔSUVmean和ΔSUVpeak)和33%(ΔSUVmax)。
核心结论
- EARL认证的常规与数字PET之间SUV变异性有限,平均差异不超过8%,仅少数病灶SUV差异超过30%,表明EARL标准化有效。
研究设计与样本
- 前瞻性头对头比较研究,纳入50例疑似或确诊癌症患者,随机分配PET扫描顺序,应用EARL兼容重建,每例最多测量5个FDG阳性病灶的SUVmean、SUVpeak、SUVmax及病灶直径。
主要发现
- 共纳入128个病灶,平均大小19±14 mm。
- dPET的SUVmean、SUVpeak、SUVmax均高于cPET,平均ΔSUV为6-8%(p < 0.001)。
- ΔSUVmax显著高于ΔSUVmean(8% vs. 6%,p = 0.002)和ΔSUVpeak(8% vs. 7%,p = 0.03)。
- 个体病灶重复性系数为27%(ΔSUVmean和ΔSUVpeak)和33%(ΔSUVmax)(p < 0.001)。
- 仅有限数量病灶显示SUV差异超过30%。
关键数值与推荐
- EARL兼容重建应用于两种扫描仪。
- 平均ΔSUV为6-8%,dPET值更高(p < 0.001)。
- 重复性系数:ΔSUVmean和ΔSUVpeak为27%,ΔSUVmax为33%。
临床怎么用
- 在EARL认证的常规与数字PET之间进行SUV比较时,应考虑平均6-8%的差异及个体病灶可达27-33%的重复性系数。
- EARL标准化可支持多中心PET SUV的可重复性,但需注意SUVmax变异性高于SUVmean和SUVpeak。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: A high SUV-reproducibility is crucial when different PET scanners are in use. We evaluated the SUV variability in whole-body FDG-PET scans of patients with suspected or proven cancer using an EARL-accredited conventional and digital PET scanner. In a head-to-head comparison we studied images of 50 patients acquired on a conventional scanner (cPET, Ingenuity TF PET/CT, Philips) and compared them with images acquired on a digital scanner (dPET, Vereos PET/CT, Philips). The PET scanning order was randomised and EARL-compatible reconstructions were applied. We measured SUVmean, SUVpeak, SUVmax and lesion diameter in up to 5 FDG-positive lesions per patient. The relative difference ΔSUV between cPET and dPET was calculated for each SUV-parameter. Furthermore, we calculated repeatability coefficients, reflecting the 95% confidence interval of ΔSUV. RESULTS: We included 128 lesions with an average size of 19 ± 14 mm. Average ΔSUVs were 6-8% with dPET values being higher for all three SUV-parameters (p < 0.001). ΔSUVmax was significantly higher than ΔSUVmean (8% vs. 6%, p = 0.002) and than ΔSUVpeak (8% vs. 7%, p = 0.03). Repeatability coefficients across individual lesions were 27% (ΔSUVmean and ΔSUVpeak) and 33% (ΔSUVmax) (p < 0.001). CONCLUSIONS: With EARL-accredited conventional and digital PET, we found a limited SUV variability with average differences up to 8%. Furthermore, only a limited number of lesions showed a SUV difference of more than 30%. These findings indicate that EARL standardisation works. TRIAL REGISTRATION: This prospective study was registered on the 31th of October 2017 at ClinicalTrials.cov. URL: https://clinicaltrials.gov/ct2/show/NCT03457506?id=03457506&rank=1.
引用
- Koopman D, Jager PL, Slump CH, et al. SUV variability in EARL-accredited conventional and digital PET. EJNMMI Res. 2019. PMID: 31823097. doi:10.1186/s13550-019-0569-7
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。