心血管18F-氟化物PET/MR与PET/CT对比研究
该研究纳入18例主动脉瓣狭窄或近期心肌梗死患者,比较18F-氟化物PET/MR与PET/CT在主动脉瓣和冠状动脉摄取评估中的一致性。结果显示径向GRE衰减校正下TBRMAX与PET/CT相当,但Dixon衰减校正表现较差,且支架伪影限制PET/MR临床应用。
要点
- 径向GRE衰减校正的PET/MR与PET/CT的主动脉瓣TBRMAX相当。
- Dixon衰减校正的PET/MR TBRMAX低于PET/CT且一致性范围较宽。
- 支架伪影使PET/MR评估冠状动脉18F-氟化物摄取具有挑战性。
核心结论
- 心血管PET/MR与PET/CT在18F-氟化物摄取评估中显示良好的视觉和定量一致性。
- PET/MR受支架相关伪影影响,目前限制其临床应用。
研究设计与样本
- 18例主动脉瓣狭窄或近期心肌梗死患者依次接受18F-氟化物PET/CT和PET/MR。
- 独立评估主动脉瓣和冠状动脉18F-氟化物摄取,比较标准Dixon与新型径向GRE衰减校正图与PET/CT的一致性。
主要发现
- 视觉上主动脉瓣18F-氟化物摄取在PET/CT与PET/MR上相似。
- 径向GRE衰减校正的TBRMAX与PET/CT相当(PET/CT 1.55±0.33 vs. PET/MR 1.58±0.34,P=0.66;95%一致性界限-27%至+25%)。
- Dixon衰减校正的TBRMAX表现较差(1.38±0.44,P=0.06;偏倚(-)14%;95%一致性界限-25%至+53%)。
- 在自体冠状动脉中,无论采用何种衰减校正方法,PET/MR与PET/CT的18F-氟化物摄取相似。
- PET/MR识别出PET/CT所识别的29个斑块中的28个,但支架在PET/MR上产生伪影使18F-氟化物摄取评估困难。
临床怎么用
- 在无支架的情况下,径向GRE衰减校正的PET/MR可用于评估主动脉瓣和冠状动脉的18F-氟化物摄取。
- 存在冠状动脉支架时,PET/MR评估18F-氟化物摄取受限,需谨慎解读。
局限性
- 样本量较小(18例)。
- 支架相关伪影目前限制PET/MR的临床应用。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: 18F-Fluoride uptake denotes calcification activity in aortic stenosis and atherosclerosis. While PET/MR has several advantages over PET/CT, attenuation correction of PET/MR data is challenging, limiting cardiovascular application. We compared PET/MR and PET/CT assessments of 18F-fluoride uptake in the aortic valve and coronary arteries. METHODS AND RESULTS: 18 patients with aortic stenosis or recent myocardial infarction underwent 18F-fluoride PET/CT followed immediately by PET/MR. Valve and coronary 18F-fluoride uptake were evaluated independently. Both standard (Dixon) and novel radial GRE) MR attenuation correction (AC) maps were validated against PET/CT with results expressed as tissue-to-background ratios (TBRs). Visually, aortic valve 18F-fluoride uptake was similar on PET/CT and PET/MR. TBRMAX values were comparable with radial GRE AC (PET/CT 1.55±0.33 vs. PET/MR 1.58 ± 0.34, P = 0.66; 95% limits of agreement - 27% to + 25%) but performed less well with Dixon AC (1.38 ± 0.44, P = 0.06; bias (-)14%; 95% limits of agreement - 25% to + 53%). In native coronaries, 18F-fluoride uptake was similar on PET/MR to PET/CT regardless of AC approach. PET/MR identified 28/29 plaques identified on PET/CT; however, stents caused artifact on PET/MR making assessment of 18F-fluoride uptake challenging. CONCLUSION: Cardiovascular PET/MR demonstrates good visual and quantitative agreement with PET/CT. However, PET/MR is hampered by stent-related artifacts currently limiting clinical application.
引用
- Andrews JPM, MacNaught G, Moss AJ, et al. Cardiovascular 18F-fluoride positron emission tomography-magnetic resonance imaging: A comparison study. J Nucl Cardiol. 2019. PMID: 31792913. doi:10.1007/s12350-019-01962-y
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。