CZT通用SPECT相机与专用心脏CZT设备的心肌灌注及功能对比

Head-to-head comparison of a CZT-based all-purpose SPECT camera and a dedicated CZT cardiac device for myocardial perfusion and functional analysis.
PubMed 文献 · J Nucl Cardiol · 2019-08-05

本研究纳入19例序贯接受低剂量99mTc-替曲膦心肌灌注显像的患者,分别用CZT通用SPECT相机(Discovery NM/CT 670)与专用心脏CZT相机(Alcyone,Discovery 530c)采集。结果显示两台设备在负荷与静息节段心肌摄取百分比、灌注诊断一致性及左心室功能与结构参数上均呈高度相关。

要点

  • 两台相机节段心肌摄取百分比在负荷(R=0.90)与静息(R=0.88)均高度相关。
  • 灌注分析诊断一致性优秀(Cohen's κ 0.85)。
  • 左心室射血分数、峰值充盈率与质量的相关性分别为R=0.95、0.97、0.98。

核心结论

  • 通用型Discovery 670 CZT-SPECT相机行心肌灌注显像可行,与当前最先进的专用心脏CZT相机表现相当。

研究设计与样本

  • 纳入19例患者,序贯使用Alcyone与Discovery 670两台相机行低剂量99mTc-替曲膦心肌灌注显像。
  • 负荷期给药148-185 MBq,静息期给药296-370 MBq。
  • 对每次扫描行灌注数据定量(示踪剂摄取百分比)与半定量分析,并比较左心室主要功能与结构参数。

主要发现

  • 两台相机节段心肌摄取百分比在负荷时R=0.90(P<0.001),静息时R=0.88(P<0.001),Bland-Altman一致性界限窄。
  • 两台相机灌注分析的诊断一致性优秀,Cohen's κ为0.85。
  • 左心室射血分数(R=0.95)、峰值充盈率(R=0.97)与质量(R=0.98)的评估均高度相关。

临床怎么用

  • 结果提示通用型CZT-SPECT相机可用于心肌灌注显像,并与专用心脏CZT设备结果可比。
  • 作者称其为初步结果,临床推广前仍需进一步验证。

局限性

  • 作者表述为初步结果(preliminary results)。
  • 样本量仅19例。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To compare the outputs of a novel all-purpose SPECT camera equipped with CZT detectors (Discovery NM/CT 670) with the state-of-the-art represented by a dedicated CZT (Alcyone, Discovery 530c) cardiac camera in patients submitted to myocardial perfusion imaging (MPI). METHODS: We included 19 patients that underwent sequential low-dose 99mTc-tetrofosmin (148-185 MBq during stress and 296-370 MBq at rest) MPI with Alcyone and Discovery 670 cameras. Quantitative (% tracer's uptake) and semi-quantitative analyses of perfusion data were performed for each scan. Moreover, major left ventricular (LV) functional and structural parameters were derived from each camera and compared. RESULTS: The two cameras showed excellent correlation for segmental myocardial % uptake at stress (R = 0.90; P < 0.001) and at rest (R = 0.88; P < 0.001) with narrow Bland-Altman limits of agreement. The level of diagnostic agreement of Discovery 670 and Alcyone cameras regarding perfusion analysis was excellent (Cohen's κ 0.85). Similarly, the two cameras showed excellent correlation in the evaluation of LV ejection fraction (R = 0.95), peak filling rate (R = 0.97), and mass (R = 0.98). CONCLUSIONS: Our preliminary results suggest that MPI with an all-purpose Discovery 670 CZT-SPECT camera is feasible, comparing well with the current state-of-the-art technology.