N13NH3-PET与99mTc-Tetrofosmin-CZT SPECT评估绝对心肌血流量及血流储备的比较
该研究在54例患者中比较了99mTc-tetrofosmin-CZT SPECT与N13NH3 PET/CT在瑞加诺生诱导充血状态下测量心肌血流量(MBF)和血流储备(MFR)的表现,并评估衰减校正(AC)的影响。结果显示PET与AC-CZT的MBF和MFR相关性合理,无AC的CZT高估MBF,建议使用CZT时进行AC。
要点
- AC-CZT与PET的静息MBF一致,但负荷MBF和MFR PET更高。
- NAC-CZT显著高估静息和负荷MBF,且与PET相关性较差。
- AC-CZT能预测MFR<2和负荷MBF<1.7,并可区分正常与冠心病患者。
核心结论
- N13NH3-PET与99mTc-Tetrofosmin-CZT SPECT测量的MBF和MFR具有合理的相关性。
- 无衰减校正的CZT会高估MBF,使用CZT测量MBF时推荐进行衰减校正。
研究设计与样本
- 研究纳入54例患者,在2周内分别于静息和血管扩张状态下接受99mTc-tetrofosmin-CZT SPECT和N13NH3 PET/CT检查。
- CZT测量MBF和MFR时分别在有衰减校正(AC)和无衰减校正(NAC)条件下进行。
主要发现
- 整体静息MBF:PET为0.76 ± 0.19 mL/min/gr,AC-CZT为0.76 ± 0.24(P = NS),NAC-CZT为1.14 ± 0.4(P < 0.001 vs PET和AC-CZT)。
- 负荷MBF:PET为1.87 ± 0.45 mL/min/gr,高于AC-CZT的1.62 ± 0.68(P < 0.0008),但低于NAC-CZT的2.36 ± 1.1(P < 0.0003)。
- MFR:PET为2.52 ± 0.56,高于AC-CZT的2.22 ± 1(P < 0.009)和NAC-CZT的2.18 ± 1.0(P < 0.004)。
- 线性回归显示PET与AC-CZT在MFR和负荷MBF上的相关性优于PET与NAC-CZT。
- ROC分析显示AC-CZT预测MFR < 2的AUC为0.75,预测负荷MBF < 1.7的AUC为0.82;区分正常与CAD患者的AUC分别为0.747(MFR)和0.892(负荷MBF)。
关键数值与推荐
- AC-CZT预测MFR < 2的AUC = 0.75,预测负荷MBF < 1.7的AUC = 0.82。
- AC-CZT区分正常与CAD患者的AUC:MFR为0.747,负荷MBF为0.892。
临床怎么用
- 使用CZT SPECT定量MBF和MFR时应进行衰减校正,以避免无校正时的高估。
- AC-CZT测得的MBF和MFR与PET有合理相关性,可用于预测异常MFR和负荷MBF以及区分正常与CAD患者。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: PET/CT is the standard for quantitative assessments of myocardial blood flow (MBF), but it requires short-lived-tracers, costly, and not widely available. SPECT with Cadmium Zinc Telluride (CZT) detectors allows dynamic acquisition and quantitation of MBF. The study aims were to compare MBF measurements by 99mTc-tetrofosmin-CZT to N13NH3 PET/CT after regadenoson-induced coronary hyperemia and to evaluate the effect of attenuation correction (AC). METHODS: 54 patients were evaluated at rest and during vasodilation by 99mTc-tetrofosmin-CZT and N13NH3 PET/CT within 2 weeks. MBF and MBF reserve (MFR) were measured by CZT with or without AC (NAC). RESULTS: The global rest MBF was 0.76 ± 0.19 mL/min/gr by PET and 0.76 ± 0.24 by AC-CZT (P = NS) and 1.14 ± 0.4 by NAC-CZT (P < 0.001 vs PET and AC-CZT). Stress MBF was higher when measured by PET than AC-CZT (1.87 ± 0.45 vs 1.62 ± 0.68 mL/min/gr, P < 0.0008), but lower than NAC-CZT (2.36 ± 1.1, P < 0.0003). The MBF reserve ratio (MFR) was higher by PET than AC-CZT (2.52 ± 0.56 vs 2.22 ± 1 (P < 0.009) and NAC-CZT (2.18 ± 1.0, P < 0.004). Linear regression was better between PET (MFR and stress MBF) and AC-CZT than between PET and NAC-CZT. ROC curve analysis showed the significant ability of AC-CZT to predict MFR < 2 and stress MBF < 1.7 (AUC = 0.75 and 0.82 respectively) and to differentiate between normal and CAD patients (AUC = 0.747 and 0.892 for MFR and stress MBF, respectively). CONCLUSIONS: Our data show a reasonable correlation between MBF and MFR measured by N13NH3-PET and 99mTc-Tetrofosmin-CZT SPECT. NAC-CZT overestimates MBF. AC is recommended when using CZT for measuring MBF.
引用
- Giubbini R, Bertoli M, Durmo R, et al. Comparison between N13NH3-PET and 99mTc-Tetrofosmin-CZT SPECT in the evaluation of absolute myocardial blood flow and flow reserve. J Nucl Cardiol. 2019. PMID: 31728817. doi:10.1007/s12350-019-01939-x
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。